Waters ACQUITY UPC2 System
Waters ACQUITY UPC2 System — Chromatographs with Flow rate 4 milliliter/minute, Max pressure 41.3 megapascal.
The ACQUITY UPC2 System is an ultra-performance convergence chromatography instrument combining a high-pressure CO2 pump, autosampler with two 384-well plate positions, and a column compartment with precise temperature control. It delivers flow rates up to 4.0 mL/min and pressures up to 413 bar, with injection volumes up to 50.0 µL and sample temperature control from 4.0 to 40.0 °C. This system is designed for analytical laboratories requiring efficient, high-resolution separations with reduced solvent consumption.
Full specifications
| Flow rate | 4.0 mL/min |
| Max pressure | 413 bar |
| Column temperature | 90 °C |
| Injection volume | 50.0 μL |
| Stroke volume mfr | 140 μL |
| Number of sample plates mfr | Total of two plates |
| Maximum sample capacity mfr | 768 in two 384-well plates |
| Number of sample injections mfr | 1 to 99 injections per sample |
| Sample temperature control mfr | 4.0 to 40.0 °C |
| Minimum sample required mfr | 5 μL residual |
| Sample carryover mfr | <0.005% or <2.000 nL |
| Column compartment temperature accuracy mfr | ±0.5 °C |
| Column compartment temperature precision mfr | ±0.1 °C |
| Column compartment temporal temperature stability mfr | ±0.3 °C |
| Column heater temperature range mfr | 20.0 to 90.0 °C |
| Pressure control stability mfr | ±7.25 psi (0.5 bar) |
| CO2 purity level mfr | 99.97% |
| Wavelength range mfr | 190 to 800 nm |
| Wavelength accuracy mfr | ±1.0 nm |
| Optical resolution mfr | 1.2 nm |
| Data rate mfr | Up to 80 Hz |
| Height | 672 mm |
| Width | 343 mm |
| Depth mfr | 71.1 cm (28.0 in.) |
| Path length mfr | 10 mm (analytical cell) |
| pH range mfr | 2.0 to 12.0 |
Description
Construction and Architecture
The ACQUITY UPC2 System integrates a CO2 pump, autosampler, column heater, and photodiode array detector into a single benchtop unit measuring 672 mm (H) × 343 mm (W) × 711 mm (D). The autosampler accommodates two 384-well plates (maximum 768 samples) and supports 1 to 99 injections per sample, with a minimum sample volume of 5 µL residual. The column compartment maintains temperatures from 20.0 to 90.0 °C with ±0.5 °C accuracy and ±0.1 °C precision, ensuring retention time stability. The system uses 99.97% pure CO2 as the primary mobile phase, with pressure control stability of ±0.5 bar. ### Behavior Under Load and in Different Media
The system operates at pressures up to 413 bar and flow rates up to 4.0 mL/min, enabling high-efficiency separations with sub-2 µm particle columns. The injection system achieves carryover of <0.005% or <2.000 nL, critical for trace analysis. The UV/Vis detector covers 190–800 nm with ±1.0 nm wavelength accuracy and 1.2 nm optical resolution, supporting a wide range of analytes. The column heater's temporal temperature stability of ±0.3 °C ensures reproducible results across long runs. ### Installation, Connection, and Control
Installation requires a CO2 supply of at least 99.97% purity and a compatible solvent delivery system. The system interfaces with Waters Empower software for control and data acquisition, with data rates up to 80 Hz. The flow path materials are compatible with pH 2.0 to 12.0, requiring careful selection of mobile phase additives. The autosampler's sample temperature control (4.0–40.0 °C) necessitates a chilled water or Peltier-based recirculating chiller, which is not included in the base delivery.
Integration & Operating Notes
Using CO2 with purity below 99.97% introduces impurities that can foul the pump, contaminate the column, and cause baseline noise. Install a high-purity CO2 source with appropriate filtration and verify purity before connection.
The system's electrical load includes the column heater, autosampler chiller, and detector; a dedicated circuit prevents voltage fluctuations that could affect retention times and detector sensitivity. Consult the site preparation guide for exact power requirements.
The system's pressure control stability is ±0.5 bar; injections initiated before stabilization can lead to retention time variability. Implement a delay in the method to allow pressure to reach setpoint after flow ramp.
CO2 pump seals wear over time, causing pressure fluctuations and leaks. Waters recommends replacement every 6 months or after 1000 hours of operation, whichever comes first. Use only Waters-certified seal kits.
Selection & Integration Q&A
Choose this model for standard analytical workflows requiring flow rates up to 4.0 mL/min and pressures up to 413 bar. For higher pressures or flow rates, consider other UPC2 variants; this system is optimized for routine method development and quality control.
The system requires CO2 of at least 99.97% purity, supplied via a 1/8-inch Swagelok fitting at a pressure of 80–100 psi. Use a two-stage regulator with stainless steel diaphragm and a coalescing filter to remove particulates and oil.
The base system does not include a CO2 regulator, chiller for sample temperature control, column, or Empower software license. Additionally, pump seal replacement kits and sample plates must be ordered separately.