Labconco Protector Airo Filtered Fume Hoods
Labconco · In Active Production

Labconco Protector Airo Filtered Fume Hoods

Labconco Protector Airo Filtered Fume Hoods — Fume hoods.

The Protector Airo is a 4 ft (121.9 cm) bench-top ductless filtered fume hood with 28.7 in interior depth, using Neutrodine filtration and a by-pass design to hold 60–100 fpm inflow. It is the 4 ft Airo modification (M2: two filter columns, four filters) within the Echo/Airo series, intended for benchtop work with moderate chemical loads where ducting is impractical.

Full specifications

Dimensions, Weight & Mechanical Design
Exterior Depth 37.7 in
Height, Lower 29.7 in
Height, Upper 35.7 in
Interior depth 28.7 in
Width 4 ft
Shipping weight mfr 1095 lb

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Heat load above 1058 W or 300 °C mandates a ducted hood

The Neutrodine filter bed cannot tolerate heat-generating equipment exceeding 1058 W or 300 °C surface temperature, nor an internal temperature rise greater than 5 °C above ambient. Exceeding these limits risks filter damage and invalid containment; specify a ducted hood instead.

User Manual, Section 3 Safety Precautions
Electrical
Provide a dedicated earthed outlet sized for 108 W total

The 4 ft hood draws 72 W fan plus 36 W light (108 W total, under 3 A). A dedicated outlet with appropriate circuit breaker and protective earthing ground must be close to the unit; 115 V wiring uses black phase and white neutral, 230 V uses brown Phase1 and blue Phase2.

User Manual, Table 2-1 and Section 4 Installation
Process Limits
Regulate service-line pressure to 40 PSI recommended

Service lines must be ¼-inch OD copper (brass for natural gas) with an easily accessible shut-off valve. Recommended operating pressure is 40 PSI; maximum allowable is 200 PSI. Use a regulator to bring supply down to the recommended value before connection.

User Manual, Section 2 Service Line Requirements
Control & I/O
Calibrate airflow sensor after filter change and annually

Airflow sensor calibration takes about 5 minutes and must be done with both side panels installed and no wiring or cords blocking the sensor inlet. After first installation or filter replacement a wrench-icon maintenance alarm appears; clear it by holding the mute button for two seconds.

User Manual, Section 4 Calibrating the Airflow Sensor
Maintenance
Replace all primary filters at 36 months or on sensor alarm

Primary Neutrodine filters must be replaced when embedded sensors detect chemicals between primary and secondary filters, or after 36 months of use. Replace all primary filters at once. This 4 ft Airo uses two columns and four filters (M2 configuration).

User Manual, Section 8 Neutrodine Filter Replacement
Mounting
Keep the matched sash weight with its own hood

The sash weight is individually matched to each hood and must not be exchanged with another unit. During installation or relocation, retain the original sash weight with its hood to preserve sash balance and airflow behavior.

User Manual, Section 4 Unpacking

Selection & Integration Q&A

When should I choose the 4 ft Airo instead of a 3 ft Airo or a 5/6/8 ft Echo?

Pick the 4 ft Airo when benchtop depth and moderate chemical load fit two filter columns (four Neutrodine filters, up to 260 CFM, 108 W). The 3 ft Airo uses one column/two filters for smaller loads; 5, 6 and 8 ft sizes exist only as Echo models, and floor-mount versions are Echo only.

What electrical and service-line provisions must be ready before the hood arrives?

A dedicated earthed outlet with circuit breaker sized for 108 W total (fan 72 W, light 36 W; under 3 A). Service lines: ¼-inch OD copper (brass for natural gas), recommended 40 PSI, maximum 200 PSI, with an accessible shut-off valve. Optional 115 V harness 9582710 or 230 V harness 9582720 for duplex outlets.

Where can the hood be placed and what duty limits apply to heat-generating work?

On a level, solid surface away from traffic, doors, windows, fans and ventilation registers, with room windows closed. Heat-generating equipment is limited to 1058 W or 300 °C surface temperature; above that, or if internal temperature rises more than 5 °C above ambient, a ducted hood is required.

What maintenance and consumables are not included in the base delivery?

Neutrodine filters (four for this 4 ft Airo), optional HEPA filter, pre-filters, airflow and chemical sensors, service fixture kits, duplex outlet kits, work surfaces, base stands and seismic supports are separate. Primary filters need replacement at 36 months or on sensor alarm; airflow sensor calibration is annual.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Protector Echo and Airo Filtered Fume Hoods Manual (2020-Present)
Manual ·5,4 МБ ·114 pp.
View PDF
PDF
Spanish-Language Protector Echo and Airo Filtered Fume Hoods Brochure
Brochure ·2,7 МБ ·20 pp.
View PDF