Labconco Crude Fiber Apparatus — Kjeldahl & extraction apparatus.

The Labconco Crude Fiber Digestion Apparatus is a six-position reflux digestion unit with individually adjustable 350 W ceramic-embedded heaters and Type 304 stainless steel condensers. It is supplied in two voltage-specific configurations, 115 V/50/60 Hz/18.3 A (3000100) and 230 V/50/60 Hz/9.11 A (3000200), both without a power cord. It prepares crude fiber samples under AOAC methods, with a 6-inch exhaust manifold and 3/8" NPT water connections.

Full specifications

Description

Construction and configuration

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Mounting
Exhaust ducting must match the 6-inch manifold connection

The manifold exhaust connection is 6 inches in diameter, sized for minimum static pressure loss while maintaining good transport velocity. Ducting or hood connections should be sized to this diameter to preserve exhaust performance; reducing or adapting the connection may compromise vapor removal.

User Manual 3002300 Rev. E, Component Identification, p.5
Control & I/O
Start-up check-out requires all six heater switches ON

During start-up and check-out, all heater control switches should be turned on to ensure proper performance. This verifies each of the six independent heater circuits and pilot lights before routine use; a position that fails to heat or indicate during this check should be investigated before samples are run.

User Manual 3002300 Rev. E, Start-Up and Check-Out, p.8
Process Limits
Beaker must be raised to the condenser head with the knob held

The elevation control knob must be turned 1/4 turn to the left while firmly held, and not released until the beaker is in position against the condenser head. Releasing early or misaligning the beaker can cause spillage or poor reflux; the procedure is part of normal operation for every sample.

User Manual 3002300 Rev. E, Normal Operation, p.9
Maintenance
Voltage-specific spares must match the installed model

Replacement heating elements are 3003100 (115 V, 350 W) and 3003200 (230 V, 350 W); infinite controls are 1315400 (120 V, 15 A) and 1315500 (230 V, 15 A). Ordering the wrong voltage variant will not function correctly. Condenser gasket 3022500 is a maintenance item subject to periodic replacement.

User Manual 3002300 Rev. E, Replacement Parts List, p.11

Selection & Integration Q&A

Should I choose the 115 V or 230 V configuration of the Crude Fiber Digestion Apparatus?

Choose 3000100 for 115 V, 50/60 Hz, 18.3 A supplies; choose 3000200 for 230 V, 50/60 Hz, 9.11 A. Both are six-position units with identical heaters and condensers. The decision is purely the available supply voltage and branch circuit capacity, since neither includes a power cord.

What plumbing and exhaust connections must be provided for installation?

Provide a 3/8" NPT cold water supply to the upper-left inlet and route the upper-right 3/8" NPT outlet to an open drain of at least 1/2" NPT for visual flow check. Exhaust ducting must match the 6-inch manifold connection to preserve transport velocity and low pressure loss.

What are the electrical supply and field-wiring requirements before first use?

Size the branch circuit for 18.3 A at 115 V (3000100) or 9.11 A at 230 V (3000200). No power cord is supplied; the main power switch must be wired per local electrical codes. During start-up check-out, turn all heater control switches on to verify every circuit and pilot light.

What maintenance items and accessories are not included in the base delivery?

The base unit includes 12 × 600 ml beakers but no power cord. Replacement parts available separately include heating elements (3003100/3003200), infinite controls (1315400/1315500), needle valve 1351600, condenser assembly 3020800, and gasket 3022500. Filtration accessories California Modified Büchner Funnel (5510000) and Oklahoma State Filter (5521000) are optional.