The Shimadzu UV-2600i/2700i Plus is a double-beam UV-Vis spectrophotometer with a sealed optical bench and integrated PC control. It covers a 185–900 nm range with high photometric accuracy, distinguishing it from single-beam or extended-range variants. It is used for quantitative absorbance and transmittance measurements in pharmaceutical, chemical, and environmental laboratories.

Full specifications

Description

Construction and positioning

Integration & Operating Notes

Mounting
Install on a level, vibration-free bench

The optical bench is sensitive to vibration and tilt; install on a rigid, level surface away from centrifuges, pumps, or other vibration sources. Exceeding ±1° tilt may misalign optics and degrade baseline stability.

Instruction Manual, Section 2.1 Installation Site
Electrical
Use a dedicated grounded outlet

Connect to a properly grounded outlet with voltage within 100–240 V AC ±10%. Avoid sharing the circuit with high-power equipment; voltage fluctuations beyond ±10% can cause lamp instability and measurement errors.

Instruction Manual, Section 2.2 Power Connection
Process Limits
Nitrogen purge required below 200 nm

For measurements below 200 nm, purge the sample compartment with nitrogen (99.99% purity) at 0.02–0.05 MPa. Insufficient purge causes absorption by oxygen, leading to inaccurate results.

Instruction Manual, Section 3.3 Nitrogen Purge
Maintenance
Replace lamps after 2,000 hours or on intensity drop

The deuterium and tungsten lamps have a typical life of 2,000 hours. Replace when the lamp intensity monitor shows a significant drop or when baseline noise increases. Always power off and unplug before replacement.

Instruction Manual, Section 5.1 Lamp Replacement
Control & I/O
Ensure PC communication is stable before starting a run

The integrated PC communicates with the spectrophotometer via USB. If the connection is lost during a measurement, the run aborts and data may be lost. Use a ferrite core on the USB cable and avoid hot-plugging.

Instruction Manual, Section 4.2 PC Connection

Selection & Integration Q&A

When should I choose the UV-2600i/2700i Plus over a single-beam or extended-range model?

Choose this double-beam model for high-stability measurements and low stray light (≤0.02% at 220 nm). For routine work above 220 nm, a single-beam model may suffice; for wavelengths beyond 900 nm, select an extended-range variant.

What are the power and PC requirements for integrating this spectrophotometer into my lab?

It requires a grounded 100–240 V AC, 50/60 Hz supply and a PC with USB port. The integrated PC runs UVProbe; ensure the outlet is not shared with noisy equipment and use a ferrite core on the USB cable.

What environmental conditions and duty-cycle limits apply to this model?

Operate at 15–35 °C and 20–80% RH (non-condensing), away from vibration and dust. Continuous operation is supported, but allow 2 hours warm-up for baseline stability; avoid rapid temperature fluctuations.

What maintenance is required and what consumables are not included in the base delivery?

Replace lamps after ~2,000 hours or when intensity drops. Base delivery excludes nitrogen regulator, tubing, and replacement lamps; order these separately. Regular cleaning of the sample compartment is recommended.