Labconco Goldfisch Fat Extractor
Labconco Goldfisch Fat Extractor — Kjeldahl & extraction apparatus.
The Labconco Goldfisch Fat Extraction Apparatus (6-unit model 35001) is a continuous reflux extractor for AOAC crude-fat determinations, reducing extraction time from 16 hours to about four hours. It operates on 115 V, 50/60 Hz at 5.2 A maximum and requires an explosion-proof hood, 3/8" NPT cold water supply, and 3/4" NPT drain. Each of the six heaters is 100 W, with panel switches controlling two heaters each.
Full specifications
Description
Construction and positioning
- The 6-unit model 35001 is a manifold-based extractor with six independent glassware sets, each comprising a sample tube, thimble, beaker, and condenser head.
- It is shipped fully assembled; operating parts (glassware, gaskets, wave washers, retainer rings) are installed after placement.
- No sibling models are listed in the verified facts; this is the sole 6-unit configuration described in the manual.
Behavior under load and in operation
- Each heater is rated 100 W; the 'HIGH' position delivers approximately 100 W per heater, while control position '5' delivers approximately 35 W per heater.
- A solvent charge of 25–35 ml typically supports 3–4 hours of refluxing.
- Pre-heating heaters for 8–10 minutes before sample introduction shortens time to reflux.
- If only one extraction is performed, the companion heater on the same panel switch should be loaded with a beaker of solvent or water to extend heater life.
- Reducing condenser water flow below the minimum causes solvent loss; no flow allows solvent to escape through pressure relief valves.
Installation, connection, and control
- Requires a dedicated 115 VAC grounded outlet (5.2 A max), a 3/8" NPT cold water supply, and a 3/4" NPT drain.
- Cooling water flow must be at least 0.75 gal/min (2.8 L/min), preferably at 55 °F (13 °C) or less.
- The apparatus must be operated inside an explosion-proof hood with proper airflow.
- Retaining rings are hand-tightened to achieve a vapor seal; do not over-tighten.
- Heater covers must be used when removing or attaching beakers to prevent solvent contact with hot heaters.
Integration & Operating Notes
The model 35001 draws up to 5.2 A at 115 V, 50/60 Hz and is supplied with a grounded 3-wire plug. Verify that the circuit can supply this current continuously and that the outlet is properly grounded to avoid shock hazard and ensure stable heater operation.
Cooling water must flow at no less than 0.75 gal/min (2.8 L/min) and preferably at 55 °F (13 °C) or lower. Insufficient flow or warm water degrades condensation, leading to solvent loss and potentially unsafe solvent vapor release through pressure relief valves.
Because extremely flammable solvents are used, the extractor must be placed inside an explosion-proof hood with adequate airflow. Site selection must account for hood space matching the unit's overall dimensions of 35.7 in × 26.9 in × 9 in.
A solvent charge of 25–35 ml is normally sufficient for 3–4 hours of refluxing. Exceeding this volume or time may not be necessary and could affect extraction efficiency or safety. Plan batch durations around approximately four hours of reflux.
Gaskets and beakers are routine maintenance items that wear out or break. Order replacements by Labconco part number: gaskets (part number not specified in digest), beakers 100 cc (part 35053), extraction thimbles 22 × 80 mm (part 35059). Keep spares to avoid downtime.
Each panel switch controls two heaters. If only one extraction is performed, load the companion heater with a beaker of solvent or water. This increases heater service life by avoiding uneven thermal cycling.
Selection & Integration Q&A
The verified facts describe only the 6-unit model 35001. No sibling models are listed, so selection versus other sizes cannot be assessed from the provided data. Choose this model when six simultaneous extractions fit your throughput and the required utilities (115 VAC, 5.2 A, water supply/drain) are available.
It requires a dedicated 115 VAC, 50/60 Hz grounded outlet (5.2 A max), a 3/8" NPT cold water supply with at least 0.75 gal/min flow at ≤55 °F, and a 3/4" NPT drain. Operation must be inside an explosion-proof hood.
The extractor is designed for batch operation with reflux times of about 4 hours per batch. It must be operated inside an explosion-proof hood. Cooling water flow must be maintained at ≥0.75 gal/min; reducing flow causes solvent loss. Pre-heat heaters for 8–10 minutes before starting samples.
The stainless steel sample tube container (35067) and tall-form beaker (35046) are not furnished as standard equipment. Routine maintenance items include gaskets and beakers (100 cc, part 35053); extraction thimbles are 22 × 80 mm (part 35059). Keep spares on hand.