Eppendorf DASbox ® Mini Bioreactor System
Eppendorf ·Made in Germany · In Active Production

Eppendorf DASbox ® Mini Bioreactor System

Eppendorf DASbox ® Mini Bioreactor System — Bioreactors & fermenters.

The DASbox Mini Bioreactor System is a 4-fold parallel bioreactor platform with direct overhead drives, TMFC gas mixing, and autoclavable glass vessels of 350 mL total volume. It supports cell culture, bacteria, yeast, fungi, insect, and stem cells, with interchangeable vessels and 20–2500 rpm agitation. Designed for scale-down process development and parallel screening in bioprocess laboratories.

Full specifications

Core Performance & Functional Specifications
Application Cell culture
Bacteria/yeast/fungi yes
Dimensions (w × D × H) 30 x 70 x 49 cm
Drive Direct overhead drives
Gas Supply TMFC; overlay and/or sparger
Impellers Marine
Insect yes
Interchangeable Vessels yes
Mammalian/animal yes
Number Of Parallel Units 24 unit
Software DASware control, other DASware optional
Speed Ranges 20...2500 rpm
Stem Cells yes
Sterilization Autoclavable
Total Volume 350 mL
Validatable no
Vessels Glass vessels
Feed lines per vessel mfr 2 (standard)/ 4 (optional)
Dimensions, Weight & Mechanical Design
Product Weight 44 kg
Typical Weight 44 kg
Electrical, Power & Drive Specifications
Typical power consumption (4-fold system without process computer) mfr 168 W (at 230 V)/ 154 W (at 115 V)
Interfaces, Connections & Control
User Interface Process computer with monitor
Standards, Safety & General Data
Standard Gas Flow Rates 0.04...5
Standard Gas Mixing 2

Description

Construction and positioning

Gas supply and mixing

Control and operation

Integration & Operating Notes

Electrical
Plan for 168 W nominal power per 4-fold system

The nominal power of 168 W applies to the 4-fold system without the process computer. When expanding to 24 parallel units, account for the cumulative power draw of all drive and control modules, and provide a dedicated circuit with appropriate overcurrent protection.

Factory Specification Sheet
Control & I/O
Process computer is required for DASware control

The system is operated via a process computer with monitor running DASware control. Without this computer, the bioreactor cannot be controlled or monitored. Integrate the computer into the lab network if remote access or data export is needed, and ensure uninterruptible power for the computer to prevent loss of control during runs.

Fermentors Bioreactors Brochure DASbox Most Compact
Maintenance
Autoclave glass vessels and replace feed lines per protocol

Glass vessels are autoclavable; follow the manufacturer's sterilization cycle. Feed lines (2 standard, 4 optional) are consumables and should be replaced between campaigns or when clogged. The direct overhead drives require periodic inspection of bearings and seals according to the service manual.

Fermentors Bioreactors Brochure DASbox Most Compact

Selection & Integration Q&A

When should I choose the DASbox Mini Bioreactor System with glass vessels instead of the single-use BioBLU 0.3 variant?

Choose glass vessels for autoclavable, repeated use with flexible experimental design and lower consumable cost. Choose BioBLU 0.3 single-use when avoiding cleaning validation, reducing turnaround time, or working with stem cells (BioBLU 0.3sc). The DASbox platform supports both, so the decision hinges on sterilization workflow and consumable budget.

What are the requirements for connecting gas supply to the DASbox TMFC inlets?

The TMFC inlets require clean, dry gas at a pressure within the controller's specified range (consult factory manual). Use appropriate tubing and fittings for the gas type (air, O2, N2, CO2). Standard gas mixing supports 2 gases; ensure each gas line is equipped with a filter and pressure regulator to protect the mass flow controllers.

What maintenance and consumables are required for the DASbox, and what is not included in base delivery?

Base delivery includes the 4-fold bioreactor system, glass vessels, and DASware control software on a process computer. Not included: feed lines (2 standard, 4 optional), sparger/overlay tubing, gas filters, and replacement drive seals. Autoclave glass vessels between runs; replace feed lines per protocol. No certificates are listed, so calibration may be optional.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Fermentors Bioreactors Brochure DASbox Most Compact
Brochure ·3,2 МБ ·4 pp.