Waters 2998 PDA (Photodiode Array) Detector
Waters ·Made in USA · In Active Production

Waters 2998 PDA (Photodiode Array) Detector

Waters 2998 PDA (Photodiode Array) Detector — Chromatography detectors.

The Waters 2998 PDA Detector is a photodiode array detector with a 512-element array and 1.2 nm digital resolution, covering 190 nm upward via a deuterium arc lamp. Its patented TaperSlit flow cell (10 mm path, 8.4 µL) supports up to 1000 psi, distinguishing it for high-resolution spectral acquisition in HPLC systems.

Full specifications

Core Performance & Functional Specifications
Detector type Photodiode Array (PDA) Detector
Wavelength range 190 nm
Wavelength accuracy mfr ±1 nm (via patented Erbium filter)
Wavelength repeatability mfr ±0.1 nm
Bandwidth mfr 1.2 nm
Photodiodes mfr 512
Digital resolution mfr 1.2 nm/pixel
Linearity range mfr ≤5% at 2.0 AU, propylparaben, 257 nm, dry analyt
Drift mfr ≤1.0 x 10 -3 AU/hour, 254 nm, 2 points/s, 1.0 s,
Sampling rate mfr Up to 80 points/s
Light source mfr Deuterium arc lamp
Flow cell design mfr Patented TaperSlit
Flow cell volume mfr 8.4 μL (analytical flow cell)
Pressure limit mfr 1000 psi (analytical flow cell)
Wetted materials mfr Alliance HPLC/ACQUITY Arc: 316 stainless-steel,
Dimensions, Weight & Mechanical Design
Path length mfr 10 mm (analytical flow cell)
Operating Conditions, Environment & Protection
Noise ≤10 x 10 -6 AU peak to peak, 254 nm, 2 points/s, 1.0 s, 30 s segments, bandwidth 3.6 nm (3-pixel bunch), dry analytical flow cell
Standards, Safety & General Data
Light source warranty mfr 2000 hours or one year (whichever comes first)

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Flow cell pressure must not exceed 1000 psi

Exceeding 1000 psi may rupture the TaperSlit flow cell, causing leaks and invalid results. Ensure system backpressure is monitored and kept below this limit during method development and operation.

Factory Specification Sheet
Maintenance
Deuterium lamp requires replacement after 2000 hours or one year

The light source warranty is limited to 2000 hours or one year, whichever comes first. Plan lamp replacement accordingly to maintain performance and avoid downtime; lamp degradation affects noise and drift specifications.

Factory Specification Sheet
Process Limits
Wetted materials vary with HPLC system; verify compatibility

For Alliance HPLC/ACQUITY Arc, wetted materials include 316 stainless steel, fused silica, PEEK, and fluoropolymer. For ACQUITY Arc Bio, they include fluoropolymer, fused silica, MP35N, PEEK, and titanium. Ensure mobile phase and sample compatibility with these materials to prevent corrosion or adsorption.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 2998 PDA Detector over other detectors in the Waters series?

Choose the 2998 PDA for full-spectrum UV/Vis detection with 512 photodiodes and 1.2 nm resolution, ideal for method development and peak purity. For single-wavelength monitoring, a UV/Vis detector may suffice; for mass confirmation, consider an MS detector.

What are the flow cell pressure and material compatibility requirements for integration with my HPLC system?

The analytical flow cell is rated to 1000 psi and uses wetted materials matched to your system: 316 stainless steel, PEEK, etc., for Alliance/ACQUITY Arc; or MP35N, titanium, etc., for ACQUITY Arc Bio. Ensure mobile phases are compatible to avoid corrosion.

Dimensional Drawings & Technical Documentation

Technical documentation