Labconco Protector Echo Filtered Fume Hoods
Labconco · In Active Production

Labconco Protector Echo Filtered Fume Hoods

Labconco Protector Echo Filtered Fume Hoods — Fume hoods.

The Protector Echo is a 4 ft (121.9 cm) ductless filtered fume hood with 28.7 in (729 mm) interior depth, using Neutrodine filtration and a by-pass airflow design. It operates at 60–100 fpm (0.31–0.51 m/s) inflow and is configured with two filter columns and four Neutrodine filters (M2). It serves as a self-contained containment and filtration unit where ducting to an exhaust stack is not available.

Full specifications

Dimensions, Weight & Mechanical Design
Exterior Depth 37.7 in
Height, Lower 29.7 in
Height, Upper 35.7 in
Interior depth 28.7 in
Width 4 ft
Shipping weight mfr 1095 lb

Description

Construction and series positioning

Airflow and load behavior

Installation, control and operation

Integration & Operating Notes

Mounting
Locate away from air disturbances and close room windows

Install on a level, solid surface away from traffic patterns, doors, windows, fans, ventilation registers and other air-handling devices that could disrupt airflow. All room windows should be closed during operation, otherwise the inflow velocity at the sash may fall outside the 60–100 fpm certification range.

User Manual, Section 2 Location and Clearance Requirements
Electrical
Provide a dedicated earthed outlet sized for 108 W total

The 4 ft hood draws 72 W for the fan and 36 W for lights (108 W total) at less than 3 A main power. A dedicated outlet with appropriate circuit breaker must be close to the product and provide a protective earthing ground; connect only the MAINS ground conductor to the earthing lug, using black/white for 115 V or brown/blue for 230 V.

User Manual, Section 2 Electrical Requirements and Section 4 Installation
Process Limits
Do not exceed 1058 W or 300 °C heat load inside the hood

Heat-generating equipment up to 1058 W or 300 °C surface temperature is tolerated; above this a ducted hood is required. Do not operate the filtered hood if internal temperature rises more than 5 °C above ambient, as filtration performance and safety depend on staying within these limits.

User Manual, Section 3 Safety Precautions
Process Limits
Regulate service-line pressure to 40 PSI, never above 200 PSI

Service lines must be ¼-inch outside diameter copper (brass for natural gas) with an easily accessible shut-off valve. Recommended operating pressure is 40 PSI; maximum allowable is 200 PSI. Use a regulator to bring line pressure down to the recommended value before connection.

User Manual, Section 2 Service Line Requirements
Control & I/O
Recalibrate airflow sensor after filter or fan service

Airflow sensor calibration is required for safe operation and takes about 5 minutes: zero the sensor with the fan off (~1 minute), then set the ~80 fpm (0.4 m/s) value with the fan on (~1 minute). Perform with both side panels installed, no wiring or cords blocking the sensor inlet, and the sash fixed 2–3 inches above working height. After first installation or filter change a maintenance alarm (wrench icon) appears; clear it by holding the mute button for two seconds.

User Manual, Section 4 Calibrating the Airflow Sensor
Control & I/O
eGuard access uses fixed Wi-Fi and Ethernet addresses

For models with the eGuard command module, the Wi-Fi network is GFH-XXXXX-XXXX (XXXX = serial number) with password GFH-2018 and browser address http://192.168.1.1. For direct Ethernet, set the laptop to IP 192.168.0.201, subnet mask 255.255.255.0, and browse to http://192.168.0.205.

User Manual, Appendix D
Maintenance
Replace all primary filters at 36 months or on sensor alarm

Primary Neutrodine filters must be replaced when the embedded sensor (molecode S, A or F) detects chemicals between primary and secondary filters, or after 36 months of use, whichever comes first. Replace all primary filters at once. The 4 ft M2 configuration holds four filters in two columns; order the correct count for the hood size.

User Manual, Section 8 Neutrodine Filter Replacement
Maintenance
Recertify at least annually and after relocation or service

Certification is required before initial use and recertification at least annually, or whenever the hood is relocated or serviced. Use ANSI/ASHRAE 110 test methodology and ANSI/AIHA/ASSP Z9.5 risk assessment if no Chemical Hygiene Plan exists. Certify at the sash position that yields the required inflow velocity within 60–100 fpm.

User Manual, Section 4 Certification and Re-Certification
Maintenance
Keep the matched sash weight with its own hood

The sash weight is individually matched to each hood and must not be exchanged with another unit. During installation or relocation, retain the original sash weight with its hood to preserve sash balance and safe operation.

User Manual, Section 4 Unpacking

Selection & Integration Q&A

When should I choose this 4 ft Echo over the 3 ft Airo or a wider Echo model?

Choose the 4 ft Echo (M2) when you need two filter columns/four Neutrodine filters and up to 260 CFM at 108 W total. The 3 ft Airo uses one column/two filters and 54 W; 5–8 ft Echo models scale to three–five columns and 162–270 W for larger work zones.

What electrical supply and wiring must be provided for this hood?

Provide a dedicated outlet with circuit breaker and protective earthing ground close to the hood. Main power to control panel, fans and lights draws less than 3 A (108 W total for the 4 ft). Use black phase/white neutral for 115 V or brown Phase 1/blue Phase 2 for 230 V, grounding only the MAINS ground conductor.

What site conditions and heat loads are acceptable for this filtered hood?

Install on a level, solid surface away from traffic, doors, windows, fans and ventilation registers, with room windows closed. Heat-generating equipment is limited to 1058 W or 300 °C surface temperature; above this, or if internal temperature rises more than 5 °C above ambient, a ducted hood is required.

What maintenance and accessories are needed beyond the base hood?

Plan primary filter replacement at 36 months or on sensor alarm (four filters for this M2 unit), annual recertification, and airflow sensor calibration after filter/fan service. Optional items not in base delivery include Neutrodine/HEPA filters, service fixture kits, duplex outlet kits, and 115 V/230 V harnesses 9582710/9582720.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Protector Echo and Airo Filtered Fume Hoods Manual (2020-Present)
Manual ·5,4 МБ ·114 pp.
View PDF
PDF
Spanish-Language Protector Echo and Airo Filtered Fume Hoods Brochure
Brochure ·2,7 МБ ·20 pp.
View PDF