Eppendorf Mastercycler ® nexus X2 - PCR Thermal Cycler
Eppendorf Mastercycler ® nexus X2 - PCR Thermal Cycler — Thermal cyclers (PCR).
The Mastercycler nexus X2 is a dual-block PCR thermal cycler with two independent aluminum thermoblocks, each accepting 64 vessels (0.2 mL and 0.5 mL) via a universal block, and Peltier heating/cooling with Triple Circuit Technology. It covers 4–99 °C block control with 3 °C/s heating and 2 °C/s cooling, and adds a 30–99 °C gradient function (1–12 °C span) for protocol optimization. It is intended as a networkable benchtop amplification engine that can also control two additional units over CAN.
Full specifications
| Block Homogeneity 20 – 72 °c | 0.3 °C |
| Control Accuracy | 0.2 °C |
| Gradient Block | ■ |
| Gradient Range | 1...12 °C |
| Gradient Temperature Control Range | 30...99 °C |
| Heating Rate | 3 °C/s |
| Heating Technology | Peltier elements, Triple Circuit Technology |
| Interfaces | CAN_in/CAN_out, Ethernet, USB |
| Lid Descent | flexlid ® technology, with thermal sample protection |
| Lid Temperature Range | 37...110 °C |
| Remote Device Monitoring And Notifications (visionize ® ) | VisioNize ® box enabled |
| Sample Capacity | 64 |
| Temperature Control Range Of The Block | 4...99 °C |
| Thermoblock | Aluminum |
| Dimensions (W × D × H) mfr | 25 × 41,2 × 33 cm / 9,8 × 16,2 × 13 in |
| Number of programs on device mfr | > 700 |
| Total Weight | 10.5 kg |
| Max. Power Consumption | 700 W |
| Power supply mfr | 230 V, 50 – 60 Hz |
| Power Plug | power plug type F |
| Temperature Control Mode | Fast, Standard, Safe |
| Cooling Rate | 2 °C/s |
Description
Construction and positioning
- Two independent aluminum thermoblocks, each with a universal format for 0.2 mL and 0.5 mL vessels; 64 samples per block.
- flexlid technology with thermal sample protection; lid temperature adjustable 37–110 °C.
- Peltier elements with Triple Circuit Technology for heating and cooling; no sibling models are listed, so this X2 configuration is the dual-block variant of the nexus line.
- 230 V, 50–60 Hz supply, type F power plug; 700 W motor power; 10.5 kg; 25 × 41.2 × 33 cm.
Thermal behavior and modes
- Block control range 4–99 °C; control accuracy 0.2 °C; block homogeneity 0.3 °C over 20–72 °C and ≤ ±0.4 °C at 90 °C.
- Ramp rates: 3 °C/s heating, 2 °C/s cooling.
- Three temperature-control modes (Fast, Standard, Safe) let the run trade speed against uniformity; the gradient function spans 1–12 °C within a 30–99 °C gradient control range for annealing optimization.
- No digitized curve points were supplied, so no melt-curve or ramp-profile data can be stated.
Control, networking and operation
- Interfaces: CAN in/CAN out, Ethernet, USB; the unit can control two other units in a network, and is VisioNize box enabled for remote monitoring and notifications.
- Intuitive graphic programming with more than 700 programs stored on the device; e-mail notification is not provided.
- The PDF digest covers consumables (twin.tec PCR plates, deepwell plates, sealing films/foils/mats) and contains no instrument installation or service procedures; installation and maintenance specifics beyond the spec sheet are therefore not documented here.
Integration & Operating Notes
The unit is specified only for 230 V, 50–60 Hz with a type F power plug and draws up to 700 W; the branch circuit and receptacle must match this, and no 115 V variant is documented for this model.
Plan CAN cabling and addressing if the X2 is to act as the controlling unit for two additional cyclers; Ethernet and USB are separate interfaces for PC/network connection and VisioNize box monitoring.
Annealing optimization using the gradient block must keep the gradient control range within 30–99 °C and the column-to-column difference within 1–12 °C; block control outside this range is not covered by the gradient function.
The flexlid with thermal sample protection operates over 37–110 °C; setting the heated lid outside this range is not supported and condensation control depends on keeping the lid within it.
The attached consumables brochure lists twin.tec PCR plates, deepwell plates and sealing films/foils/mats as separate catalog items; none of these, nor interface cables, are part of the base instrument delivery.
Selection & Integration Q&A
Pick the X2 when two independent 64-sample protocols must run simultaneously or when one unit must control two further cyclers over CAN. A single-block unit is sufficient for one protocol at a time and does not offer the two-unit network control role.
A 230 V, 50–60 Hz circuit with a type F receptacle sized for 700 W, plus CAN in/out cabling if it controls two other units; Ethernet and USB are available for PC and VisioNize box connection.
Block control is 4–99 °C, the heated lid 37–110 °C, and gradient operation 30–99 °C with a 1–12 °C span. Homogeneity is 0.3 °C from 20–72 °C and ≤ ±0.4 °C at 90 °C, so verify uniformity at your highest hold temperature.
PCR plates (twin.tec 96/384, skirted or unskirted), deepwell plates and sealing films, foils or mats are separate catalog items; interface cables for CAN/Ethernet/USB and the VisioNize box are also not part of the base delivery.