Eppendorf Mastercycler ® X50 - PCR Thermocycler
Eppendorf ·Made in Germany · In Active Production

Eppendorf Mastercycler ® X50 - PCR Thermocycler

Eppendorf Mastercycler ® X50 - PCR Thermocycler — Thermal cyclers (PCR).

The Mastercycler X50 is a 96-well Peltier thermal cycler with a silver thermoblock, six Peltier elements, and a 4–99 °C block range. It offers a 2D gradient over 12 columns and 8 rows (1–30 °C) and four control modes (Fast, Intermediate, Standard, Safe). It is used for PCR amplification and thermal cycling protocols requiring precise temperature uniformity.

Full specifications

Core Performance & Functional Specifications
Block Homogeneity 20 – 72 °c 0.2 °C
Control Accuracy 0.15 °C
Gradient Range 1...30 °C
Gradient Temperature Control Range 30...99 °C
Heating Rate 10 °C/s
Heating Technology Six peltier elements
Sample Capacity 96
Temperature Control Range Of The Block 4...99 °C
Thermoblock Silver
2D-Gradient block mfr over 12 columns and 8 rows
Interfaces, Connections & Control
Temperature Control Mode Fast, Intermediate, Standard, Safe

Description

Construction and positioning

Thermal behavior and control modes

Networking and operation

Integration & Operating Notes

Process Limits
Do not exceed 99 °C block or gradient setpoints

The block control range tops out at 99 °C and the gradient control range at 99 °C. Setting higher targets is outside the verified operating envelope and may invalidate results or stress the Peltier elements; keep protocols within 4–99 °C (block) and 30–99 °C (gradient).

Factory Specification Sheet
Process Limits
Gradient span limited to 1–30 °C across 12 columns

The 2D gradient covers 12 columns and 8 rows with a span of 1–30 °C, and gradient control is only available between 30 and 99 °C. Annealing optimization steps below 30 °C cannot use the gradient function; plan column temperatures accordingly.

Factory Specification Sheet
Control & I/O
One unit can act as network master for up to 9 slaves

The X50 can control 9 other units in a network but cannot itself be controlled by another unit. In a multi-instrument setup, designate this unit as the master or verify that the controlling unit supports the required network role before wiring.

Factory Specification Sheet
Maintenance
Consumables and sealing options are not included

The attached consumables brochure lists twin.tec PCR plates (96/384, skirted, semi-skirted, unskirted, divisible), films, foils, mats, and plate lids as separate catalog items. Base delivery does not include plates or sealing consumables; specify them separately for the 96-well format.

Consumables Brochure Plates Unlimited

Selection & Integration Q&A

When should I choose the Mastercycler X50 rather than a 384-well variant?

Choose the X50 for 96-well plates and 0.1/0.2 mL tubes; it does not support 384-well plates. If your throughput requires 384-well format, this model is not suitable and a 384-well-capable instrument must be selected instead.

What consumables and adjacent labware must be specified with this unit?

The 96-well block accepts twin.tec PCR plates (skirted, semi-skirted, unskirted, divisible) and 0.1/0.2 mL tubes. Sealing films, foils, mats, and plate lids are separate catalog items and are not part of base delivery; order them alongside the instrument.

Can this unit be placed in a multi-instrument network, and what role can it take?

Yes. The X50 can control up to 9 other units in a network, but it cannot itself be controlled by another unit. It must therefore serve as the master in any networked configuration; verify cabling and master assignment before commissioning.

What maintenance or service items are not included in base delivery?

No maintenance schedule or service-kit details are provided in the supplied documents. The only accessory information available concerns consumables (plates, films, foils, mats, lids), which are catalog items separate from the instrument; confirm service intervals with the manufacturer.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Consumables Brochure Plates Unlimited
Brochure ·6,1 МБ ·19 pp.
View PDF