The Xslicer SMX 10101020 is a microfocus X-ray inspection system with a 90 kV / 250 µA (10 W) source and a 1.4-megapixel flat-panel detector, achieving 5 µm theoretical spatial resolution on the JIMA RT RC-05 chart. It accommodates samples up to 350 × 450 mm, 100 mm high, and 5 kg, with a 60° tilt stroke for oblique fluoroscopy. This model is intended for non-destructive internal inspection of small-to-medium components where sub-10 µm feature visibility is required.

Full specifications

Core Performance & Functional Specifications
Detector pixel count mfr Approximately 1.4 million pixels (Xslicer SMX-1010)
Detector pixel count mfr Approximately 2.8 million pixels (Xslicer SMX-1020)
Detector area size (SMX-1000 Plus) mfr 62 mm × 57 mm or more mm (SMX-1000 Plus)
Detector area size mfr 114 mm × 62 mm or more mm (Xslicer SMX-1010/1020)
Panoramic image size (Simple) (Equivalent to 2K (full HD)) mfr Approx. 2 million pixels (Equivalent to 2K (full HD))
Panoramic image size (Normal) (Equivalent to 4K) mfr Approx. 8 million pixels (Equivalent to 4K)
Panoramic image size (Fine) (Equivalent to 8K) mfr Approx. 32 million pixels (Equivalent to 8K)
Panoramic imaging time (Simple) mfr 115 sec (SMX-1010)
Panoramic imaging time (Simple) mfr 100 sec (SMX-1020)
Panoramic imaging time (Normal) mfr 135 sec (SMX-1010)
Panoramic imaging time (Normal) mfr 120 sec (SMX-1020)
Panoramic imaging time (Fine) mfr 590 sec (SMX-1010)
Panoramic imaging time (Fine) mfr 495 sec (SMX-1020)
Inspection start time after sample replacement mfr 5 seconds
Exterior image acquisition time (SMX-1000 Plus) mfr 00:22 mm:ss (SMX-1000 Plus)
Exterior image acquisition time mfr 00:04 mm:ss (Xslicer SMX-1020)
Movement to start position time (SMX-1000 Plus) mfr 00:12 mm:ss (SMX-1000 Plus)
Movement to start position time mfr 00:04 mm:ss (Xslicer SMX-1020)
25-sample step feed time (SMX-1000 Plus) mfr 03:28 mm:ss (SMX-1000 Plus)
25-sample step feed time mfr 02:17 mm:ss (Xslicer SMX-1020)
CT calibration time reduction (Automated calibration) mfr 80 % (Automated calibration)
Maximum sample size (Continuous rotation) mfr MAX × 125 mm (Continuous rotation)
Electrical, Power & Drive Specifications
X-ray generator type mfr 90 kV microfocus kV (SMX-1010/1020)

Description

Construction and positioning

Integration & Operating Notes

Mounting
Sample weight must not exceed 5 kg

The manipulator and stage are rated for a maximum sample mass of 5 kg. Exceeding this can cause stage deflection, positioning errors, or mechanical damage, and may invalidate the 5 µm resolution specification.

Factory Specification Sheet
Process Limits
Sample height limited to 100 mm

The cabinet geometry allows a maximum sample height of 100 mm. Taller parts may collide with the X-ray tube or detector during tilt or magnification changes, risking damage to the system and the sample.

Factory Specification Sheet
Electrical
Requires 0.5 kVA single-phase supply

The system draws 0.5 kVA from a 100–240 VAC ±10%, 50/60 Hz supply. Ensure the circuit can accommodate the inrush current and is dedicated to avoid interference with other sensitive instruments.

Factory Specification Sheet
Cooling & Media
Ambient must stay within 10–30 °C and 40–80% RH

Operation outside these conditions can affect detector performance and X-ray output stability. Condensation must be avoided; if the system is moved from a cold environment, allow it to acclimatize before powering up.

Factory Specification Sheet
Maintenance
No user-serviceable parts inside; annual PM recommended

The X-ray source and detector are sealed assemblies. Only qualified service personnel should perform maintenance. Schedule annual preventive maintenance to verify dose leakage (≤1 µSv/h) and resolution performance.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the SMX 10101020 over other Xslicer series models?

Choose this model when you need 5 µm theoretical resolution and a 90 kV source for samples up to 5 kg and 100 mm high. For larger or denser samples, a higher-kV variant in the series would be required; for sub-micron resolution, a higher-magnification model is needed.

What are the power and environmental requirements for integrating this system into my lab?

It requires a single-phase 100–240 VAC ±10%, 50/60 Hz supply at 0.5 kVA, and an environment maintained at 10–30 °C and 40–80% RH without condensation. Ensure the bench or floor can support 700 kg and that the circuit is dedicated to avoid electrical noise.

Can I inspect a sample that is 120 mm tall?

No. The maximum sample height is 100 mm. Exceeding this may cause collision with the X-ray tube or detector during tilt or magnification, potentially damaging the system and the sample.

What maintenance and accessories are required beyond the base delivery?

The base delivery includes the cabinet, X-ray source, detector, and stage. No user-serviceable parts are inside; annual preventive maintenance by qualified personnel is recommended. A calibration micro chart (e.g., JIMA RT RC-05) is needed for resolution verification and is not included.

Dimensional Drawings & Technical Documentation

Dimensional drawings

XSlicer SMX-1010/1020 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
c251-e039
Datasheet ·11,1 МБ ·16 pp.
View PDF