Eppendorf Centrifuge 5804/ 5804 R - Medium-throughput, High Versatility
Eppendorf ·Made in Germany · In Active Production

Eppendorf Centrifuge 5804/ 5804 R - Medium-throughput, High Versatility

Eppendorf Centrifuge 5804/ 5804 R - Medium-throughput, High Versatility — Centrifuges.

The Centrifuge 5804 is the non-refrigerated benchtop member of Eppendorf's 58xx family, driven by a 900 W motor and supplied for 230 V with a type F plug. It reaches 200–14,000 rpm and 20,913 × g with fixed-angle rotors, while swing-bucket and plate rotors are limited to 4,500 × g and 2,250 × g respectively. It serves medium-throughput pelleting and plate processing where ambient-temperature runs are acceptable.

Full specifications

Core Performance & Functional Specifications
Acceleration/braking Ramps 10
Blood Collection Tubes 56 mm
Capacity 4 gram
Conical Tubes 40 mL
Dimensions (w × D × H) 46,6 × 55,0 × 33,7 cm
Max. Rcf 20913 g
Max. Rcf With Fixed-angle Rotor 20913 g
Max. Rcf With Plate Rotor 2250 g
Max. Rcf With Swing-bucket Rotor 4500 g
Number Of Programs 35
Plates 10
Rotational speed 200...14000 rpm
Rotor A-4-44 (4x100 Ml) 65 dB
Rotor F-34-6-38 51 dB
Rotor Fa-45-6-30 (6×50 Ml) 55 dB
Timer 1 - 99 min
Rotors available mfr 12
Dimensions, Weight & Mechanical Design
Footprint (dimensions W/o Front Panel, Wxd) 47 cm
Height (with Open Lid) 74 cm
Product Weight 55 kg
Electrical, Power & Drive Specifications
Max. Power Consumption 900 W
Voltage 230 V
Interfaces, Connections & Control
Power Plug power plug type F
Operating Conditions, Environment & Protection
Cooling non-refrigerated
Noise level 65 dB

Description

Construction and series positioning

Rotor-dependent performance

Drive and control

Gaps in the supplied data

Integration & Operating Notes

Electrical
230 V supply with type F (Schuko) plug required

The unit is wired for 230 V mains and terminated with a type F plug, so the bench outlet must be a compatible Schuko socket or an adapter must be provided. The 900 W motor implies supply and circuit capacity sized for the full drive load, not just standby.

Factory Specification Sheet
Process Limits
Never exceed the RCF limit of the installed rotor type

Maximum RCF differs by rotor format: 20,913 × g for fixed-angle, 4,500 × g for swing-bucket, and 2,250 × g for plate rotors. Setting a speed that exceeds the installed rotor's limit risks rotor failure; the run speed must be chosen from the rotor in use, not from the instrument maximum.

Factory Specification Sheet
Control & I/O
Program memory and ramps must be set per rotor and application

The controller offers 35 program slots and 10 acceleration/braking ramps. Because ramps and stored speeds are recalled per program, each program should be verified against the rotor installed at run time; a program written for a fixed-angle rotor is not valid for a swing-bucket or plate rotor.

Factory Specification Sheet
Cooling & Media
Ambient-temperature operation only; no refrigeration available

This model is non-refrigerated and no temperature range is specified. Heat-sensitive samples that require holding below ambient must be run on the refrigerated 5804 R sibling instead; on the 5804, run duration and sample load should be planned around warming from the 900 W drive.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 5804 rather than the refrigerated 5804 R or the larger 5810/5810 R?

Pick the 5804 when runs are at ambient temperature and 4,500 × g swing-bucket / 20,913 × g fixed-angle performance covers the protocol. Choose the 5804 R if samples must be held cold; choose the 5810/5810 R when higher capacity than the 5804's rotor set is needed.

What rotor and adapter set do I need to buy alongside the base unit, and how does rotor choice change the achievable g-force?

Twelve rotor types are supported. Fixed-angle rotors reach 20,913 × g, swing-bucket rotors 4,500 × g, and plate rotors 2,250 × g. Rotors and adapters are separate items; the base unit alone does not define the achievable RCF or the vessel formats.

What maintenance and consumable items are outside the base delivery?

Rotors, buckets and adapters are separate order items, and no maintenance schedule or service-interval data was supplied in the available documents. Confirm rotor inspection intervals and spare-part lists with Eppendorf before commissioning; the base unit includes no rotor.