Eppendorf Mastercycler ® nexus X2 - PCR Thermal Cycler
Eppendorf ·Made in Germany · In Active Production

Eppendorf Mastercycler ® nexus X2 - PCR Thermal Cycler

Eppendorf Mastercycler ® nexus X2 - PCR Thermal Cycler — Thermal cyclers (PCR).

The Mastercycler nexus X2 is a dual-block PCR thermal cycler with two independent aluminum thermoblocks, each accepting 64 vessels (0.2 mL and 0.5 mL) via a universal block, and Peltier heating/cooling with Triple Circuit Technology. It covers 4–99 °C block control with 3 °C/s heating and 2 °C/s cooling, and adds a 30–99 °C gradient function (1–12 °C span) for protocol optimization. It is intended as a networkable benchtop amplification engine that can also control two additional units over CAN.

Full specifications

Core Performance & Functional Specifications
Block Homogeneity 20 – 72 °c 0.3 °C
Control Accuracy 0.2 °C
Gradient Block ■
Gradient Range 1...12 °C
Gradient Temperature Control Range 30...99 °C
Heating Rate 3 °C/s
Heating Technology Peltier elements, Triple Circuit Technology
Interfaces CAN_in/CAN_out, Ethernet, USB
Lid Descent flexlid ® technology, with thermal sample protection
Lid Temperature Range 37...110 °C
Remote Device Monitoring And Notifications (visionize ® ) VisioNize ® box enabled
Sample Capacity 64
Temperature Control Range Of The Block 4...99 °C
Thermoblock Aluminum
Dimensions (W × D × H) mfr 25 × 41,2 × 33 cm / 9,8 × 16,2 × 13 in
Number of programs on device mfr > 700
Dimensions, Weight & Mechanical Design
Total Weight 10.5 kg
Electrical, Power & Drive Specifications
Max. Power Consumption 700 W
Power supply mfr 230 V, 50 – 60 Hz
Interfaces, Connections & Control
Power Plug power plug type F
Temperature Control Mode Fast, Standard, Safe
Operating Conditions, Environment & Protection
Cooling Rate 2 °C/s

Description

Construction and positioning

Thermal behavior and modes

Control, networking and operation

Integration & Operating Notes

Electrical
Provide a 230 V, 50–60 Hz supply with type F receptacle

The unit is specified only for 230 V, 50–60 Hz with a type F power plug and draws up to 700 W; the branch circuit and receptacle must match this, and no 115 V variant is documented for this model.

Factory Specification Sheet
Control & I/O
CAN in/out allows one unit to control two others in a network

Plan CAN cabling and addressing if the X2 is to act as the controlling unit for two additional cyclers; Ethernet and USB are separate interfaces for PC/network connection and VisioNize box monitoring.

Factory Specification Sheet
Process Limits
Gradient runs are confined to 30–99 °C with 1–12 °C span

Annealing optimization using the gradient block must keep the gradient control range within 30–99 °C and the column-to-column difference within 1–12 °C; block control outside this range is not covered by the gradient function.

Factory Specification Sheet
Process Limits
Lid temperature must be set within 37–110 °C

The flexlid with thermal sample protection operates over 37–110 °C; setting the heated lid outside this range is not supported and condensation control depends on keeping the lid within it.

Factory Specification Sheet
Maintenance
Consumables and sealing are separate from the instrument

The attached consumables brochure lists twin.tec PCR plates, deepwell plates and sealing films/foils/mats as separate catalog items; none of these, nor interface cables, are part of the base instrument delivery.

Consumables Brochure Plates Unlimited

Selection & Integration Q&A

When should I choose the nexus X2 rather than a single-block nexus unit?

Pick the X2 when two independent 64-sample protocols must run simultaneously or when one unit must control two further cyclers over CAN. A single-block unit is sufficient for one protocol at a time and does not offer the two-unit network control role.

What supply and network infrastructure does the X2 require?

A 230 V, 50–60 Hz circuit with a type F receptacle sized for 700 W, plus CAN in/out cabling if it controls two other units; Ethernet and USB are available for PC and VisioNize box connection.

What are the thermal operating limits I must respect in a run?

Block control is 4–99 °C, the heated lid 37–110 °C, and gradient operation 30–99 °C with a 1–12 °C span. Homogeneity is 0.3 °C from 20–72 °C and ≤ ±0.4 °C at 90 °C, so verify uniformity at your highest hold temperature.

What consumables and accessories are not included with the instrument?

PCR plates (twin.tec 96/384, skirted or unskirted), deepwell plates and sealing films, foils or mats are separate catalog items; interface cables for CAN/Ethernet/USB and the VisioNize box are also not part of the base delivery.