Labconco Protector Premier Laboratory Hoods
Labconco · In Active Production

Labconco Protector Premier Laboratory Hoods

Labconco Protector Premier Laboratory Hoods — Fume hoods.

The Protector Premier 8' laboratory hood is a by-pass, vertical-rising sash fume hood with a 48" interior height and 28" maximum operating sash opening. It requires 1640 CFM at 100 fpm sash open and 0.28" static pressure, distinguishing it as the widest standard model in the Premier line. It serves as the primary containment device in laboratory exhaust systems, factory-rated to ASHRAE 110-1995 AM<0.05.

Full specifications

Core Performance & Functional Specifications
Airflow At 100 Fpm Sash Open 705 CFM
Built-in Blower Airflow 725 CFM
Static pressure at 100 fpm sash open mfr 0.28 s.p.
Built-in blower maximum external static pressure mfr 0.17 S.P.
Dimensions, Weight & Mechanical Design
Hood width mfr 8 ft
Electrical, Power & Drive Specifications
Built-in blower motor HP mfr 1 HP

Description

Construction and Series Positioning

Airflow and Containment Behavior

Installation and Operation

Integration & Operating Notes

Mounting
Exhaust duct must be supported independently of the hood

The weight of the exhaust ductwork must be carried by separate supports, not by the hood superstructure. Supporting duct load on the hood can distort the cabinet, misalign the sash, and compromise airflow containment.

Instruction Manual, Chapter 3 Exhaust Connection / p.17
Process Limits
Face velocity must be certified within 60–100 fpm before use

The hood is designed for an average face velocity of 60 to 100 fpm at the sash opening. Operating outside this range risks inadequate containment or excessive turbulence; certification by a qualified contractor is required after installation and after any change to the exhaust system.

Instruction Manual, Chapter 3 Certifying / p.21; Chapter 8 / p.42
Process Limits
Natural gas requires the brass-tubed service fixture only

Natural gas must be connected only to the service fixture that is pre-plumbed with brass tubing. Standard copper supply lines are not suitable due to sulfur content deterioration. Oxygen must never be used with any standard service fixture; contact Labconco for oxygen-compatible options.

Instruction Manual, Chapter 3 Service Lines / p.20–21
Maintenance
Keep sash weight matched to its original hood

The sash weight is individually matched to the specific hood during factory assembly. Exchanging it with another unit can cause incorrect sash balance, leading to sash drift or difficulty in operation. During service or relocation, ensure the weight remains with its original hood.

Instruction Manual, Chapter 3 Sash Weight Release / p.15

Selection & Integration Q&A

When should I choose the 8' Protector Premier hood with remote blower instead of an integral blower model?

Choose the 8' remote-blower model when you need the widest standard working width and have an external exhaust system capable of 1640 CFM at 0.28" static pressure. Integral blower models are available only in 4', 5', and 6' widths and include a built-in blower (1180 CFM, 1 HP) for installations without remote exhaust.

What are the requirements for the exhaust duct and blower connected to this hood?

The exhaust connection is 12" nominal (12.75" OD) duct. The blower must deliver 1640 CFM at 100 fpm sash open against 0.28" static pressure, and maintain transport velocity of 1000–2500 fpm in the duct. Ductwork weight must be supported independently of the hood.

What maintenance and accessories are required that are not included with the base hood?

The base hood does not include service fixtures, blower, ductwork, or work surface. Routine maintenance per Chapter 6 includes sash weight check and LED lighting inspection. Optional accessories such as Guardian Airflow Monitors, sash stops, and additional service fixtures must be ordered separately and installed per Chapter 7.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Protector Laboratory Fume Hoods Brochure – XStream, Premier and XL
Brochure ·3,5 МБ ·24 pp.
View PDF
PDF
Protector Premier Laboratory Hoods User Manual
Manual ·3,2 МБ ·70 pp.
View PDF