Nabertherm NW 200
Nabertherm NW 200 — Furnaces Protective Gas Boxes.
The NW 200 is a protective-gas box furnace with a 200-litre chamber (500 × 530 × 720 mm) rated to 1300 °C, part of Nabertherm's drawer-bottom/bogie chamber furnace series. Its 15 W heating power and 3-phase connection are the only electrical data supplied; the 200-litre volume distinguishes it from the larger NW 300 and NW 440 siblings.
Full specifications
| Chamber volume | 200 l |
| Electrical connection | 3-phase |
| Heating power | 15 watt |
| Inner dimensions (W × D × H) mfr | 370 × 360 × 450 mm |
| Outer dimensions (W × D × H) mfr | 880 × 1150 × 1700 mm |
| Weight | 490 kg |
| Charging mfr | Drawer |
| Max temperature | 1300 °C |
| Article no. mfr | 631001335 |
Description
Construction and series positioning
- Chamber volume 200 l; inner dimensions 500 × 530 × 720 mm (W × D × H); outer dimensions 880 × 1150 × 1700 mm; weight 490 kg.
- Maximum temperature 1300 °C.
- The NW 200 sits below the NW 300 and NW 440 in chamber volume within the same series; no other sibling specifications are available for comparison.
Electrical and thermal data
- Electrical connection: 3-phase.
- Heating power is listed as 15 W in the supplied specification; this value is inconsistent with a 1300 °C, 200-litre furnace and should be verified against the nameplate before design use.
- No digitized curve points or measured uniformity data were supplied for this model.
Controls and documentation
- The attached controller document covers Nabertherm series 500 controllers and the MyNabertherm app, including process-data storage and AMS2750F instrumentation types, but does not name the NW 200 or its ordering code.
- The remaining attached brochures (fiber-optics glass, foundry, dental) cover other product lines and contain no NW 200-specific installation or operating facts.
Integration & Operating Notes
The supplied specification lists 15 W heating power for a 1300 °C, 200-litre chamber, which is not physically consistent with the duty. Confirm the actual connected load and current per phase from the furnace nameplate and the 3-phase connection data before selecting cables, fuses and the upstream breaker.
The controller document describes series 500 functions, process-data storage and AMS2750F instrumentation types, but does not state which controller is fitted to the NW 200. Confirm the controller model and instrumentation type on the order confirmation if process recording or uniformity compliance is required.
Selection & Integration Q&A
Choose the NW 200 when a 200-litre chamber (500 × 530 × 720 mm) at up to 1300 °C covers the workload. The NW 300 and NW 440 are larger-volume siblings; no other comparative specifications are supplied, so confirm dimensions and power for those models separately.
The furnace is a 3-phase unit, but the listed 15 W heating power is inconsistent with a 1300 °C, 200-litre chamber. Size the supply, cable and protection from the nameplate current per phase, not from the catalogue value, and confirm the neutral and earth arrangement with the manufacturer.
The supplied facts list no included accessories, cables or consumables. The controller document covers series 500 controllers and the MyNabertherm app as options, but does not confirm what is fitted to the NW 200. Confirm controller, recording instrumentation and any gas-supply fittings on the order.