Shimadzu NOA 7100
Shimadzu ·Made in Japan · In Active Production

Shimadzu NOA 7100

Shimadzu NOA 7100 — Gas analyzers.

The Shimadzu NOA 7100 is a transportable, all-in-one NOx/O2 gas analyzer combining atmospheric-pressure chemiluminescence NOx detection with limiting-current zirconia O2 measurement. Built-in pump, filter, and electric cooler enable direct sampling without external pretreatment. It serves continuous emissions monitoring and process gas analysis where portability and multi-range flexibility are required.

Full specifications

Core Performance & Functional Specifications
Measurement components mfr NOx, O2 (NOA-7100)
Measurement components (NOA-7100 variant) mfr NOx, NO, NO2 (*1) (NOA-7100 variant)
Measurement range mfr NOx: 0-25/50/100/250/500/1000/2500/4000 (NOA-7100)
Measurement range (NOA-7100 variant) mfr NOx, NO, NO2: 250/500/1000/2500/4000 ppm ppm (NOA-7100 variant)
Gas sample amount mfr Approx. 2 L/min (NOA-7100)
Gas sample amount (NOA-7100 variant) mfr 100 mL/min (NOA-7100 variant)
External drain separator mfr Fitted (NOA-7100)
External drain separator (NOA-7100 variant) mfr Not fitted (NOA-7100 variant)
Small flow rate set mfr Not fitted (NOA-7100)
Small flow rate set (NOA-7100 variant) mfr Fitted (NOA-7100 variant)
Communication set mfr Not fitted (NOA-7100)
Communication set (NOA-7100 variant) mfr Fitted (NOA-7100 variant)
NOx-NO switching solenoid valve set mfr Not fitted (NOA-7100)
NOx-NO switching solenoid valve set (NOA-7100 variant) mfr Fitted (NOA-7100 variant)
Coexisting gas CO (O2 measurement) mfr Less than 1 vol% (O2 measurement)
Coexisting gas CO2 (NOx, NO, NO2 measurement) mfr 30 vol% or less vol% (NOx, NO, NO2 measurement)
Coexisting gas CO2 (O2 measurement) mfr Less than 30 vol% (O2 measurement)
Coexisting gas Hydrocarbon (O2 measurement) mfr Less than 1 vol% (O2 measurement)
Coexisting gas N2O (O2 measurement) mfr Must not be contained (O2 measurement)
Coexisting gas SO3 (NOx, NO, NO2 measurement) mfr 50 ppm or less ppm (NOx, NO, NO2 measurement)
Coexisting gas SO3 (O2 measurement) mfr Must not be contained (O2 measurement)
Coexisting gas Other flammable gases (NOx, NO, NO2 measurement) mfr Must not be contained (NOx, NO, NO2 measurement)

Description

Construction and positioning

Integration & Operating Notes

Mounting
Avoid direct sunlight and radiant heat during installation

The permissible ambient temperature range of 5–40°C is valid only when the analyzer is shielded from direct sunlight and radiant heat sources; exposure can cause internal overheating and measurement drift, invalidating the specified accuracy.

Factory Specification Sheet, Permissible Ambient Temperature
Electrical
Select correct voltage series and account for inrush current

Power requirements offer 100 V series (410 VA) or 200 V series (700 VA steady 200 VA). Ensure the supply circuit can handle the higher apparent power during startup; undersized wiring may cause voltage drops affecting analyzer stability.

Factory Specification Sheet, Power Requirements
Process Limits
Sample gas must be within pressure and dust limits at inlet

At the NOA-7100 inlet port, sample pressure must remain between -980 and +980 Pa and dust concentration ≤0.1 g/Nm³. Exceeding these limits can clog the built-in filter or damage the pump, leading to measurement errors or downtime.

Factory Specification Sheet, Sample Gas Conditions
Control & I/O
Analog output channels share a common ground

The 0–1 VDC transmission outputs are insulated as a group but individual channels are not isolated from each other. When connecting to a data logger or PLC, ensure that all channel returns are referenced to the same potential to avoid ground loops and signal errors.

Factory Specification Sheet, Transmission Output

Selection & Integration Q&A

When should I choose the NOA-7100 over other NOx/O2 analyzers in the Shimadzu series?

Choose the NOA-7100 when you need a transportable, all-in-one unit with built-in pump, filter, and cooler for field or temporary monitoring. For fixed installations with external conditioning, a rack-mount model may be more suitable.

What are the requirements for connecting the analog outputs to a data acquisition system?

The NOA-7100 provides 0–1 VDC, 3-channel outputs with a common ground (individual channels not isolated). Ensure the receiving device has a high-impedance input and is referenced to the same ground to prevent signal errors.

What maintenance items and accessories are not included with the base unit?

Calibration gases, external filters, and replacement parts are not included. Regular zero/span checks are required to meet drift specifications. USB flash drives for data export and LAN cables are also user-supplied.