Eppendorf New Brunswick Innova ® 42/42R - Stackable Incubator Shaker
Eppendorf ·Made in Germany · In Active Production

Eppendorf New Brunswick Innova ® 42/42R - Stackable Incubator Shaker

Eppendorf New Brunswick Innova ® 42/42R - Stackable Incubator Shaker — Laboratory shakers.

The Innova 42/42R is a stackable incubator shaker with a triple-eccentric counterbalanced drive and solid-state DC brushless motor, offered in non-refrigerated (42) and refrigerated (42R) configurations. It covers 10–80 °C with 0.25 °C uniformity and 25–400 rpm (25–300 rpm for flasks >500 mL when stacked). It serves as a benchtop, under-bench or stacked incubation platform for temperature-controlled shaking.

Full specifications

Core Performance & Functional Specifications
Audible And Visual Alarms yes
Available Program Modes -232
Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) B, U, S
Door Opening opening to front
Drive Mechanism triple-eccentric counterbalanced drive
Gassing Manifold yes
Internal Convenience Light yes
Memory non volatile with automatic power failure restart
Motor Type Solid state, DC brushless motor
Multi-step Programming yes
Photosynthetic Lighting no
Platform Size (cm/in) 46 cm
Refrigerated −
Slide-out Platform no
Temperature range 10...80 °C
Temperature uniformity 0.25 °C
Timer mfr 0.1 – 99.9 h, continuous
Dimensions (W × D × H) mfr 63,5 × 74,5 × 81,8 cm / 25 × 29,3 × 32,8 in
In chamber power receptacle mfr yes
Speed range mfr 25 – 400 rpm; 25 – 300 rpm when using flasks > 5
Steps per program mfr 15
Stored programs mfr 4
Dimensions, Weight & Mechanical Design
Product Weight 98.5 kg
External depth (with lid/door open) mfr 131 cm (51.6 in)
Electrical, Power & Drive Specifications
Power supply mfr 230 V ± 10 %, 50 Hz
Interfaces, Connections & Control
Power Plug type F/type I
Operating Conditions, Environment & Protection
Water/humidity Monitoring yes

Description

Construction and series positioning

Behavior and operating envelope

Control, connection and operation

Integration & Operating Notes

Mounting
Allow 131 cm depth clearance for the open door

The external depth with lid/door open is 131 cm, so the installation envelope must accommodate this swing plus service access; benchtop, under-bench and stacked configurations are supported, but the 98.5 kg unit weight requires a load-rated surface or stacking frame.

Factory Specification Sheet
Process Limits
Stacked operation caps speed at 300 rpm for flasks >500 mL

When the device is stacked and flasks larger than 500 mL are used, the speed range is limited to 25–300 rpm instead of 25–400 rpm; program setpoints above 300 rpm must be avoided in that configuration to prevent exceeding the stated envelope.

Factory Specification Sheet
Control & I/O
Non-volatile memory restarts automatically after power failure

The controller stores programs in non-volatile memory and performs an automatic power-failure restart; engineers should account for this behavior in protocols where an unintended restart after mains interruption is not acceptable, and use the audible/visual alarms to detect fault states.

Factory Specification Sheet
Maintenance
No maintenance interval is stated in the supplied documents

The supplied factory specification and brochure digest do not define service intervals, consumables or accessory cables; these must be obtained from the full instruction manual before commissioning, and the in-chamber power receptacle and gassing manifold should be checked as part of site acceptance.

Shakers Brochure Incubator Shakers Incubator Shaker Portfolio

Selection & Integration Q&A

When should I choose the Innova 42/42R rather than another incubator shaker in the portfolio?

Choose this model when you need a stackable, front-door incubator shaker covering 10–80 °C with 0.25 °C uniformity and up to 400 rpm. The supplied facts list it as non-refrigerated; the portfolio also covers refrigerated (R) variants, so confirm the R configuration if sub-ambient control is required.

What supply and adjacent-system requirements apply to this shaker?

It requires 230 V ± 10 %, 50 Hz with a type F or type I plug. An in-chamber power receptacle and gassing manifold are provided for ancillary devices; no external drive or controller is needed because the solid-state DC brushless motor and non-volatile program memory are integral.

What maintenance and accessories are defined for this model?

The supplied documents state no service intervals, consumable list or included cable set; only the type F/type I power plug is specified. Request the full instruction manual for maintenance scheduling and confirm accessory scope (platform, gassing manifold connections) before ordering.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Shakers Brochure Incubator Shakers Incubator Shaker Portfolio
Brochure ·673,8 КБ ·1 pp.
View PDF