Peter Huber Kältemaschinenbau CC-304B Wärme-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-304B Wärme-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-304B Wärme-Badumwälzthermostat — Thermostatic baths.

The CC-304B is a 5-litre stainless-steel bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 28–300 °C without cooling (20–300 °C with water cooling, -20–300 °C with refrigerator). Its 0.02 K stability and up to 3 kW heating distinguish it from the larger CC-308B and CC-315B siblings. It serves as a precise temperature source for external circuits or direct bath immersion.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 300C Series — Model Comparison & Selection

Model Current consumption, ADimensions w d hFuse rating, AWeight, kgAmbient temperature limit, °CController
CC-304B Wärme-Badumwälzthermostat Current 14210 x 335 x 39216145–40Pilot ONE
CC-308B Wärme-Badumwälzthermostat 14242 x 404 x 39216185–40Pilot ONE
CC-315B Wärme-Badumwälzthermostat 16335 x 382 x 43320225–40Pilot ONE

Full specifications

Core Performance & Functional Specifications
Bath volume 5 l
Dimensions (w × D × H) 210 x 335 x 392 mm
Maximum delivery flow rate 25 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 28...300 °C
with cooling unit: -20...300 °C · with water cooling: 20...300 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 2, 2 - 3 kW
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump; 0, 4 bar)
Bath opening WxD/BD mfr 130 x 100 / 155 mm
Filling volume min. mfr 3, 2 l (min)
Dimensions, Weight & Mechanical Design
Weight 14 kg
Electrical, Power & Drive Specifications
Current consumption 14 A (max)
Fuse rating 16 A
Power supply 200-240V 1~/2~ 50/60Hz V; Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation and control

Integration & Operating Notes

Electrical
Branch circuit must be rated for 14 A continuous with 16 A fuse

The unit draws up to 14 A at 200–240 V; the supply circuit must be protected by a 16 A fuse and, for UL markets, a UL 489-listed breaker. Undersized wiring or a lower-rated breaker will trip during heat-up at full heating power.

CC-304B Technical Data Sheet, Specifications
Process Limits
Cooling method fixes the minimum achievable temperature

Without cooling the bath cannot go below 28 °C; water cooling lowers the limit to 20 °C, and only a refrigerator reaches -20 °C. The cooling utility must be decided before specifying the unit, because it changes the usable setpoint range and the required site services.

CC-304B Technical Data Sheet, Overview
Process Limits
Bath fluid viscosity must stay below 50 mm²/s at temperature

The pump is rated for a maximum kinematic viscosity of 50 mm²/s. At low temperatures many thermofluids become too viscous, causing cavitation, poor flow and possible pump overload; select a fluid whose viscosity at the lowest operating temperature remains within this limit.

CC-304B Technical Data Sheet, Specifications
Electrical
Order the country-specific plug with the unit

Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. Without this, the unit may arrive with a plug incompatible with the site socket, delaying commissioning.

CC-304B Technical Data Sheet, Installation
Control & I/O
Set adjustable overtemperature protection before operation

The unit includes adjustable overtemperature protection per DIN 12876. It must be configured for the specific bath fluid and application during commissioning; an incorrect setting either trips prematurely or fails to protect the fluid and system if control is lost.

CC-304B Technical Data Sheet, Safety

Selection & Integration Q&A

When should I choose the CC-304B rather than the CC-308B or CC-315B?

Pick the CC-304B for a 5-litre bath when external circuit volume and immersion load are small; the CC-308B and CC-315B offer larger baths for greater thermal stability under heavy external loads or larger immersed samples.

What pump connections and fluid viscosity limits apply to the external circuit?

Pump ports are M16x1 male, so hoses and adapters must match this thread. The bath fluid must have a kinematic viscosity no higher than 50 mm²/s at the lowest operating temperature to avoid pump overload.

What are the electrical supply and branch protection requirements for installation?

Supply is 200–240 V, 1~ or 2~, 50/60 Hz, drawing up to 14 A; provide a 16 A fuse and, for UL markets, a UL 489-listed breaker. Voltage and frequency tolerances of ±10% apply only if not opposing.

What maintenance and accessories are needed beyond the base delivery?

The base unit includes a power cable with country-specific plug only if specified at order. Bath fluid, hoses with M16x1 fittings, and any cooling-water or refrigerator circuit are not included and must be sourced separately; overtemperature protection must be set at commissioning.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-304B Technical Data Sheet
Datasheet ·95,7 КБ ·2 pp.
View PDF