Shimadzu GC 2014
Shimadzu ·Made in Japan · In Active Production

Shimadzu GC 2014

Shimadzu GC 2014 — Chromatographs with Flow rate 0…100 milliliter/minute.

The Shimadzu GC-2014 is a dual-packed injection gas chromatograph with a 15.8 L column oven programmable in 1 °C steps up to 400 °C and a dual flow-rate differential detector control system. It supports FID, TCD, ECD, FPD and FTD detectors with dynamic ranges from 10^3 to 10^7, and up to 20 cooling program steps. It is intended for routine packed-column GC analysis in laboratory method development and quality control.

Flow rate
0…100 milliliter/minute

Full specifications

Core Performance & Functional Specifications
Flow rate 0 ~ 100 mL/min
Dimensions 250 (W) × 360 (H) × 175 (D) mm
Column oven capacity mfr 15.8 L
Column oven temperature accuracy mfr Set value (K) ± 1%
Column oven total program time mfr 9999.99 minutes max.
Column oven linear heating rate (up to 400°C, 100V/120V) mfr 7 °C/min (up to 400°C, 100V/120V)
Column oven linear heating rate (up to 400°C, 200V/230V) mfr 15 °C/min (up to 400°C, 200V/230V)
Detector types (for capillary/packed columns) mfr FID, TCD, ECD, FPD, FTD (for capillary/packed columns)
FID dynamic range mfr 10^7
TCD dynamic range mfr 10^5
ECD dynamic range (63Ni 370 MBq radioactive source) mfr 10^4 (63Ni 370 MBq radioactive source)
FPD dynamic range mfr P: 10^4
FTD dynamic range mfr N: 10^3
Injection unit types mfr Dual packed
Programmable flow steps mfr 7 steps
Programmed rate setting range mfr -400 ~ 400
Pressure setting range mfr 0 ~ 970 kPa
Interfaces, Connections & Control
Detector flow control system mfr Dual flow rate differential system
Materials, Media & Process Compatibility
Column oven temperature range (using liquid CO2) mfr (Ambient + 10°C) ~ 400 °C (using liquid CO2)
Operating Conditions, Environment & Protection
Column oven cooling rate (at 25°C ambient temperature) mfr 300°C ~ 50°C in 6 min max. (at 25°C ambient temperature)

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection and control

Integration & Operating Notes

Electrical
Supply voltage determines maximum heating rate

On 100/120 V supplies the oven linear heating rate is limited to 7 °C/min up to 400 °C; on 200/230 V supplies it reaches 15 °C/min. Method transfer between sites with different mains voltages will change run times and must be re-validated.

Factory Specification Sheet
Cooling & Media
Liquid CO2 cooling sets the lower oven limit

When liquid CO2 is used for cooling, the column oven range becomes (ambient + 10 °C) to 400 °C. Methods requiring sub-ambient starts must account for this offset and for the 6 min cool-down from 300 °C to 50 °C at 25 °C ambient.

Factory Specification Sheet
Process Limits
FPD detector temperature must not exceed 350 °C

FID, TCD, ECD and FTD detectors may be set up to 400 °C, but the FPD is limited to 350 °C. Oven or detector methods written for other detectors must be checked before running an FPD to avoid exceeding its limit.

Factory Specification Sheet
Control & I/O
Flow programming is limited to seven steps

Programmed flow rate settings range from -400 to 400 with a pressure setting range of 0 to 970 kPa, across a maximum of 7 programmable flow steps. Complex multi-ramp pressure programs beyond seven steps cannot be entered on this model.

Factory Specification Sheet
Maintenance
ECD source requires radioactive-material handling procedures

The ECD uses a 63Ni 370 MBq radioactive source. Installation, leak testing and eventual disposal must follow the radioactive-material procedures in the instruction manual; the document digest does not list these procedures and the full manual must be consulted.

Instruction Manual c184-e014

Selection & Integration Q&A

When should I choose the GC-2014 dual-packed configuration instead of a capillary-oriented variant?

Choose this model when the method uses packed columns and dual packed injection; its injection unit type is dual packed. Capillary capability is documented, but packed-column work is the configuration this model is specified around.

What detector and flow-control requirements must adjacent system parts meet?

The detector flow control is a dual flow-rate differential system with a 0–100 mL/min flow range and pressure setting up to 970 kPa. Detector choice sets the temperature ceiling: 400 °C for FID/TCD/ECD/FTD, 350 °C for FPD.

What maintenance and consumables are not included in base delivery?

The document digest for c184-e014 lists no accessories, cables or consumables in base delivery, and no service intervals. The full instruction manual must be consulted for maintenance schedules and the ECD radioactive-source handling procedure.

Dimensional Drawings & Technical Documentation

Dimensional drawings

GC 2014 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
c184-e014
Datasheet ·2,8 МБ ·14 pp.
View PDF