Shimadzu Nexera Asaprep

Shimadzu Nexera Asaprep — Chromatographs.

The Shimadzu Nexera Asaprep is a preparative-scale HPLC system within the Nexera series, configured for automated purification of target compounds from complex matrices. It integrates a high-flow pump, fraction collector, and UV detector into a single platform, enabling unattended isolation of milligram-to-gram quantities. Its primary purpose is to serve as a bridge between analytical method development and downstream characterization or bioassay.

Direct drop-in replacements

0%
match
⚖️ Compare
0%
match
⚖️ Compare

Full specifications

Description

Construction and Series Positioning

The Nexera Asaprep is a dedicated preparative HPLC system, distinct from analytical Nexera models by its higher flow-rate capacity and integrated fraction collection. Unlike analytical siblings that prioritize sub-2 µm particle separations at low flow, the Asaprep is optimized for larger-bore columns and higher sample loads. The system architecture includes a binary or quaternary pump, autosampler with enlarged injection volume, UV-Vis detector with preparative flow cell, and a fraction collector—all under a single control platform. ### Behavior Under Load and Media

Operating with typical preparative solvents (e.g., water/acetonitrile or methanol gradients), the system maintains stable flow and pressure up to its specified maximum. The fraction collector triggers on peak slope or threshold, allowing collection of targeted peaks even in overloaded conditions. The system tolerates higher sample concentrations and viscous solutions than analytical configurations, though exact limits depend on the installed pump head and tubing. ### Installation, Control, and Operation

Installation requires a bench space with adequate ventilation and solvent waste containment. The system is controlled via Shimadzu LabSolutions software, which manages method transfer from analytical to preparative scale. Connections use standard 1/16" or 1/8" fittings, and the fraction collector accepts various tube racks. Operation specifics, including leak checks and column equilibration, follow the standard Nexera series manual.

Integration & Operating Notes

Mounting
Bench must support full solvent load and collector weight

The system with filled solvent reservoirs and a loaded fraction collector can exceed 100 kg; ensure the bench is level and rated for at least 150 kg to prevent vibration-induced baseline noise and collector misalignment.

Nexera Series Pre-Installation Guide
Electrical
Requires dedicated 15 A circuit for pump and detector

The pump and detector each draw up to 6 A at 100–240 V AC; sharing a circuit with other high-draw equipment may cause voltage fluctuations that affect detector baseline stability. Use a line conditioner if the supply is noisy.

Nexera Series Site Preparation Manual
Control & I/O
Fraction collector must be enabled before pump start

The LabSolutions method should include a pre-run check that verifies the collector is in the correct rack position and the waste line is clear; otherwise, the pump will not start, preventing accidental spillage of purified fractions.

LabSolutions Preparative Method Guide
Process Limits
Avoid solvent gradients exceeding 5% per minute at high flow

Rapid composition changes at preparative flow rates can cause pressure spikes and poor peak shape; ramp gradients at ≤5% B per minute when flow exceeds 20 mL/min to stay within the pump's dynamic response.

Nexera Asaprep Application Note: Method Transfer
Maintenance
Replace pump seals every 1000 L of delivered solvent

At typical preparative flows, seal wear accelerates; track cumulative volume via the software counter and replace seals at 1000 L or if pressure ripple exceeds 0.5 MPa, whichever comes first, to prevent leaks and baseline drift.

Nexera Series Maintenance Manual, Section 5.1

Selection & Integration Q&A

When should I choose the Nexera Asaprep over an analytical Nexera model for purification?

Choose Asaprep when target compound mass per run exceeds 1 mg and you need automated fraction collection; analytical Nexera models lack the high-flow pump and collector capacity for preparative loads.

What are the requirements for connecting the Asaprep to an existing fraction collector or MS detector?

The Asaprep's outlet uses a standard 1/8" OD tube; ensure the collector's inlet fitting matches and that the MS source can tolerate preparative flow rates (often requiring a split). Verify control via analog I/O or LabSolutions driver.

What environmental conditions and duty-cycle limits apply to the Asaprep?

Operate at 4–35 °C and 20–80% RH non-condensing; continuous duty is supported, but allow 30 min warm-up for detector stability. Avoid direct sunlight and vibration sources.

What maintenance items and accessories are not included in the base delivery?

Base delivery excludes solvent reservoirs, waste containers, column, fraction collection racks, and spare seals. Plan to order these separately; also schedule quarterly PM kits (seals, check valves) and annual detector lamp replacement.