Shimadzu Lightnirs
Shimadzu ·Made in Japan · In Active Production

Shimadzu Lightnirs

Shimadzu Lightnirs — Life science instruments.

The Shimadzu Lightnirs is a compact near-infrared tissue oximeter using 3-wavelength semiconductor lasers and an avalanche photodiode to measure Oxy-Hb, Deoxy-Hb, and Total-Hb. Its 253×222×68 mm, 1.6 kg housing accepts 15 V from an AC adapter or lithium-ion battery and provides 3 digital and 10 analog inputs. It serves as a portable monitoring unit in life-science and clinical research settings.

Full specifications

Core Performance & Functional Specifications
Light source 3-wavelength near-infrared semiconductor laser
Wavelength 780, 805, 830 nm
Measurement channels (Holder Type A, forehead) (Holder Type A, one unit) mfr 2×8 (22ch) ch (Holder Type A, one unit)
Measurement channels (Holder Type A, dorsolateral prefrontal area) (Holder Type A, one unit) mfr 2×4 (10ch)×2 ch (Holder Type A, one unit)
Measurement channels (Holder Type A, language area) (Holder Type A, one unit) mfr 2×4 (10ch)×2 ch (Holder Type A, one unit)
Measurement channels (Holder Type A, dorsolateral prefrontal cortex + prefrontal area) (Holder Type A, two-unit connecti mfr 3×10 (47ch) ch (Holder Type A, two-unit connecting system)
Measurement channels (Holder Type A, supplementary motor area) (Holder Type A, two-unit connecting system (both)) mfr 3×5 (24ch)×2 ch (Holder Type A, two-unit connecting system (both))
Measurement channels (Holder Type B, primary motor area) (Holder Type B) mfr 2×8 (22ch) ch (Holder Type B)
Measurement channels (Holder Type B, visual area and frontal pole) (Holder Type B) mfr 2×4 (10ch)×2 ch (Holder Type B)
Measurement channels (Holder Type B, frontal pole + visual cortex + somatosensory) (Holder Type B) mfr 2×4+4×6 (48ch) ch (Holder Type B)
Measurement channels (Holder Type B, motor area) (Holder Type B) mfr 4×8 (52ch) ch (Holder Type B)
Optical fiber pairs (selectable optical fiber) mfr 8 pairs pairs (selectable optical fiber)
Number of subjects (study data) mfr 8 (study data)
Measured region (study data) mfr Frontal pole 3 × 3 (12-channel) (study data)
Rest period (block design) mfr 10 sec (block design)
Oral tasting period (block design) mfr 20 sec (block design)
Rest period after swallowing (block design) mfr 30 sec (block design)
Jelly sample amount (4 types of jelly) mfr about 5 g (4 types of jelly)
Scalable system configuration (system configuration) mfr 4 up to 40 sets (system configuration)
L-type optic fiber cable expansion (L-type optic fiber cable) mfr +12 sets (L-type optic fiber cable)

Description

Integration & Operating Notes

Electrical
Use only a 15 V source rated for the instrument's input

The Lightnirs accepts 15 V input from either the supplied AC adapter or a lithium-ion battery. Connecting a supply outside this rating risks damage to the laser and detector electronics; verify adapter output and battery voltage before use.

Factory Specification Sheet
Environment
Keep ambient at or below 30 °C and 45–85 % RH

Operation is specified at 30 °C and 45–85 % relative humidity with no condensation. Exceeding these conditions can degrade optical measurement stability and may invalidate results; provide climate control or avoid humid, condensing locations.

Factory Specification Sheet
Control & I/O
Plan cabling for 3 digital and 10 analog input channels

The external input interface exposes 3 digital and 10 analog channels. Integrators must provide compatible signal levels and connectors, and should document channel mapping for synchronization with external acquisition or trigger systems.

Factory Specification Sheet

Selection & Integration Q&A

What power and signal interfaces must adjacent equipment provide?

Provide a 15 V DC source via AC adapter or lithium-ion battery, and route 3 digital plus 10 analog channels to the external input connector. Confirm signal levels and connector types against the instrument before wiring.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c297-e103
Datasheet ·2,0 МБ ·8 pp.
View PDF
PDF
c297-e089
Datasheet ·5,6 МБ ·24 pp.
View PDF
PDF
c297-e153
Datasheet ·328,4 КБ ·2 pp.
View PDF