Shimadzu GC 2014
Shimadzu GC 2014 — Chromatographs with Flow rate 0…100 milliliter/minute.
The Shimadzu GC-2014 is a dual-packed injection gas chromatograph with a 15.8 L column oven programmable in 1 °C steps up to 400 °C and a dual flow-rate differential detector control system. It supports FID, TCD, ECD, FPD and FTD detectors with dynamic ranges from 10^3 to 10^7, and up to 20 cooling program steps. It is intended for routine packed-column GC analysis in laboratory method development and quality control.
Full specifications
| Flow rate | 0 ~ 100 mL/min |
| Dimensions | 250 (W) × 360 (H) × 175 (D) mm |
| Column oven capacity mfr | 15.8 L |
| Column oven temperature accuracy mfr | Set value (K) ± 1% |
| Column oven total program time mfr | 9999.99 minutes max. |
| Column oven linear heating rate (up to 400°C, 100V/120V) mfr | 7 °C/min (up to 400°C, 100V/120V) |
| Column oven linear heating rate (up to 400°C, 200V/230V) mfr | 15 °C/min (up to 400°C, 200V/230V) |
| Detector types (for capillary/packed columns) mfr | FID, TCD, ECD, FPD, FTD (for capillary/packed columns) |
| FID dynamic range mfr | 10^7 |
| TCD dynamic range mfr | 10^5 |
| ECD dynamic range (63Ni 370 MBq radioactive source) mfr | 10^4 (63Ni 370 MBq radioactive source) |
| FPD dynamic range mfr | P: 10^4 |
| FTD dynamic range mfr | N: 10^3 |
| Injection unit types mfr | Dual packed |
| Programmable flow steps mfr | 7 steps |
| Programmed rate setting range mfr | -400 ~ 400 |
| Pressure setting range mfr | 0 ~ 970 kPa |
| Detector flow control system mfr | Dual flow rate differential system |
| Column oven temperature range (using liquid CO2) mfr | (Ambient + 10°C) ~ 400 °C (using liquid CO2) |
| Column oven cooling rate (at 25°C ambient temperature) mfr | 300°C ~ 50°C in 6 min max. (at 25°C ambient temperature) |
Description
Construction and positioning
- The GC-2014 is a dual-packed injection GC with a 15.8 L column oven and a dual flow-rate differential detector control system. It is the packed-column-oriented modification within the GC-2014 family; capillary capability is documented but the injection unit type is dual packed.
- Detector options include FID, TCD, ECD (63Ni 370 MBq), FPD and FTD, with maximum detector temperatures of 400 °C (FID, TCD, ECD, FTD) and 350 °C (FPD).
- The oven accepts 1 °C temperature setting steps and up to 20 cooling program steps, with a total program time of up to 9999.99 min.
Behavior under load and in different modes
- Linear heating rate is voltage-dependent: 7 °C/min up to 400 °C on 100/120 V supplies, and 15 °C/min up to 400 °C on 200/230 V supplies. Cooling from 300 °C to 50 °C takes 6 min at 25 °C ambient.
- Oven temperature accuracy is set value (K) ± 1%, with a temperature variation coefficient of 0.01 °C/°C. With liquid CO2 cooling the oven range is (ambient + 10 °C) to 400 °C.
- Detector dynamic ranges differ by type: FID 10^7, TCD 10^5, ECD 10^4, FPD 10^4 (P), FTD 10^3 (N). Flow programming provides 7 steps with a programmed rate setting range of -400 to 400 and a pressure setting range of 0 to 970 kPa.
Installation, connection and control
- The unit measures 250 (W) × 360 (H) × 175 (D) mm. Flow rate range is 0 to 100 mL/min.
- The document digest for c184-e014 does not provide detailed installation, connection or control facts; these must be taken from the full instruction manual.
Integration & Operating Notes
On 100/120 V supplies the oven linear heating rate is limited to 7 °C/min up to 400 °C; on 200/230 V supplies it reaches 15 °C/min. Method transfer between sites with different mains voltages will change run times and must be re-validated.
When liquid CO2 is used for cooling, the column oven range becomes (ambient + 10 °C) to 400 °C. Methods requiring sub-ambient starts must account for this offset and for the 6 min cool-down from 300 °C to 50 °C at 25 °C ambient.
FID, TCD, ECD and FTD detectors may be set up to 400 °C, but the FPD is limited to 350 °C. Oven or detector methods written for other detectors must be checked before running an FPD to avoid exceeding its limit.
Programmed flow rate settings range from -400 to 400 with a pressure setting range of 0 to 970 kPa, across a maximum of 7 programmable flow steps. Complex multi-ramp pressure programs beyond seven steps cannot be entered on this model.
The ECD uses a 63Ni 370 MBq radioactive source. Installation, leak testing and eventual disposal must follow the radioactive-material procedures in the instruction manual; the document digest does not list these procedures and the full manual must be consulted.
Selection & Integration Q&A
Choose this model when the method uses packed columns and dual packed injection; its injection unit type is dual packed. Capillary capability is documented, but packed-column work is the configuration this model is specified around.
The detector flow control is a dual flow-rate differential system with a 0–100 mL/min flow range and pressure setting up to 970 kPa. Detector choice sets the temperature ceiling: 400 °C for FID/TCD/ECD/FTD, 350 °C for FPD.
The document digest for c184-e014 lists no accessories, cables or consumables in base delivery, and no service intervals. The full instruction manual must be consulted for maintenance schedules and the ECD radioactive-source handling procedure.