Waters 3767 Sample Manager (Injector/Collector)
Waters ·Made in USA · In Active Production

Waters 3767 Sample Manager (Injector/Collector)

Waters 3767 Sample Manager (Injector/Collector) — Laboratory automation & equipment.

The Waters 3767 Sample Manager is a dual-mode autosampler for analytical and preparative LC, supporting injection volumes from 20 µL to 30 mL via interchangeable fixed loops. It handles up to 15 microplates or 544 test tubes, with <0.75% RSD injection reproducibility and <0.05% carryover. Its 6000 psi high-pressure side and 100 psi collection side enable high-throughput sample introduction and fraction collection in a single unit.

Full specifications

Core Performance & Functional Specifications
Sample capacity (Injection) mfr Up to 15 microplates (up to 384-wells); Up to 72
Syringe delivery accuracy mfr >99.0% at full stroke with 1000-μL syringe
Injection reproducibility mfr <0.75 RSD; 20-μL injection volume, 20-μL fixed l
Carryover mfr <0.05%
Injection volume (Analytical) mfr 20 μL standard (fixed loop), (0.150-in valve pas
Injection volume (Preparative) mfr 1 mL standard (fixed loop), (0.024-in valve pass
System pressure ratings mfr 6000 psi high pressure side, including injector;
Wash pump mfr >30 mL/min flow through standard tubing and need
Available sample loops mfr 20 uL (standard analytical); 1.0 mL (standard pr
Supported syringes mfr 1.0 mL (standard); 2.5 mL, 5.0 mL, 12.5 mL
Injection cycle time mfr <30 s based on standard wash cycle
Custom rack calibration mfr Control software
Communications mfr MassLynx v4.2 or higher; Ethernet to PC/workstat
Dimensions without hood mfr Width: 959 mm (37.75 in); Depth: 629 mm (24.75 i
Dimensions with hood mfr Width: 959 mm (37.75 in); Depth: 629 mm (24.75 i
Operating temperature mfr 15 to 40 °C (59 to 104 °F)
Fume hood (optional) mfr 4-in. duct with >1.25 CFM, minimum face velocity
Dimensions, Weight & Mechanical Design
Weight 86.2 kg

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Collection side and wash system limited to 100 psi

The high-pressure side (injector) is rated to 6000 psi, but the collection side and syringe/needle wash are limited to 100 psi. Exceeding 100 psi on the collection side can damage the valve or tubing. Use appropriate back-pressure regulation when collecting fractions.

Factory Specification Sheet
Control & I/O
TTL outputs must be compatible with peripheral device logic

The rear-panel TTL (open collector) outputs are used to trigger fraction collectors or other peripherals. Open-collector outputs require pull-up resistors or compatible input circuitry on the peripheral. Verify voltage and current ratings to avoid missed triggers or damage.

Factory Specification Sheet
Environment
Fume hood duct must provide >1.25 CFM and 10 ft/min face velocity

If the optional fume hood is used, connect a 4-in. duct capable of >1.25 CFM and maintain a minimum face velocity of 10 ft/min. Inadequate exhaust may allow solvent vapors to escape into the laboratory, posing health and safety risks.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 3767 Sample Manager instead of a dedicated analytical autosampler?

Choose the 3767 when you need both analytical injection (20 µL loop) and preparative fraction collection (up to 30 mL loops) in one instrument. A dedicated analytical autosampler lacks the 100 psi collection side and large-loop capability.

What are the requirements for connecting the 3767 to a mass spectrometer and a fraction collector?

The 3767 connects to a PC via Ethernet running MassLynx v4.2 or higher. TTL outputs on the rear panel trigger peripherals like fraction collectors. Ensure the MS can accept the 6000 psi high-pressure side and that the collector's input is compatible with open-collector signals.

What maintenance items and accessories are not included in the base delivery?

The base unit includes a 20 µL analytical loop and 1.0 mL syringe. Additional loops (1, 2, 5, 10, 30 mL), syringes (2.5, 5, 12.5 mL), the fume hood, and custom racks are not included. Wash solvents and replacement tubing are consumables. Regular maintenance includes checking wash solvent levels and replacing syringes/loops per the user manual.

Dimensional Drawings & Technical Documentation

Technical documentation