Shimadzu Inspexio 7000
Shimadzu Inspexio 7000 — Spectrometers.
The inspeXio 7000 is a microfocus X-ray CT system with a 225 kV/1000 µA tube and a 16-inch flat-panel detector, housed in a shielded cabinet of 2170 × 1350 × 1858 mm and 3100 kg. It supports normal, half, offset, FS, 2DCT, and CBCT scan modes, with SOD and SID axes enabling geometric magnification. The system is used for non-destructive internal inspection and dimensional analysis of industrial components.
Full specifications
| Detector types | Flat panel detector |
| Input resolution mfr | 14 megapixels (inspeXio 7000) |
| Processing speed improvement (HPCinspeXio ver. 3.0) mfr | 50 times faster (HPCinspeXio ver. 3.0) |
| Sample rotation (CT scanning) mfr | 360 degrees (CT scanning) |
| MPR image FOV (MPR image) mfr | ø207.6 mm (MPR image) |
| Cross-sectional image FOV (GFRP) (GFRP cross-sectional image) mfr | ø20 mm (GFRP cross-sectional image) |
| CR image FOV (CR image) mfr | 10 × 20 mm (CR image) |
| Oblique image FOV (Oblique image) mfr | ø30 mm (Oblique image) |
| MPR image FOV (void analysis) (MPR image) mfr | ø42 mm (MPR image) |
| Module type (CORE Series) mfr | Option module for X-ray CT system (CORE Series) |
| Compatible CT systems (Microfocus X-ray CT System) mfr | inspeXio 7000, SMX-225CT FPD HR Plus (Microfocus X-ray CT System) |
| Module variant (dual-energy scanning, NEW) mfr | CORE XDEF (dual-energy scanning, NEW) |
| Module variant (de-noising technology, NEW) mfr | CORE Xtract (de-noising technology, NEW) |
| Module variant (ultra-fast reconstruction) mfr | CORE Boost (ultra-fast reconstruction) |
| Scanning technique (CORE XDEF) mfr | dual-energy scanning (CORE XDEF) |
| Image processing (CORE Xtract) mfr | proprietary algorithm eliminating only image noi (CORE Xtract) |
| Reconstruction (CORE Boost) mfr | high-speed parallel processing (CORE Boost) |
| Image reconstruction function (CORE Boost) mfr | ROI image reconstruction (CORE Boost) |
| Artifact reduction (CORE XDEF) mfr | metal artifacts reduced (CORE XDEF) |
| Contrast optimization (CORE XDEF) mfr | per material (CORE XDEF) |
| X-Ray Detector mfr | Flat panel detector |
| X-Ray Detector Size mfr | 16 inch |
| X-Ray Detector Shades of Gray mfr | 16 |
| Max. Input Resolution (for Offset Scan) mfr | 14,000,000 pixels |
| Two-Dimensional CT mfr | 4096 |
| Cone-Beam CT mfr | 4096 |
| Version mfr | 3.0 |
| For 1024 × 1024 pixel CT images mfr | 10 sec |
| Positioning via an Exterior Camera mfr | Yes |
| Recommend Scanning mfr | Yes |
| SOD Axis *1 mfr | 890 mm |
| SID Axis *2 mfr | 3 levels |
| CT-Z Axis mfr | 300 mm |
| Scan Modes mfr | 3 |
| CT Data Acquisition Time mfr | 10 sec |
| Main Unit mfr | 200/220 |
| External Leakage Dose mfr | 0.5 |
| Option mfr | Operation Box. Exterior Camera Retraction Unit |
| Layout dimension (Layout and dimensional drawings) mfr | 500 mm (Layout and dimensional drawings) |
| Shield Box Size and Mass mfr | 2170 |
| Cross-sectional image FOV (lithium-ion battery) (Lithium-ion rechargeable battery) mfr | ø18.4 mm (Lithium-ion rechargeable battery) |
| Cross-sectional image FOV (lithium polymer battery) (Rectangular lithium polymer battery) mfr | ø15 mm (Rectangular lithium polymer battery) |
| Rated Power mfr | 135 W |
| Max. Tube Voltage mfr | 225 kV |
| Max. Tube Current mfr | 1000 |
| Control Computer *6 *7 mfr | 100 V |
| Ground mfr | 100 ohm |
| 3D CT Scan Region Display Function mfr | Yes |
| CE, KC mfr | Yes |
Description
Construction and positioning
- The inspeXio 7000 integrates a microfocus X-ray source (max. 225 kV, 1000 µA) and a 16-inch flat-panel detector with 16-bit grayscale (65,536 shades) inside a radiation-shielded cabinet. - The shield box measures W2170 × D1350 × H1858 mm and weighs approximately 3100 kg, requiring a stable floor and adequate access. ### Scan modes and imaging behavior
- Six scan modes are available: normal, half, offset, FS, 2DCT, and CBCT. Offset scan yields a maximum input resolution of approximately 14,000,000 pixels; 2DCT and CBCT produce 4096 × 4096 images. - The SOD axis has a fixed 890 mm travel, while the SID axis switches between three levels (600, 800, 1200 mm), allowing geometric magnification adjustment. - CT data acquisition time is user-selectable from 10 seconds to 60 minutes; a 1024 × 1024 pixel CT image is reconstructed in 10 seconds or less after acquisition. - The system includes a 3D CT scan region display function and recommended scanning feature to aid setup. ### Installation and control
- The main unit requires AC 200/220/230/240 V (tap switching) ±10%, 50/60 Hz, 2 kVA; the control computer requires AC 100–240 V ±10%, 50/60 Hz, 1.5 kVA. - A Type-D ground with 100 Ω max. ground resistance is mandatory. - External leakage dose is 0.5 µSv/h or less, and the system carries CE and KC markings. - Positioning can be performed via an exterior camera; an optional operation box and exterior camera retraction unit are available.
Integration & Operating Notes
The main unit must be connected to a 200/220/230/240 V AC supply with the correct tap selected; incorrect tap setting or voltage outside ±10% can cause malfunction or damage. The supply must be rated for 2 kVA.
The shielded cabinet weighs approximately 3100 kg and measures 2170 × 1350 × 1858 mm. Ensure the floor and access route can accommodate this load and size; inadequate support may affect alignment and safety.
A dedicated Type-D ground with resistance not exceeding 100 Ω is mandatory for safe operation and to meet leakage dose limits. Poor grounding can compromise safety and image quality.
An exterior camera is available for positioning the sample; an optional retraction unit and operation box are offered. These are not included in the base delivery and must be specified separately.
Selection & Integration Q&A
The main unit needs AC 200–240 V ±10%, 50/60 Hz, 2 kVA with tap switching; the control computer needs AC 100–240 V ±10%, 50/60 Hz, 1.5 kVA. A Type-D ground with ≤100 Ω resistance is required.