Labconco Protector Airo Filtered Fume Hoods
Labconco Protector Airo Filtered Fume Hoods — Fume hoods.
The Protector Airo is a 4 ft (121.9 cm) bench-top ductless filtered fume hood with 28.7 in interior depth, using Neutrodine filtration and a by-pass design to hold 60–100 fpm inflow. It is the 4 ft Airo modification (M2: two filter columns, four filters) within the Echo/Airo series, intended for benchtop work with moderate chemical loads where ducting is impractical.
Full specifications
| Exterior Depth | 37.7 in |
| Height, Lower | 29.7 in |
| Height, Upper | 35.7 in |
| Interior depth | 28.7 in |
| Width | 4 ft |
| Shipping weight mfr | 1095 lb |
Description
Construction and series positioning
- Bench-top, ductless filtered fume hood; 4 ft nominal width (121.9 cm) with 28.7 in (729 mm) interior depth and 55.7 in exterior depth.
- Two-part height: lower 29.7 in, upper 47.7 in; shipping weight 1095 lb.
- Within the Echo/Airo family, this is the 4 ft Airo bench-top variant (M2 filtration: 2 columns, 4 Neutrodine filters). The 3 ft Airo uses one column/two filters; 5, 6 and 8 ft sizes exist only as Echo models. Floor-mount versions are Echo only.
- Uses the patented Protector hood design (U.S. Patent 6,461,233) with Neutrodine universal filter technology and a by-pass feature for ductless operation.
Behavior under load and media
- Designed inflow velocity range 60–100 fpm (0.31–0.51 m/s); velocity is adjusted by sash position and verified by the airflow sensor.
- Heat-generating equipment is limited to 1058 W or 300 °C surface temperature; beyond this, or if internal temperature rises more than 5 °C above ambient, a ducted hood is required.
- Filtration is monitored by embedded sensors (molecode S, A or F) between primary and secondary filters; primary filters must be replaced when chemicals are detected or after 36 months.
Installation, control and operation
- Requires a level, solid surface away from traffic, doors, windows, fans and ventilation registers; room windows closed.
- Dedicated electrical outlet with circuit breaker and protective earthing; main power to control panel, fans and lights draws less than 3 A. For this 4 ft width, fan 72 W, light 36 W, total 108 W, airflow up to 260 CFM.
- Service lines: ¼-inch OD copper (brass for natural gas), recommended 40 PSI, maximum 200 PSI, with accessible shut-off valve.
- Command module with Smart Halo indicator and mute button; optional eGuard module provides INFO/LOGS/SETTINGS pages, face-velocity and filtration status, and alarm screens. Alarm mute silences for ten minutes only.
- Airflow sensor calibration (~5 min) required for safe operation, with side panels installed and sensor inlet unobstructed; annual calibration recommended. Certification before first use and at least annually thereafter.
Integration & Operating Notes
The Neutrodine filter bed cannot tolerate heat-generating equipment exceeding 1058 W or 300 °C surface temperature, nor an internal temperature rise greater than 5 °C above ambient. Exceeding these limits risks filter damage and invalid containment; specify a ducted hood instead.
The 4 ft hood draws 72 W fan plus 36 W light (108 W total, under 3 A). A dedicated outlet with appropriate circuit breaker and protective earthing ground must be close to the unit; 115 V wiring uses black phase and white neutral, 230 V uses brown Phase1 and blue Phase2.
Service lines must be ¼-inch OD copper (brass for natural gas) with an easily accessible shut-off valve. Recommended operating pressure is 40 PSI; maximum allowable is 200 PSI. Use a regulator to bring supply down to the recommended value before connection.
Airflow sensor calibration takes about 5 minutes and must be done with both side panels installed and no wiring or cords blocking the sensor inlet. After first installation or filter replacement a wrench-icon maintenance alarm appears; clear it by holding the mute button for two seconds.
Primary Neutrodine filters must be replaced when embedded sensors detect chemicals between primary and secondary filters, or after 36 months of use. Replace all primary filters at once. This 4 ft Airo uses two columns and four filters (M2 configuration).
The sash weight is individually matched to each hood and must not be exchanged with another unit. During installation or relocation, retain the original sash weight with its hood to preserve sash balance and airflow behavior.
Selection & Integration Q&A
Pick the 4 ft Airo when benchtop depth and moderate chemical load fit two filter columns (four Neutrodine filters, up to 260 CFM, 108 W). The 3 ft Airo uses one column/two filters for smaller loads; 5, 6 and 8 ft sizes exist only as Echo models, and floor-mount versions are Echo only.
A dedicated earthed outlet with circuit breaker sized for 108 W total (fan 72 W, light 36 W; under 3 A). Service lines: ¼-inch OD copper (brass for natural gas), recommended 40 PSI, maximum 200 PSI, with an accessible shut-off valve. Optional 115 V harness 9582710 or 230 V harness 9582720 for duplex outlets.
On a level, solid surface away from traffic, doors, windows, fans and ventilation registers, with room windows closed. Heat-generating equipment is limited to 1058 W or 300 °C surface temperature; above that, or if internal temperature rises more than 5 °C above ambient, a ducted hood is required.
Neutrodine filters (four for this 4 ft Airo), optional HEPA filter, pre-filters, airflow and chemical sensors, service fixture kits, duplex outlet kits, work surfaces, base stands and seismic supports are separate. Primary filters need replacement at 36 months or on sensor alarm; airflow sensor calibration is annual.