Waters ACQUITY UPLC I-Class PLUS System
Waters ACQUITY UPLC I-Class PLUS System — Chromatographs with Flow rate 2 milliliter/minute, Max pressure 124.1 megapascal.
The ACQUITY UPLC I-Class PLUS System is a binary, stainless-steel flow path UPLC chromatograph with active pre-heating, rated to 18,000 psi (124 MPa) and 2.0 mL/min. It supports columns up to 150 mm × 4.6 mm I.D. and up to 20 sample plates with the Sample Organizer, distinguishing it within the I-Class series by its PLUS configuration. It serves as the high-pressure separation front end for PDA, TUV, FLR, CA, RI, ELS, or mass spectrometric detection in regulated analytical workflows.
Full specifications
| Flow rate | 2.0 mL/min |
| Max pressure | 18000 psi |
| Column temperature | 4–40 °C |
| Sample flow path mfr | Stainless-steel |
| Sample carryover mfr | <0.001% (Caffeine) under UV conditions |
| Pump options mfr | Binary |
| Solvent conditioning mfr | Active Pre-heating |
| Column capacity mfr | Up to 2 columns of 150 mm length and up to 4.6 m |
| Maximum sample capacity mfr | Up to 20 plates (with Sample Organizer) |
| Detector options mfr | PDA Detector, TUV Detector, FLR Detector, Charge |
| External control mfr | Empower Software, MassLynx Software, waters_conn |
| Height | 711 mm |
| Width | 343 mm |
| Depth mfr | 71.2 cm (28.0 in.) |
| pH Range mfr | 1–12.5 |
Description
Construction and positioning
- Binary pump with stainless-steel sample flow path; active pre-heating conditions solvent before the column. - Column compartment holds up to two columns of 150 mm length and up to 4.6 mm I.D.; temperature controlled from 4 to 40 °C. - Sample capacity extends to 20 plates when the Sample Organizer is fitted. ### Behavior under load and in different media
- Maximum operating pressure is 18,000 psi (≈124.1 MPa) at a maximum flow of 2.0 mL/min. - Compatible with mobile phases from pH 1 to 12.5, suiting acidic through strongly basic separations. - Sample carryover is specified as <0.001% (caffeine) under UV conditions, supporting low-level analyte work. ### Installation, control and operation
- External control is available through Empower, MassLynx, waters_connect, or UNIFI Scientific Information System. - Detector options include PDA, TUV, FLR, charged aerosol, RI, ELS, and mass spectrometers. - Physical envelope: 711 mm height, 343 mm width, 712 mm depth. - No PDF digests were attached, so detailed installation, connection, and control specifics cannot be cited beyond the factory specification sheet.
Integration & Operating Notes
The pump and flow path are rated to 18,000 psi (124.1 MPa) and 2.0 mL/min. Operating near both maxima simultaneously raises wear and leak risk; verify column and tubing pressure ratings before method transfer from lower-pressure HPLC.
The stainless-steel flow path tolerates pH 1 to 12.5. Outside this window, corrosion and altered retention can occur; confirm buffer and additive pH before running extended gradients.
The system can be driven by Empower, MassLynx, waters_connect, or UNIFI. The selected platform determines network, licensing, and data-integrity configuration, so decide before installation to avoid rework.
The <0.001% caffeine carryover figure under UV conditions is a specification, not a guarantee over time. Establish a preventive maintenance interval for seals, needles, and injection components to preserve it.
Selection & Integration Q&A
Detector options include PDA, TUV, FLR, charged aerosol, RI, ELS, and mass spectrometers. Control is via Empower, MassLynx, waters_connect, or UNIFI. Confirm the detector's pressure and flow ratings against the 18,000 psi and 2.0 mL/min system limits before connection.
Dimensional Drawings & Technical Documentation