Shimadzu Xseeker 8000
Shimadzu Xseeker 8000 — Spectrometers.
The Xseeker 8000 is a flat-panel-detector X-ray CT system with a 160 kV / 1.2 mA generator, delivering 100 µm spatial resolution at a 550 mm source-to-rotation distance. It accommodates samples up to φ150 mm × H320 mm and 10 kg, with a maximum CT-scan region of φ85 mm × H70 mm at SRD 475 mm. It is intended for non-destructive internal inspection and dimensional analysis within analytical laboratory workflows.
Full specifications
Core Performance & Functional Specifications
| Maximum input resolution mfr | 5.6 million pixels (XSeeker 8000) |
| Input resolution (offset scan mode) (XSeeker 8000, offset scan mode) mfr | 5.4 million pixels (XSeeker 8000, offset scan mode) |
| Scanning field of view diameter (XSeeker 8000, offset scan mode) mfr | 95 mm (XSeeker 8000, offset scan mode) |
| Maximum sample diameter mfr | φ300 mm (XSeeker 8000) |
| Scanning time (XSeeker 8000, minimum) mfr | 12 s (XSeeker 8000, minimum) |
| Data acquisition time mfr | 10 s (XSeeker 8000) |
Dimensions, Weight & Mechanical Design
| Height | 893 mm (XSeeker 8000) |
| Width | 526 mm (XSeeker 8000) |
| Detector pixel pitch (XSeeker 8000, flat panel detector) mfr | 50 μm (XSeeker 8000, flat panel detector) |
| Depth mfr | 650 mm (XSeeker 8000) |
| Weight mfr | 290 kg (XSeeker 8000) |
Electrical, Power & Drive Specifications
| X-ray tube voltage mfr | 160 kV (XSeeker 8000) |
Description
Construction and positioning
- The Xseeker 8000 integrates a 160 kV microfocus X-ray generator with a flat panel detector in a single cabinet, weighing 290 kg. - The detector provides approximately 5.6 million pixels, enabling high-resolution CT reconstruction. ### Behavior under load and modes
- Spatial resolution is specified as 100 µm at SRD = 550 mm; at SRD = 475 mm the maximum CT-scan region is φ85 mm × H70 mm. - CT data acquisition can be performed in either 12 seconds or 5 minutes, offering a trade-off between scan speed and data quality. - Maximum sample size is φ150 mm × H320 mm at SRD = 550 mm, with a weight limit of 10 kg including the sample setting jig. ### Installation, connection, and operation
- The main unit and control computer each require a single-phase 100–240 V AC ±10%, 50/60 Hz supply at 1.0 kVA. - A D-type grounding with resistance ≤100 Ω is mandatory. - Operating environment must be maintained at 10–30 °C and 45–80% RH without condensation. - External leakage dose is specified as ≤1 µSv/h, confirming adequate shielding for laboratory use.