Eppendorf New Brunswick Innova ® 44/44R - Stackable Incubator Shaker
Eppendorf ·Made in Germany · In Active Production

Eppendorf New Brunswick Innova ® 44/44R - Stackable Incubator Shaker

Eppendorf New Brunswick Innova ® 44/44R - Stackable Incubator Shaker — Laboratory shakers.

The New Brunswick Innova 44/44R is a floor-standing, stackable incubator shaker built on a triple-eccentric counterbalanced drive with a solid-state DC brushless motor. It covers 10–80 °C with 0.25 °C uniformity and 25–400 rpm at a 2.5 cm orbit, distinguishing it as the economical workhorse of the Innova stackable line. It serves as a high-capacity incubation and shaking platform for microbial and cell culture in shared laboratory infrastructure.

Full specifications

Core Performance & Functional Specifications
Audible And Visual Alarms yes
Available Program Modes -232
Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) F, S
Door Opening glide up
Drive Mechanism triple-eccentric counterbalanced drive
Gassing Manifold yes
Internal Convenience Light yes
Memory non volatile with automatic power failure restart
Motor Type Solid state, DC brushless motor
Multi-step Programming yes
Photosynthetic Lighting no
Platform Size (cm/in) 76 cm
Refrigerated −
Slide-out Platform yes
Temperature range 10...80 °C
Temperature uniformity 0.25 °C
Timer mfr 0.1 – 99.9 h, continuous
Dimensions (W × D × H) mfr 134,9 × 84,3 × 66 cm / 53,1 × 33,2 × 26 in
Speed range mfr 25 – 400 rpm with 2.5 cm orbit; 25 – 300 rpm wit
Steps per program mfr 15
Stored programs mfr 4
Dimensions, Weight & Mechanical Design
External Height (with Lid/door Open) 113.9
Product Weight 238 kg
Electrical, Power & Drive Specifications
Power supply mfr 230 V ± 10 %, 50 Hz
Interfaces, Connections & Control
Power Plug type F/type I
Operating Conditions, Environment & Protection
Water/humidity Monitoring no

Description

Construction and series positioning

Behavior under load and in different modes

Installation, control and operation

Integration & Operating Notes

Mounting
Stack height caps the usable speed envelope

The drive is rated 25–400 rpm only as a single unit at the 2.5 cm orbit; two stacked units are limited to 25–300 rpm and three stacked units to 25–250 rpm. Plan the stack configuration before writing process setpoints, because a method validated at 400 rpm cannot be run in a multi-unit stack.

Factory Specification Sheet, Speed range
Control & I/O
Power-failure restart resumes the running program automatically

Memory is non-volatile with automatic power-failure restart, so after a mains interruption the shaker restarts the stored program without operator action. Anyone relying on a manual restart as a safety step must account for this behavior, and alarms should be wired to a remote monitoring point if unattended operation is intended.

Factory Specification Sheet, memory
Process Limits
No humidity or photosynthetic control in this model

Water/humidity monitoring is absent and photosynthetic lighting is not fitted; the gassing manifold is the only atmosphere-related interface. Processes requiring controlled humidity or light cycles must use the flagship S44i variant instead, or provide external conditioning.

Factory Specification Sheet, water-humidity monitoring and photosynthetic lighting
Maintenance
Service intervals are not documented in the supplied materials

The flyer digest contains no maintenance schedule, error-code list, or consumable/accessory list for the 44/44R. The user must obtain the full instruction manual for drive service intervals, platform accessories, and alarm-code handling before commissioning; none of these are included in the base delivery documentation provided here.

Shakers Flyer Innova S44i / Innova 44/44R

Selection & Integration Q&A

When should I choose the Innova 44/44R rather than the flagship Innova S44i?

Pick the 44/44R for heating-only incubation at 10–80 °C with 25–400 rpm shaking and no humidity or light control. Choose the S44i when you need refrigerated operation below ambient, photosynthetic lighting, or the flagship user interface; the 44/44R is the economical workhorse of the stackable pair.

Can I run a 400 rpm method in a two-unit stack?

No. Two stacked units are limited to 25–300 rpm and three stacked units to 25–250 rpm; 400 rpm is available only as a single unit at the 2.5 cm orbit. Revalidate any method above the stacked limit or run it in a single unstacked unit.

What maintenance and accessories are not included in the base delivery?

The supplied flyer documents no maintenance schedule, error-code list, platform clamps, flask holders, or replacement drive parts; these must be ordered separately and the full instruction manual obtained. Plan for periodic drive and platform inspection per that manual, since no interval is stated in the available facts.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Shakers Flyer Innova S44i New Brunswick Innova 44 44R Stackable Incuba
Brochure ·775,8 КБ ·2 pp.
View PDF