Labconco Purifier HEPA Filtered Safety Enclosures
Labconco Purifier HEPA Filtered Safety Enclosures — Laminar flow cabinets & clean benches.
The Purifier HEPA Filtered Safety Enclosure is a Class I-style containment unit available in 2', 3', and 4' widths, using an integral motorized impeller to draw room air past the operator and push it through a HEPA filter rated at least 99.99% efficient for 0.3-micron particles. Face velocity is adjustable from 60 to 105 fpm, with 75–105 fpm recommended for biohazardous operations. Filtered exhaust may be recirculated or ducted outside via a thimble connection and remote blower.
Full specifications
| Shipping weight mfr | 621 lbs. |
| Current | 12 A |
| Energy Requirements | 150 W |
Description
Construction and Series Positioning
- The Purifier HEPA Filtered Enclosure is a filtered containment enclosure distinct from the Purifier Class I Filtered Enclosure (which adds UV) and from the Logic/Axiom biological safety cabinets in the broader Labconco line.
- Available in 2', 3', and 4' widths; part numbers span 3980200–3980423 depending on width and configuration.
- Integral motorized impeller is vibration-isolated; airflow is horizontal, guided by aerodynamic air foils, an upper sash foil, a zoned rear perforated baffle, and side-entry air foils.
- The plenum and HEPA filter are jacketed by negative pressure so leaks are recaptured and refiltered.
- Internal depth is 23 inches; the sash is 1/4-inch tempered safety glass and must be down for normal operation.
Airflow and Containment Behavior
- Face velocity is set via a speed control behind the front panel: 75–105 fpm for biohazardous operations, 60–100 fpm for weighing procedures. Containment is maximized within the biohazard range; quietest operation is at the lowest appropriate velocity.
- Exhaust volume, noise, and static pressure are tabulated for 60, 75, 80, 90, 100, and 105 fpm. At 75 fpm: 2' model 110 CFM, 49–55 dB(A), max 220 ft equivalent 6" duct, 0.19" external static pressure; 3' model 165 CFM, 52–56 dB(A), max 65 ft equivalent 6" duct, 0.12" external static pressure; 4' model 220 CFM, 54–58 dB(A), max 60 ft equivalent 6" duct, 0.19" external static pressure.
- The upper dilution air supply introduces 5–7% of the required air volume and must not be blocked.
- HEPA filters are leak-checked at the factory; a second leak check is required before use and at least annually thereafter.
- ASHRAE 110-1995 as-manufactured rating: less than 0.05 ppm at 4 lpm (AM <0.05). NSF/ANSI 49 biological containment testing at 75 fpm inflow.
Installation, Connection, and Control
- 115V models connect via supplied power cord to the IEC inlet on the rear; at 50 Hz maximum airflow is reduced by 17%. 230V models ship without a plug and require a locally appropriate plug.
- Optional thimble exhaust connections (6" for house exhaust, 5" for FilterMate Portable Exhauster) allow ducting outside; ductwork weight must be supported independently of the enclosure superstructure, and an adjustable damper or valve must be installed when connected to a house exhaust system.
- The enclosure must be placed on a level, structurally adequate base cabinet and a smooth, rigid, chemical-resistant work surface at least as wide as the enclosure, with a dished recess for primary spills. The front of the enclosure must align within 0.36" of the front of the work surface. Four mounting screws secure the enclosure; a minimum 1/8" clearance under the screw heads allows the rear and front air-foil flanges to slide into place.
- Accessory Odor Control carbon filters are gravity-installed with the gasket side down on the exhaust side of the HEPA filter after removing the upper perforated cover. Carbon filters provide no particulate protection and are for trace nuisance odors only; work-area concentrations must not exceed the IDLH and exhaust must not exceed the TWA.
Integration & Operating Notes
115V models connect via the supplied power cord to the IEC inlet on the rear. At 50 Hz, the blower's maximum airflow is reduced by 17%, so face velocity and exhaust sizing must account for this when commissioning on 50 Hz supplies. 230V models ship without a plug and require a locally appropriate plug to be fitted.
The speed control behind the front panel sets face velocity. For biohazardous operations, set 75–105 fpm; for weighing procedures, 60–100 fpm. Containment is maximized within the biohazard range, and quietest operation is at the lowest appropriate velocity. The sash must be down for normal operation; use the spring-loaded latch only for loading and cleaning.
When the enclosure is connected directly to a house exhaust system, an adjustable damper or valve must be installed to control airflow properly, especially when one house system serves multiple enclosures. Optional thimble exhaust connections are 6" for house exhaust and 5" for FilterMate Portable Exhauster. The weight of the exhaust ductwork must be supported independently of the enclosure superstructure, and duct material must match the enclosure, procedures, and chemical applications.
HEPA filters ship installed with the gasket on the downstream side and are leak-checked at Labconco, but a second leak check is required before use and at least annually thereafter. Chapter 6 of the manual covers the routine maintenance schedule, decontamination, determination of when to replace HEPA and carbon filters, HEPA filter leak testing, setting inflow face velocity, calibrating and operating the airflow monitor, calculating carbon filter life, initial certification and re-certification, and replacement of fluorescent/UV lamps, motorized impeller, and speed control.
Selection & Integration Q&A
Choose the Purifier HEPA Filtered Enclosure for particulate containment without UV or the Class II personnel/product protection of Logic/Axiom cabinets. The Class I Filtered Enclosure adds UV for surface decontamination. Logic/Axiom cabinets are Class II biological safety cabinets with different containment principles and should be selected when operator protection from bioaerosols is required.
Use a thimble exhaust connection: 6" for house exhaust or 5" for FilterMate. When connected to house exhaust, install an adjustable damper or valve to control airflow. Support ductwork weight independently of the enclosure. Duct material must match the enclosure, procedures, and chemicals. For 2' model at 75 fpm, max 220 ft equivalent 6" duct and 0.19" external static pressure; 3' model max 65 ft and 0.12"; 4' model max 60 ft and 0.19".
HEPA filter leak check before use and at least annually. Routine maintenance per Chapter 6 includes decontamination, filter replacement determination, face velocity setting, airflow monitor calibration, carbon filter life calculation, certification/re-certification, and replacement of lamps, impeller, and speed control. Accessories not included: HEPA and carbon filters (part numbers by width), thimble exhaust connections (3924400 for 2', 3924401 for 3', 3924402 for 4'), Guardian Airflow Monitor, work surfaces, utility shelves, remote blowers, and adjustable dampers. 230V models require a locally appropriate plug.