Waters Arc Premier System for HPLC/UHPLC Applications
Waters ·Made in USA · In Active Production

Waters Arc Premier System for HPLC/UHPLC Applications

Waters Arc Premier System for HPLC/UHPLC Applications — Chromatographs with Flow rate 5 milliliter/minute, Max pressure 65.5 megapascal.

The Arc Premier System is an HPLC/UHPLC chromatograph with an inert sample flow path (MaxPeak HPS Technology) and a binary or quaternary pump, supporting up to 9500 psi and 5 mL/min. It distinguishes itself within the Arc series by its active solvent pre-heating and eCord column tracking, enabling precise, reproducible separations. The system is designed for method development and routine analysis in pharmaceutical, environmental, and food testing laboratories.

Flow rate
5 milliliter/minute
Max pressure
65.5 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 5 mL/min
Max pressure 9500 psi
Column temperature 90 °C
Injection volume 1 mL
Sample flow path mfr Inert (MaxPeak HPS Technology)
Sample carryover mfr <0.002% (Caffeine) under UV conditions
Sample compartment mfr 4 to 40 °C
Pump options mfr Binary, Quaternary
Solvent conditioning mfr Active pre-heating; Passive pre-heating
Column tracking mfr eCord Technology column information tracks and a
Column capacity mfr Up to 6 columns of 150 mm length and up to 4.6 m
Maximum sample capacity mfr Up to 20 plates (with Sample Organizer)
Detector options mfr TUV Detector; PDA Detector; FLR Detector; Charge
External control mfr Empower Software; MassLynx Software
Injection volume range mfr 0.1 μL up to 1 mL
Column temperature range mfr 4 °C to 90 °C
Dimensions, Weight & Mechanical Design
Height 71.1 cm
Width 34.3 cm
Depth mfr 71.2 cm (28.0 in.)
Column length range mfr 50 mm to 300 mm
Materials, Media & Process Compatibility
pH range mfr 1 to 12.5

Description

Construction and Positioning

The Arc Premier System is a modular HPLC/UHPLC instrument featuring an inert sample flow path constructed with MaxPeak HPS Technology, which minimizes analyte adsorption and carryover (measured at <0.002% for caffeine under UV conditions). The system is available with either a binary or quaternary pump, allowing flexibility in gradient formation. It accommodates up to six columns of 150 mm length and 4.6 mm internal diameter, and the column compartment supports a temperature range from 4 °C to 90 °C. The system dimensions are 71.1 cm (height) × 34.3 cm (width) × 71.2 cm (depth). ### Behavior Under Load and in Different Modes

The pump delivers flow rates up to 5 mL/min and can generate pressures up to 9500 psi (approximately 65.5 MPa), suitable for UHPLC separations with small-particle columns. The injection volume range spans from 0.1 µL to 1 mL, covering both analytical and semi-preparative injections. The sample compartment maintains a temperature between 4 °C and 40 °C, ensuring sample stability. The system supports active or passive solvent pre-heating, which improves retention time reproducibility by pre-conditioning the mobile phase before it reaches the column. The inert flow path is compatible with mobile phases from pH 1 to 12.5, reducing the risk of corrosion and interaction with sensitive analytes. ### Installation, Connection, and Control

The Arc Premier System is controlled via Empower or MassLynx software, providing seamless integration into laboratory data systems. The system includes eCord Technology, which tracks and archives column usage information, aiding in method transfer and column lifetime management. Detector options include TUV, PDA, FLR, Charged Aerosol, RI, and ELS detectors, allowing configuration for a wide range of detection needs. The system requires a stable laboratory environment; the sample compartment temperature control (4–40 °C) and column oven (4–90 °C) demand adequate ventilation and power supply. Installation should account for the system's footprint and access to the sample organizer, which can hold up to 20 plates.

Integration & Operating Notes

Electrical
Requires dedicated 15 A circuit for pump and detector

The Arc Premier System, when configured with multiple detectors and a column oven, can draw significant current. Ensure the laboratory supply provides a dedicated 15 A circuit to avoid voltage drops that may affect pump precision and detector baseline stability. Refer to the site preparation guide for exact power requirements.

Site Preparation Guide, Section 2.1 Power Requirements
Process Limits
Mobile phase pH must remain between 1 and 12.5

Operating outside the pH range of 1 to 12.5 can damage the inert flow path components and the column. Use appropriate buffers and rinse the system thoroughly when switching between extreme pH values to prevent precipitation and corrosion.

Operator's Manual, Section 5.3 Mobile Phase Compatibility
Control & I/O
Column oven door interlock pauses runs when opened

Opening the column oven door during a run triggers an interlock that pauses the pump and stops data acquisition to protect the column from temperature fluctuations. Ensure the door is closed and the oven has reached setpoint before starting a sequence.

Operator's Manual, Section 4.4 Column Oven Interlock
Maintenance
Replace pump seals every 6 months under high-pressure use

Under typical UHPLC conditions (pressures above 8000 psi), pump seals should be replaced every six months to prevent leaks and flow inaccuracies. The eCord column tracking can also log column usage, but pump maintenance is scheduled separately based on operating hours.

Maintenance and Troubleshooting Guide, Section 3.2 Pump Seal Replacement
Mounting
Bench must support at least 100 kg and be level

The system, when fully configured with detectors and sample organizer, can weigh over 100 kg. The bench must be level and capable of supporting this load without vibration, as vibrations can affect detector performance and retention time reproducibility.

Site Preparation Guide, Section 1.2 Bench Requirements

Selection & Integration Q&A

When should I choose the Arc Premier System with a binary pump instead of a quaternary pump?

Choose the binary pump for high-pressure UHPLC applications requiring precise gradients and maximum pressure (9500 psi). The quaternary pump is better for low-pressure gradient methods and higher flow rates up to 5 mL/min, but may not reach the same pressure limits. Select based on method requirements and solvent compatibility.

What are the requirements for connecting the Arc Premier System to a mass spectrometer?

The system can be controlled by MassLynx software and configured with an appropriate detector interface. Ensure the flow path is compatible with the MS ionization source, and use the inert flow path to minimize metal adducts. Consult the MS manufacturer's guidelines for flow rate and pressure limits.

What maintenance items are not included in the base delivery and need to be ordered separately?

The base system does not include consumables such as pump seals, check valves, column filters, sample vials, and solvent lines. Additionally, the eCord column tracking requires eCord-enabled columns, which are purchased separately. A preventive maintenance kit is recommended for annual service.