Shimadzu Up Series
Shimadzu Up Series — Laboratory balances.
The Shimadzu Up Series is a top-loading laboratory balance with a 220 g capacity and 0.001 g readability, featuring a 110 mm round pan and external calibration. It delivers stable weighing with approximately 1 s response time and ±3 ppm/°C temperature stability, making it suitable for routine laboratory weighing tasks.
Full specifications
| Capacity | 220 g (UP223X / UP223Y) UP423X / UP423Y: 420 g · UP623X / UP623Y: 620 g · UP823X / UP823Y: 820 g · UP1023X / UP1023Y: 1020 g · UP2202X / UP2202Y: 2200 g · UP4202X / UP4202Y / UP4201X / UP4201Y: 4200 g · UP6202X / UP6202Y: 6200 g · UP422X / UP422Y: 420 g · UP822X / UP822Y: 820 g · UP8201X / UP8201Y: 8200 g |
| Linearity | ± 0.002 g (UP223X / UP223Y / UP423X / UP423Y / UP623X / UP623Y) UP823X / UP823Y / UP1023X / UP1023Y: ± 0.003 g |
| Resolution | 0.001 g (small pan model) |
| Weighing range mfr | 100 to 220 g (UP223X / UP223Y) |
| Weighing range mfr | 100 to 420 g (UP423X / UP423Y) |
| Weighing range mfr | 100 to 620 g (UP623X / UP623Y) |
| Weighing range mfr | 400 to 820 g (UP823X / UP823Y) |
| Weighing range mfr | 500 to 1020 g (UP1023X / UP1023Y) |
| Weighing range mfr | 1000 to 2200 g (UP2202X / UP2202Y) |
| Weighing range mfr | 1000 to 4200 g (UP4202X/Y, UP4201X/Y) |
| Weighing range mfr | 1000 to 6200 g (UP6202X / UP6202Y) |
| Weighing range mfr | 100 to 420 g (UP422X / UP422Y) |
| Weighing range mfr | 100 to 820 g (UP822X / UP822Y) |
| Minimum display value mfr | 0.01 g (UP2202X/Y, UP4202X/Y, UP6202X/Y, UP422X/Y, UP822X/Y) |
| Minimum display value mfr | 0.1 g (UP4201X/Y, UP8201X/Y) |
Description
Construction and Positioning
- Single-range top-loading balance with a 110 mm diameter weighing pan and 3.4 kg housing. - External span calibration requires a 100–220 g weight, indicating manual calibration rather than internal motorized calibration. ### Performance Characteristics
- Capacity: 220 g; readability: 0.001 g; repeatability: 0.001 g; linearity: 0.002 g. - Response time is approximately 1 second, supporting rapid weighing cycles. - Sensitivity stability is ±3 ppm/°C over 10–30 °C, ensuring minimal drift in temperature-controlled laboratories. - The balance is intended for benchtop use in a stable environment; external calibration weight must be supplied separately.
Integration & Operating Notes
Sensitivity stability is specified as ±3 ppm/°C only within the 10–30 °C range. Operating outside this range may increase drift and reduce weighing accuracy; ensure the balance is placed away from drafts, heat sources, and direct sunlight.
Selection & Integration Q&A
Operate within 10–30 °C to maintain ±3 ppm/°C sensitivity stability. Avoid vibrations, drafts, and rapid temperature changes. No duty-cycle limits are specified; continuous operation is typical for laboratory balances.