Peter Huber Kältemaschinenbau CC-200BX Brückenthermostat
Peter Huber Kältemaschinenbau CC-200BX Brückenthermostat — Thermostatic baths.
The CC-200BX is a bridge thermostat with a telescopic arm (up to 884 mm) and 128 mm immersion depth, designed to clamp onto an open bath. Its 200–240 V single- or two-phase unit delivers 1.5–2.1 kW heating and reaches 28–200 °C uncooled, 20–200 °C with water cooling, or -20–200 °C with a refrigerator. It circulates thermofluid through external circuits via a variable-speed pressure/suction pump.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Brueckenthermostate Series — Model Comparison & Selection
| Model | Current consumption, A | Dimensions w d h | Max delivery flow rate, liter/minute | Temperature range, °C | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-200BX Brückenthermostat Current | 10 | 345 x 200 x 326 | 22 | 28–200 | 7 | 5–40 |
| CC-300BX Brückenthermostat | 16 | 345 x 190 x 392 | 25 | 28–300 | 13 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 345 x 200 x 326 mm |
| Maximum delivery flow rate | 27 l/min (pressure pump max.; pressure 0, 7 bar) suction pump max.; pressure 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 28...200 °C with cooling unit: -20...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Weight | 7 kg |
| Current consumption | 10 A |
| Power supply | 200-240V 1~/2~ 50/60Hz V; Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Bridge-mounted thermostat: the housing and all wetted parts are stainless steel; the pressure and suction pumps are high-resistant plastic.
- Telescopic arm extends to 884 mm and the immersion depth is 128 mm, so the unit spans an open bath rather than forming a self-contained tank.
- Within the Brückenthermostate series, the CC-200BX is the smaller sibling of the CC-300BX; the CC-200BX is the choice when the 128 mm immersion depth and 884 mm arm reach suit the existing bath, while the CC-300BX is the neighbor for larger bath geometries.
- Safety class III/FL with adjustable overtemperature protection to DIN 12876; pressure side rated Art. 4.3 PED.
Thermal behavior and media
- Temperature range is configuration-dependent: 28–200 °C without cooling, 20–200 °C with water cooling, and -20–200 °C with a refrigerator. The uncooled range cannot go below 28 °C, so sub-ambient work requires the water-cooling or refrigerator configuration.
- Heating capacity 1.5–2.1 kW; temperature stability 0.02 K at 70 °C.
- Thermofluid kinematic viscosity must not exceed 50 mm²/s, which constrains fluid selection at the low end of the range.
- Pump: variable-speed pressure and suction pump; pressure side up to 27 l/min at 0.7 bar, suction side up to 22 l/min at 0.4 bar. Pump connections are M16x1 male.
Installation, control and operation
- Power: 200–240 V, 1~ or 2~, 50/60 Hz, max. current 10 A; external branch-circuit protection to UL 489 required, fuse 10–16 A. Voltage and frequency tolerances of ±10% apply only if they do not run in opposite directions.
- Single- and two-phase units ship with a power cable and country-specific plug, which must be specified at ordering.
- Controller: Pilot ONE; interfaces Ethernet, USB host and device, RS232; external Pt100 sensor connection.
- Ambient 5–40 °C, IP20, dry indoor location only; Class A Group 1 to EN55011, so additional EMC measures may be needed in residential or light-industrial settings.
Integration & Operating Notes
The unit draws up to 10 A at 200–240 V and requires external branch-circuit protection according to UL 489 with a fuse rated between 10 A and 16 A. Verify the site supply and protection device before installation; the country-specific plug must be specified at ordering.
Voltage and frequency tolerances are ±10% only if they do not run in the opposite direction. A supply at -10% voltage together with +3% frequency is outside the permitted envelope and must be corrected before commissioning.
Without cooling the operating range starts at 28 °C; water cooling extends it to 20 °C and a refrigerator to -20 °C. Select the cooling configuration against the lowest required process temperature before specifying the unit.
The maximum permissible kinematic viscosity is 50 mm²/s. At low temperatures viscosity rises, so the fluid must be chosen so that it stays within this limit across the whole operating range; otherwise pump delivery and heat transfer degrade.
The immersion depth is 128 mm and the telescopic arm reaches up to 884 mm. Confirm that the bath geometry allows the heater and sensor to be immersed to this depth and that the bridge arm spans the bath opening.
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar. Calculate the pressure drop of the connected external loop; if it exceeds the pump capability, flow falls below the required rate.
IP20 protection and an ambient limit of 5–40 °C rule out wet or dusty locations. Class A Group 1 to EN55011 means additional EMC measures may be needed in residential or light-industrial environments.
The unit has adjustable overtemperature protection according to DIN 12876 and safety class III/FL. Adjust the limit to the bath fluid and process before unattended operation; the stainless-steel wetted parts and plastic pumps define the cleaning and fluid-compatibility practice.
Selection & Integration Q&A
Pick the CC-200BX when the existing bath matches its 128 mm immersion depth and 884 mm telescopic arm reach. The CC-300BX is the neighbor for larger bath geometries; confirm the arm span and immersion depth against your bath before deciding.
A 200–240 V, 1~ or 2~, 50/60 Hz supply with max. 10 A, external branch-circuit protection to UL 489, and a fuse of 10–16 A. The country-specific plug must be specified at ordering.
Ambient 5–40 °C, IP20, dry indoor location only. Voltage and frequency tolerances of ±10% apply only if they do not run in opposite directions. Class A Group 1 to EN55011 may require extra EMC measures in light-industrial settings.
The power cable and country-specific plug must be specified at ordering. External branch-circuit protection to UL 489 is site-provided. The external Pt100 sensor, thermofluid, and cooling water or refrigerator for sub-ambient operation are separate.