Peter Huber Kältemaschinenbau DC30
Peter Huber Kältemaschinenbau DC30 — Immersion coolers.
The DC30 is a flow-through cooler with no integral pump, rated for -30 to 50 °C and 0.2 kW cooling at 15 °C evaporating temperature. Its M16x1 male fluid connection and 220–240 V 1~/2~ supply distinguish it from pumped siblings DC31/DC32. It serves as a compact chiller stage in external circulation systems.
Peter Huber Kältemaschinenbau Durchflusskuehler Series — Model Comparison & Selection
Full specifications
| Dimensions (w × D × H) | 190 x 250 x 360 mm |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | no pump |
| Temperature range | -30...50 °C |
| Weight | 16 kg |
| Fuse rating | 2 A |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 1, 4 A |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 15:0, 2; 0:0, 15; -20:0, 07 °C / kW (at ambient 20°C) |
| Safety class | I / NFL |
Description
Construction and series positioning
- The DC30 is a pump-less flow-through cooler: the fluid circuit is driven by an external pump or system pressure, not by an onboard pump (pump type: no pump; pump speed adjustable: no).
- Fluid connection is a single M16x1 AG male thread; mating hoses/fittings must match this thread.
- Within the DC series, the DC30 is the base no-pump variant; siblings DC31 and DC32 are separate models whose configurations are not described here.
- Dimensions 190 × 250 × 360 mm, weight 16 kg, protection IP20, safety class I / NFL.
Thermal behavior and capacity
- Operating temperature range: -30 to 50 °C.
- Cooling capacity at 20 °C room temperature: 0.2 kW at 15 °C, 0.15 kW at 0 °C, 0.07 kW at -20 °C evaporating temperature. Capacity falls steeply as the operating point drops, so sizing must use the capacity at the actual evaporating temperature, not the 15 °C figure.
- Capacity data are valid at 20 °C ambient; higher room temperature may reduce cooling capacity.
- Refrigerant is R-290 (propane, A3, H220) with a 0.05 kg charge.
Electrical and control
- Supply: 220–240 V 1~/2~ 50/60 Hz; max current 1.4 A; fuse rating 2 A.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; combinations such as -5% voltage with +2% frequency are not permitted.
- Single/two-phase units are delivered with a power cable and country-specific plug, which must be specified at ordering.
- EMC classification: Class A, Group 1 to EN55011 — industrial environment; may cause radio interference in residential areas.
Installation and environment
- Ambient temperature must remain 5–40 °C.
- IP20 enclosure requires a dry, protected location without water or dust ingress.
- Space requirements around the unit must be respected per the operating conditions; clearance affects air-cooled condenser performance and service access.
Integration & Operating Notes
The cooler is rated only for an operating temperature range of -30 to 50 °C. Operating outside this window risks damage or invalid performance; select the unit only for applications whose fluid temperature stays inside this envelope.
Cooling capacity is 0.2 kW at 15 °C, 0.15 kW at 0 °C and 0.07 kW at -20 °C. Because capacity drops sharply as the operating point falls, the 15 °C figure must not be used for sizing a low-temperature duty; use the capacity at the real evaporating temperature.
The refrigerant is R-290 (propane, A3, H220) with a 0.05 kg charge. The installation area must be assessed for flammable refrigerant handling and ignition sources before commissioning.
Room temperature must be verified before commissioning. Capacity figures are valid at 20 °C ambient; if the ambient temperature rises, cooling capacity may drop, so warmer rooms require derating.
The device is IP20 and must not be exposed to water or dust ingress. Plan the installation in a dry, protected area and respect the space requirements around the unit for condenser airflow and service access.
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%. Combinations such as -5% voltage with +2% frequency are not allowed; verify the site supply quality against this combined limit before connection.
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering. Confirm the destination plug type so the correct cable set is supplied.
The equipment is EMC Class A, Group 1 to EN55011 and may cause radio interference in residential areas. It is intended for industrial environments; assess EMC compatibility with nearby sensitive equipment.
Selection & Integration Q&A
Choose the DC30 when the circulation is driven by an external pump or system pressure: it has no integral pump and a single M16x1 male fluid connection. Siblings DC31/DC32 are separate models; their configurations are not covered by the DC30 datasheet, so verify their pump and connection data before substituting.
The fluid side needs an external pump or system pressure and hoses/fittings matching the M16x1 AG male thread. The electrical side needs 220–240 V 1~/2~ 50/60 Hz, max 1.4 A, protected at 2 A, with voltage tolerance ±5% only together with frequency tolerance ±2%.
Install in a dry, protected location: IP20 forbids water or dust ingress. Ambient must stay between 5 and 40 °C, and capacity is rated at 20 °C room temperature, so warmer rooms require derating. Respect the space requirements around the unit for airflow and service access.
Single/two-phase units come with a power cable and country-specific plug, which must be specified at ordering. The pump is not included (no pump), and mating hoses/fittings for the M16x1 male connection are not part of the base delivery. No maintenance schedule is given in the available datasheet digest.