Waters ACQUITY RDa Detector
Waters ACQUITY RDa Detector — Mass spectrometers with Resolution 10000.
The ACQUITY RDa Detector is a high-resolution time-of-flight mass spectrometer with an electrospray ionization source, covering 50–2000 m/z in both positive and negative modes (extended ranges to 9000 and 5000 m/z respectively). It delivers 10,000 FWHM resolution and <2.5 ppm mass accuracy, distinguishing it as a dedicated intact-mass and data-independent acquisition detector within the Waters ACQUITY series. It is intended for integration into UPLC/MS systems for accurate mass confirmation and profiling.
Full specifications
| Resolution | 10000 FWHM |
| Ion source | Electrospray Ionization (ESI) |
| Scan speed | 1 to 20 Hz |
| Vacuum system mfr | fully automated and differentially pumped vacuum |
| Acquisition modes mfr | Full scan (MS); Full scan with fragmentation (da |
| Positive ion MS sensitivity mfr | Intensity ≥22,000 ions/counts per second, measur |
| Negative ion MS sensitivity mfr | Intensity ≥13,750 ions/counts per second, measur |
| Chromatographic dynamic range mfr | ≥4 orders of magnitude |
| Regulatory approvals/marks mfr | CE, CB, NRTL (CAN/US), RCM |
| Software mfr | Systems supported in waters_connect Software ver |
| Mass range | Positive ion: 50–2000 or 400–9000 m/z; Negative ion: 50–2000 or 400–5000 m/z |
| Mass analyzer mfr | high resolution, high stability Tof mass analyze |
| Mass measurement accuracy mfr | RMS error <2.5 ppm |
| Mass scale calibration accuracy mfr | <1 ppm with internal Lock Mass correction over r |
Description
Construction and architecture
- High-resolution, high-stability Tof mass analyzer with a differentially pumped vacuum system comprising an air-cooled split-flow turbo pump and one rotary backing pump (oil-free option available). - Acquisition modes: full scan MS and full scan with fragmentation (data-independent acquisition). - Mass range selectable: positive ion 50–2000 or 400–9000 m/z; negative ion 50–2000 or 400–5000 m/z. ### Performance under load
- Resolution 10,000 FWHM; mass measurement accuracy RMS error <2.5 ppm; mass scale calibration accuracy <1 ppm with internal Lock Mass correction over 50–2000 m/z. - Sensitivity: positive ion ≥22,000 ions/counts per second from m/z 556.28; negative ion ≥13,750 from m/z 554.26. - Chromatographic dynamic range ≥4 orders of magnitude; scan speed 1–20 Hz. ### Installation, connection and control
- Controlled via waters_connect Software version 4.0.0; no other software compatibility is stated. - Vacuum system is fully automated; the turbo pump is air-cooled, requiring adequate ventilation. - Regulatory approvals: CE, CB, NRTL (CAN/US), RCM.
Integration & Operating Notes
The split-flow turbo pump is air-cooled; installation must provide sufficient clearance and ambient airflow to prevent overheating. The rotary backing pump is oil-free only if that option is specified.
Positive ion mode offers 50–2000 or 400–9000 m/z; negative ion mode offers 50–2000 or 400–5000 m/z. Exceeding the selected range or using a range not supported in the chosen polarity will yield no valid data.
Mass scale calibration accuracy of <1 ppm is achieved only with internal Lock Mass correction over 50–2000 m/z. Without Lock Mass, accuracy degrades to the RMS <2.5 ppm level; routine verification of Lock Mass reference is necessary.
Selection & Integration Q&A
The detector must be controlled by waters_connect Software version 4.0.0. It uses an ESI source; no specific LC coupling requirements are stated, but the system must support the selected acquisition mode (full scan or full scan with fragmentation).
The vacuum system is fully automated; the rotary backing pump is oil-free only if that option is ordered. No maintenance schedule, cables, or consumables are listed in the supplied facts; Lock Mass reference materials are needed for <1 ppm calibration.