Labconco Protector Echo Filtered Fume Hoods
Labconco Protector Echo Filtered Fume Hoods — Fume hoods.
The Protector Echo is a 4 ft (121.9 cm) ductless filtered fume hood with 28.7 in (729 mm) interior depth, using Neutrodine filtration and a by-pass airflow design. It operates at 60–100 fpm (0.31–0.51 m/s) inflow and is configured with two filter columns and four Neutrodine filters (M2). It serves as a self-contained containment and filtration unit where ducting to an exhaust stack is not available.
Full specifications
| Exterior Depth | 37.7 in |
| Height, Lower | 29.7 in |
| Height, Upper | 35.7 in |
| Interior depth | 28.7 in |
| Width | 4 ft |
| Shipping weight mfr | 1095 lb |
Description
Construction and series positioning
- The Echo is a bench-top, ductless filtered fume hood built on the patented Protector hood design (U.S. Patent No. 6,461,233) with Neutrodine universal filter technology and a by-pass airflow feature.
- This 4 ft model (M2 configuration) uses two Neutrodine filter columns holding four filters; the 3 ft Airo sibling uses one column/two filters, while 5, 6 and 8 ft Echo models use three, four and five columns respectively.
- Exterior depth is 55.7 in; lower height 29.7 in and upper height 47.7 in. Shipping weight is 1095 lb.
Airflow and load behavior
- Face velocity is adjustable within 60–100 fpm (0.31–0.51 m/s) by sash position: raising the sash lowers velocity, lowering it raises velocity. Certification must be performed at the sash position that yields the required inflow velocity.
- The 4 ft hood moves up to 260 CFM through two filter modules; fan power is 72 W and light power 36 W, for 108 W total.
- Heat-generating equipment is limited to 1058 W or 300 °C surface temperature; beyond this, or if internal temperature rises more than 5 °C above ambient, a ducted hood is required.
Installation, control and operation
- The hood requires a dedicated outlet with circuit breaker and protective earthing ground; main power to control panel, fans and lights draws less than 3 amps.
- 115 V wiring uses black phase and white neutral; 230 V uses brown Phase 1 and blue Phase 2. Ground only the MAINS ground conductor to the protective earthing lug.
- Service lines are ¼-inch OD copper (brass for natural gas) with an accessible shut-off valve; recommended operating pressure is 40 PSI, maximum 200 PSI.
- The command module provides a Smart Halo indicator and mute button; the optional eGuard module exposes INFO, LOGS and SETTINGS pages showing face velocity, fan, filtration and maintenance states.
- Airflow sensor calibration takes about 5 minutes and must be done with both side panels installed and no obstruction at the sensor inlet.
Integration & Operating Notes
Install on a level, solid surface away from traffic patterns, doors, windows, fans, ventilation registers and other air-handling devices that could disrupt airflow. All room windows should be closed during operation, otherwise the inflow velocity at the sash may fall outside the 60–100 fpm certification range.
The 4 ft hood draws 72 W for the fan and 36 W for lights (108 W total) at less than 3 A main power. A dedicated outlet with appropriate circuit breaker must be close to the product and provide a protective earthing ground; connect only the MAINS ground conductor to the earthing lug, using black/white for 115 V or brown/blue for 230 V.
Heat-generating equipment up to 1058 W or 300 °C surface temperature is tolerated; above this a ducted hood is required. Do not operate the filtered hood if internal temperature rises more than 5 °C above ambient, as filtration performance and safety depend on staying within these limits.
Service lines must be ¼-inch outside diameter copper (brass for natural gas) with an easily accessible shut-off valve. Recommended operating pressure is 40 PSI; maximum allowable is 200 PSI. Use a regulator to bring line pressure down to the recommended value before connection.
Airflow sensor calibration is required for safe operation and takes about 5 minutes: zero the sensor with the fan off (~1 minute), then set the ~80 fpm (0.4 m/s) value with the fan on (~1 minute). Perform with both side panels installed, no wiring or cords blocking the sensor inlet, and the sash fixed 2–3 inches above working height. After first installation or filter change a maintenance alarm (wrench icon) appears; clear it by holding the mute button for two seconds.
For models with the eGuard command module, the Wi-Fi network is GFH-XXXXX-XXXX (XXXX = serial number) with password GFH-2018 and browser address http://192.168.1.1. For direct Ethernet, set the laptop to IP 192.168.0.201, subnet mask 255.255.255.0, and browse to http://192.168.0.205.
Primary Neutrodine filters must be replaced when the embedded sensor (molecode S, A or F) detects chemicals between primary and secondary filters, or after 36 months of use, whichever comes first. Replace all primary filters at once. The 4 ft M2 configuration holds four filters in two columns; order the correct count for the hood size.
Certification is required before initial use and recertification at least annually, or whenever the hood is relocated or serviced. Use ANSI/ASHRAE 110 test methodology and ANSI/AIHA/ASSP Z9.5 risk assessment if no Chemical Hygiene Plan exists. Certify at the sash position that yields the required inflow velocity within 60–100 fpm.
The sash weight is individually matched to each hood and must not be exchanged with another unit. During installation or relocation, retain the original sash weight with its hood to preserve sash balance and safe operation.
Selection & Integration Q&A
Choose the 4 ft Echo (M2) when you need two filter columns/four Neutrodine filters and up to 260 CFM at 108 W total. The 3 ft Airo uses one column/two filters and 54 W; 5–8 ft Echo models scale to three–five columns and 162–270 W for larger work zones.
Provide a dedicated outlet with circuit breaker and protective earthing ground close to the hood. Main power to control panel, fans and lights draws less than 3 A (108 W total for the 4 ft). Use black phase/white neutral for 115 V or brown Phase 1/blue Phase 2 for 230 V, grounding only the MAINS ground conductor.
Install on a level, solid surface away from traffic, doors, windows, fans and ventilation registers, with room windows closed. Heat-generating equipment is limited to 1058 W or 300 °C surface temperature; above this, or if internal temperature rises more than 5 °C above ambient, a ducted hood is required.
Plan primary filter replacement at 36 months or on sensor alarm (four filters for this M2 unit), annual recertification, and airflow sensor calibration after filter/fan service. Optional items not in base delivery include Neutrodine/HEPA filters, service fixture kits, duplex outlet kits, and 115 V/230 V harnesses 9582710/9582720.