The Shimadzu Nexera Asaprep is a preparative-scale HPLC system within the Nexera series, configured for automated purification of target compounds from complex matrices. It integrates a high-flow pump, fraction collector, and UV detector into a single platform, enabling unattended isolation of milligram-to-gram quantities. Its primary purpose is to serve as a bridge between analytical method development and downstream characterization or bioassay.
Direct drop-in replacements
Full specifications
Description
Construction and Series Positioning
The Nexera Asaprep is a dedicated preparative HPLC system, distinct from analytical Nexera models by its higher flow-rate capacity and integrated fraction collection. Unlike analytical siblings that prioritize sub-2 µm particle separations at low flow, the Asaprep is optimized for larger-bore columns and higher sample loads. The system architecture includes a binary or quaternary pump, autosampler with enlarged injection volume, UV-Vis detector with preparative flow cell, and a fraction collector—all under a single control platform. ### Behavior Under Load and Media
Operating with typical preparative solvents (e.g., water/acetonitrile or methanol gradients), the system maintains stable flow and pressure up to its specified maximum. The fraction collector triggers on peak slope or threshold, allowing collection of targeted peaks even in overloaded conditions. The system tolerates higher sample concentrations and viscous solutions than analytical configurations, though exact limits depend on the installed pump head and tubing. ### Installation, Control, and Operation
Installation requires a bench space with adequate ventilation and solvent waste containment. The system is controlled via Shimadzu LabSolutions software, which manages method transfer from analytical to preparative scale. Connections use standard 1/16" or 1/8" fittings, and the fraction collector accepts various tube racks. Operation specifics, including leak checks and column equilibration, follow the standard Nexera series manual.
Integration & Operating Notes
The system with filled solvent reservoirs and a loaded fraction collector can exceed 100 kg; ensure the bench is level and rated for at least 150 kg to prevent vibration-induced baseline noise and collector misalignment.
The pump and detector each draw up to 6 A at 100–240 V AC; sharing a circuit with other high-draw equipment may cause voltage fluctuations that affect detector baseline stability. Use a line conditioner if the supply is noisy.
The LabSolutions method should include a pre-run check that verifies the collector is in the correct rack position and the waste line is clear; otherwise, the pump will not start, preventing accidental spillage of purified fractions.
Rapid composition changes at preparative flow rates can cause pressure spikes and poor peak shape; ramp gradients at ≤5% B per minute when flow exceeds 20 mL/min to stay within the pump's dynamic response.
At typical preparative flows, seal wear accelerates; track cumulative volume via the software counter and replace seals at 1000 L or if pressure ripple exceeds 0.5 MPa, whichever comes first, to prevent leaks and baseline drift.
Selection & Integration Q&A
Choose Asaprep when target compound mass per run exceeds 1 mg and you need automated fraction collection; analytical Nexera models lack the high-flow pump and collector capacity for preparative loads.
The Asaprep's outlet uses a standard 1/8" OD tube; ensure the collector's inlet fitting matches and that the MS source can tolerate preparative flow rates (often requiring a split). Verify control via analog I/O or LabSolutions driver.
Operate at 4–35 °C and 20–80% RH non-condensing; continuous duty is supported, but allow 30 min warm-up for detector stability. Avoid direct sunlight and vibration sources.
Base delivery excludes solvent reservoirs, waste containers, column, fraction collection racks, and spare seals. Plan to order these separately; also schedule quarterly PM kits (seals, check valves) and annual detector lamp replacement.