Labconco CT-Series Freeze Dryers
Labconco · In Active Production

Labconco CT-Series Freeze Dryers

Labconco CT-Series Freeze Dryers — Freeze dryers.

The Labconco CT-Series is a cold-trap freeze dryer available in three collector-temperature classes (-50°C, -84°C, -105°C) and two manifold configurations (4-port, 8-port). It is placed upstream of a vacuum pump and removes water or solvent from frozen material by sublimation, with collector temperature selected 15–20°C below the sample eutectic point.

Full specifications

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection, and control

Integration & Operating Notes

Mounting
Keep 3 in (7.6 cm) clear on both sides and rear

Airflow enters the right side and exits the left side of the cabinet, so walls or adjacent equipment closer than 3 inches (7.6 cm) to the sides or back will degrade refrigeration performance. Place the unit on a stable, flat, level surface.

User Manual, Section 2 Location, Space and Clearance Requirements
Process Limits
Collector must be 15–20°C below sample eutectic point

Select the -50°C, -84°C, or -105°C class from the sample eutectic temperature, not from ice volume. Samples at -40 to -60°C eutectic require cascade refrigeration below -75°C or a dry-ice/solvent trap; the standard -50°C collector is insufficient for these.

User Manual, Freeze Dry Rates and Freeze Dry Capacity
Process Limits
Do not exceed instantaneous ice load of the collector

Maximum instantaneous ice load is 1.6 L for -50°C, 2.1 L for -84°C, and 1.6 L for -105°C. Add pre-frozen samples one at a time and allow about 5 minutes of vacuum recovery; if vacuum does not recover, capacity has been exceeded and the sample should be removed.

User Manual, Start-up and Freeze Dry Capacity
Electrical
Dedicated circuit with protective earth is mandatory

All models are single-phase and require a dedicated outlet with an appropriate circuit breaker and a protective earthing ground. Typical operating currents range from 2.6 A to 13 A depending on model and voltage; for example, 8702X00 requires 115 V/60 Hz/13 A, while 8700X30/-40/-50/-70/-80/-90 require 230 V/50 Hz/3 A at 2.6 A typical.

User Manual, Electrical Requirements
Electrical
Facility power must stay within ±10% of nominal

Main supply voltage fluctuations must not exceed ±10% of nominal, and transient overvoltages up to 1500 V (115 V models) or 2500 V (230 V models) are allowed per Installation Category II. Verify facility power quality before installation; the environment is Pollution Degree 2, indoor use only.

User Manual, Appendix C Environmental Conditions
Process Limits
Vacuum pump must reach below 0.02 mbar

The user-supplied vacuum pump must achieve less than 0.02 mbar and its inlet fitting must accept 0.50 ID hose. Recommended operating vacuum for many applications is 0.133 mbar or less. Flammable solvents require an explosion-proof or solvent-suitable pump; aggressive vapors require PTFE or PTFE-coated wetted parts.

User Manual, Exhaust and Vacuum Pump Requirements
Process Limits
Never start the pump with liquid in the trap or tubing

Liquid present in the Cold Trap or vacuum tubing can be drawn into the pump and damage it. During shutdown, allow ice to melt without chipping, drain immediately, and flush with water before any restart. Attach the grounding clip to the solvent catch pot when draining to eliminate electrostatic spark ignition risk.

User Manual, Shutdown and Electrical Connection
Maintenance
Empty and clean the trap after every run

Empty the trap after each run, clean spills, and clean the manifold and gasket with a mild non-abrasive cleaner. After acid use, clean immediately. Check vacuum pump oil level and condition (change if cloudy, particle-laden, or discolored; level between MIN and MAX), check protective earth continuity annually, inspect hoses and gaskets, and clean the condenser with a vacuum brush.

User Manual, Recommended Maintenance Schedule
Environment
Optimal ambient is 70–75°F (21–24°C)

Refrigeration performance is adversely affected by high ambient temperature. The permitted operating range is 41° to 104°F (5° to 40°C) with maximum 80% RH decreasing linearly to 50% at 104°F. Install in a temperature-controlled room and keep flammable or hazardous solvents at least 12 inches (300 mm) away.

User Manual, Section 2 and Appendix C

Selection & Integration Q&A

When should I choose the -84°C or -105°C class instead of the standard -50°C CT freeze dryer?

Choose from the sample eutectic temperature: the collector must be 15–20°C colder. Samples at -40 to -60°C eutectic require cascade refrigeration below -75°C, so the -50°C class is insufficient; use the -84°C or -105°C class or a dry-ice/solvent trap. The -84°C class also has the largest instantaneous ice load (2.1 L vs 1.6 L).

What vacuum pump and hose interface does the CT-Series require, and can I use my existing pump?

The pump must achieve less than 0.02 mbar and its inlet fitting must accept 0.50 ID hose; recommended operating vacuum is 0.133 mbar or less. The pump plugs into a separate receptacle from the Cold Trap. Flammable solvents require an explosion-proof or solvent-suitable pump, and aggressive vapors require PTFE or PTFE-coated wetted parts.

What are the room, clearance, and power-quality limits for installing a CT freeze dryer?

Indoor use only, ambient 41–104°F (5–40°C) with optimal 70–75°F (21–24°C), max 80% RH decreasing to 50% at 104°F. Keep 3 in (7.6 cm) clear on both sides and rear because air enters right and exits left. Supply voltage must stay within ±10% of nominal on a dedicated earthed circuit.

What maintenance and consumables are required, and what is not included in the base delivery?

Empty and clean the trap after each run, clean manifold and gasket with mild non-abrasive cleaner, check pump oil (change if cloudy or discolored), check earth continuity annually, and clean the condenser with a vacuum brush. The vacuum pump, pump oil, and 0.50 ID hose are user-supplied; the manual's Appendix A lists replaceable parts such as gaskets, valves, and sensors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CT-Series Freeze Dryers Manual
Manual ·1,5 МБ ·41 pp.
View PDF