Shimadzu Nexera UC
Shimadzu Nexera UC — Chromatographs.
The Shimadzu Nexera UC is a supercritical fluid chromatography (SFC) system configured for unified chromatography, combining SFC and LC separation modes in a single instrument. Its primary purpose is to serve as a versatile chromatographic tool for method development and routine analysis in pharmaceutical, food, and environmental testing.
Direct drop-in replacements
Full specifications
| Number of Capsicum varieties analyzed mfr | 16 |
| Number of compounds extracted and identified mfr | 41 |
| Analysis time mfr | less than 20 min |
| Sample amount mfr | 1 g |
| Adsorbent powder amount mfr | 1 g |
| Gradient program mfr | 0-2 min, 0% B; 2-17 min 80% B |
| Pressure capacity (LC-30ADSF CO2 pump) mfr | 66 MPa (LC-30ADSF CO2 pump) |
| Pressure capacity (LC-40D modifier/makeup pump) mfr | 44 MPa (LC-40D modifier/makeup pump) |
| Pressure capacity (LC-40D XR modifier/makeup pump) mfr | 70 MPa (LC-40D XR modifier/makeup pump) |
| Pressure capacity (LC-40D XS modifier/makeup pump) mfr | 105 MPa (LC-40D XS modifier/makeup pump) |
| Pressure capacity (LC-40D X3 modifier/makeup pump) mfr | 130 MPa (LC-40D X3 modifier/makeup pump) |
| Pressure capacity (SIL-40 autosampler) mfr | 44 MPa (SIL-40 autosampler) |
| Pressure capacity (SIL-40C autosampler) mfr | 44 MPa (SIL-40C autosampler) |
| Pressure capacity (SIL-40 XR autosampler) mfr | 80 MPa (SIL-40 XR autosampler) |
| Pressure capacity (SIL-40C XR autosampler) mfr | 80 MPa (SIL-40C XR autosampler) |
| Pressure capacity (SIL-40C XS autosampler) mfr | 105 MPa (SIL-40C XS autosampler) |
| Pressure capacity (SIL-40C X3 autosampler) mfr | 130 MPa (SIL-40C X3 autosampler) |
| Rinsing function (SIL-40 autosampler) mfr | Optional (SIL-40 autosampler) |
| Rinsing function (SIL-40C autosampler) mfr | Optional (SIL-40C autosampler) |
| Rinsing function (SIL-40 XR autosampler) mfr | Optional (SIL-40 XR autosampler) |
| Rinsing function (SIL-40C XR autosampler) mfr | Optional (SIL-40C XR autosampler) |
| Rinsing function (SIL-40C XS autosampler) mfr | Optional (SIL-40C XS autosampler) |
| Rinsing function (SIL-40C X3 autosampler) mfr | Standard (SIL-40C X3 autosampler) |
| Maximum sample capacity (with rack changer) mfr | 48 samples (with rack changer) |
| Maximum microplate capacity (with up to three optional plate changers) mfr | 44 microplates (with up to three optional plate changers) |
Description
Construction and Architecture
The Nexera UC is a modular ultra-high-performance supercritical fluid chromatography (UHPSFC) system. It integrates a CO2 delivery pump, a modifier pump, an autosampler, a column oven, a backpressure regulator, and a photodiode array (PDA) or mass spectrometry (MS) detector. The system is designed to operate with compressed carbon dioxide as the primary mobile phase, with optional organic modifiers. This configuration allows seamless switching between SFC and LC modes, enabling unified chromatography workflows. ### Behavior Under Load and Modes
The system operates under high pressure, with CO2 as the main mobile phase. The backpressure regulator maintains system pressure to keep CO2 in a supercritical state. The modular design allows for gradient elution with modifiers, and the autosampler supports high-throughput injection sequences. The column oven provides precise temperature control to ensure reproducible retention times. The system is compatible with a range of column chemistries, from sub-2 µm particles to monolithic columns, and can be coupled with various detectors. ### Installation, Connection, and Control
The Nexera UC requires a supply of high-purity CO2 (≥99.9%) and appropriate exhaust ventilation due to CO2 release. It is controlled via Shimadzu LabSolutions software, which provides integrated control of pumps, autosampler, oven, and detectors. The system requires a stable power supply and a network connection for data transfer. Installation should be performed by trained Shimadzu service engineers to ensure proper configuration and safety compliance.
Integration & Operating Notes
The system requires a continuous supply of high-purity CO2 (≥99.9%) at sufficient pressure and flow. Impurities or pressure fluctuations can cause baseline noise, retention time shifts, or system shutdown. A two-stage regulator and filters are recommended to ensure stable delivery.
The backpressure regulator and exhaust lines release CO2 into the laboratory environment. The installation site must have proper ventilation to prevent CO2 accumulation. A dedicated exhaust system or fume hood is recommended, especially in small rooms.
The system includes safety interlocks that prevent injection or pump operation if the backpressure regulator is not functioning or if CO2 pressure is insufficient. The control software will display an error and halt the run to avoid damage to the column or detector.
The CO2 pump seals and inline filters should be replaced every 6 months or after 1000 hours of operation, whichever comes first. Failure to do so can lead to leaks, pressure instability, and reduced column life. Maintenance logs should be kept for compliance.
Selection & Integration Q&A
Choose Nexera UC when you need faster separations, reduced organic solvent consumption, and compatibility with both SFC and LC modes. It excels in chiral separations and lipophilic compounds. For simple reversed-phase LC methods, a conventional HPLC may be more cost-effective.
The Nexera UC can be coupled to MS via a dedicated interface. Ensure the MS can handle high CO2 flow and has a compatible ion source. The backpressure regulator must be positioned after the MS or use a split flow to avoid pressure fluctuations affecting MS performance.
The base system does not include CO2 cylinders, regulators, filters, or exhaust tubing. Columns, solvents, and sample vials are also not included. A preventive maintenance kit (seals, filters) and a CO2 purity certificate are recommended for initial setup.