Peter Huber Kältemaschinenbau Piccolo 280 OLÉ Ultra-kompakter Laborkühler mit Peltiertechn
Peter Huber Kältemaschinenbau Piccolo 280 OLÉ Ultra-kompakter Laborkühler mit Peltiertechn — Recirculating chillers with Noise level 60 dB.
The Piccolo 280 OLÉ is a refrigerant-free Peltier-based recirculating chiller with a 4–70 °C working range and a 0.1–0.2 l internal circuit. Its 0.03 kW cooling capacity at 4 °C and 1.85 l/min pressure pump suit small, low-load loops where refrigerant handling is restricted.
Full specifications
| Noise level | 60 dB(A) |
| Pump connection | Connector with hose barb |
| Pump speed adjustable | no |
| Pump type | Pressure pump |
| Temperature range | 4...70 °C |
| Weight | 13 kg |
| Power supply | 100-240V 1~/2~ 50/60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Peltier (thermoelectric) cooling, refrigerant-free; no compressor, no refrigerant logistics or leak checks.
- Single model in the Piccolo series with no sibling variants listed; the OLÉ controller and OLED display define this modification.
- Compact 13 kg unit, IP20, safety class I / NFL per DIN 12876, pressure equipment Art. 4.3 PED.
Thermal behavior and media
- Working range 4–70 °C; cooling capacity is strongly temperature-dependent: only 0.03 kW at 4 °C, so low-temperature set points cannot support significant heat loads.
- Rated output is valid at 20 °C ambient; capacity falls as ambient rises, so enclosure ventilation and room temperature control are required.
- Internal cooling circuit volume is only 0.1–0.2 l, giving low thermal inertia and fast response but little buffering against load spikes.
- Pressure pump delivers up to 1.85 l/min at up to 0.95 bar; pump speed is not adjustable.
Installation, control and operation
- Accepts 100–240 V, 1~ or 2~, 50/60 Hz; ships with power cable and country-specific plug that must be specified at order.
- Upstream fuse must be sized between 10 A minimum and 16 A maximum.
- OLÉ controller with OLED display, 0.1 K resolution, USB (Device) and RS232 interfaces.
- Autostart after power failure restarts the unit automatically; assess against process safety concept.
- Class A, Group 1 per EN 55011: intended for industrial electromagnetic environments, not residential.
Integration & Operating Notes
At 4 °C the unit provides only 0.03 kW, so any external heat load approaching this value will prevent the set point from being reached or held. Verify the actual load at the lowest required temperature before selecting this model.
Published output is valid at 20 °C room temperature; higher ambient lowers cooling capacity. Install in a ventilated location and account for the actual room temperature when sizing the loop.
The manufacturer specifies a minimum fuse of 10 A and a maximum of 16 A. A breaker outside this window risks nuisance tripping or inadequate protection; coordinate the site circuit before commissioning.
Single- and two-phase 100–240 V units ship with a power cable and a country-specific plug. Specify the plug type at order time to avoid rework at installation.
The unit is rated IP20 and must not be exposed to water spray or dust. Place it in a dry, sheltered position; ambient temperature must remain between 5 °C and 40 °C for correct operation.
The autostart function automatically restarts the circulator after mains interruption. For unattended or sensitive processes, confirm whether this behavior is acceptable or must be disabled in the safety concept.
The pressure pump delivers up to 1.85 l/min at up to 0.95 bar and is not speed-adjustable. Size hoses and external circuit components so that pressure drop stays within this head at the required flow.
The internal cooling circuit holds only 0.1–0.2 l. Select fluid volume and expansion provisions accordingly; the low inertia gives fast response but little damping against sudden load changes.
Selection & Integration Q&A
Choose it when a refrigerant-free Peltier chiller is required and the load at the lowest set point is very small (0.03 kW at 4 °C). No sibling variants are listed for this series, so compare against other series if higher capacity or lower temperatures are needed.
Confirm the external circuit pressure drop fits the fixed pump (max 1.85 l/min at 0.95 bar) and that the 100–240 V, 1~/2~, 50/60 Hz supply is protected by a 10–16 A fuse. Specify the country-specific plug at order.
Install in a dry, protected location: IP20 enclosure, ambient 5–40 °C. Cooling capacity is rated at 20 °C and falls at higher ambient, so ventilation and room temperature control are required.
The unit ships with a power cable and country-specific plug (specify at order). No accessories, external hoses, fittings or coolant are listed as included; plan pump connections (hose barb) and fluid separately.