Shimadzu Xslicer SMX 6010
Shimadzu ·Made in Japan · In Active Production

Shimadzu Xslicer SMX 6010

Shimadzu Xslicer SMX 6010 — Spectrometers.

The Xslicer SMX-6010 is a microfocus X-ray inspection system with a flat panel detector and a 5-axis stage, achieving 1 µm spatial resolution. It accommodates samples up to 470 × 420 × 100 mm and 5 kg, with fluoroscopic and CT scan ranges of 350 × 350 mm. This model is suited for non-destructive internal inspection of small-to-medium components requiring fine feature resolution.

Full specifications

Core Performance & Functional Specifications
Detector types Flat panel detector
Resolution (SMX-6010, JIMA chart analysis) mfr 1 μm (SMX-6010, JIMA chart analysis)
External X-ray leakage mfr Maximum 1 μSv/h (SMX-6010)
Stage axes mfr Five-axis (X, Y, Z, detector tilt and rotation) (SMX-6010)
CT imaging type mfr Planar CT (PCT) standard; Vertical CT (VCT) opti (SMX-6010)
Spatial Resolution mfr 1
Maximum sample size mfr 470
Fluoroscopic Inspection Stroke mfr 460 mm
CT Scan Range mfr 350 mm
Laminographic Angle (CT Scan Angle) mfr 45
Fluoroscopy FOV (on Carbon Plate) mfr 0.75 mm
CT FOV (on Carbon Plate) mfr 3 to 30 mm
External Leakage Dose mfr 1
Dimensions, Weight & Mechanical Design
Maximum sample weight mfr 200 g (SMX-6010)
Weight mfr 2,450 kg
Electrical, Power & Drive Specifications
X-ray tube type mfr 160-kV open tube kV (SMX-6010)
X-Ray Output mfr 160 kV
Power Requirements mfr 200 V
Operating Conditions, Environment & Protection
Operating Environment Conditions mfr 15 to 30

Description

Construction and Positioning

The SMX-6010 is a planar X-ray inspection system built around a microfocus X-ray tube (160 kV, 100 µA, 16 W) and a flat panel detector. The 5-axis stage provides X, Y, Z translation plus ±180° rotation and 60° tilt, enabling both fluoroscopic and computed tomography (CT) modes. With 1 µm spatial resolution (JIMA chart), it targets applications where fine feature detection is critical. ### Behavior Under Load and in Different Modes

In fluoroscopic mode, the field of view on a carbon plate ranges from 0.75 mm × 1.3 mm to 20 mm × 38 mm, allowing magnification by positioning the sample close to the source. CT scan range is 350 mm × 350 mm, with a laminographic angle of 45° or 60°. The CT field of view depends on the angle: 3–30 mm at 45° and 3–14 mm at 60°, trading off between larger volume and higher resolution. Maximum sample size is 470 × 420 × 100 mm and 5 kg, which defines the load limit for the stage. ### Installation, Connection, and Operation

Power requirements are 200 V AC ±10%, 1.5 kVA, with ground resistance ≤100 Ω. The system must operate in an environment of 15–30 °C and 35–80% relative humidity without condensation. External leakage dose is ≤1 µSv/h, ensuring safe operation with proper shielding.

Integration & Operating Notes

Electrical
Requires 200 V AC supply with dedicated grounding

The system draws 1.5 kVA at 200 V AC ±10% and requires a ground resistance of 100 Ω or less. Ensure the supply circuit can handle the inrush current and is properly grounded to avoid damage and ensure safety.

Factory Specification Sheet
Environment
Operate within 15–30 °C and 35–80% RH non-condensing

Ambient conditions outside this range may affect X-ray tube stability and detector performance. Condensation must be avoided to prevent electrical hazards and component damage.

Factory Specification Sheet
Mounting
Sample weight must not exceed 5 kg

The 5-axis stage is rated for a maximum sample weight of 5 kg. Exceeding this limit can cause stage deformation, positioning inaccuracy, or motor overload, potentially damaging the system.

Factory Specification Sheet
Process Limits
CT field of view depends on laminographic angle

At 45° laminographic angle, CT FOV is 3–30 mm; at 60°, it is 3–14 mm. Choose the angle based on the required volume versus resolution trade-off.

Factory Specification Sheet

Selection & Integration Q&A

What are the power and grounding requirements for integrating this system?

The system requires a 200 V AC ±10% supply with 1.5 kVA capacity and a ground resistance of 100 Ω or less. Ensure the electrical installation meets these specifications to avoid operational issues.

What are the environmental limits for operating the SMX-6010?

Operate the system in ambient temperatures of 15–30 °C and relative humidity of 35–80% without condensation. These conditions ensure stable X-ray output and detector performance.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c251-e040
Datasheet ·4,7 МБ ·16 pp.
View PDF