Waters LC Prep AutoPurification System for Liquid Chromatography
Waters LC Prep AutoPurification System for Liquid Chromatography — Chromatographs with Flow rate 150 milliliter/minute, Max pressure 41.37 megapascal.
The Waters LC Prep AutoPurification System is a preparative-scale liquid chromatography instrument configured with a binary high-pressure gradient pump, delivering up to 150 mL/min at pressures to 6000 psi. Its 0.35 mL delay volume and 20 µL analytical injection loop (1.0 mL preparative loop standard) distinguish it as a dedicated purification platform within the Waters LC Prep series. It is intended for automated fraction collection and compound isolation in pharmaceutical, biotech, and chemical research workflows.
Full specifications
| Flow rate | 150.0 mL/min |
| Max pressure | 6000 psi |
| Injection volume | 20 µL |
| Delay volume mfr | 0.35 mL |
| Number of solvents mfr | Up to 4: A1 or A2 and B1 or B2 |
| Size (W x H x D, cm) mfr | 28 x 43.2 x 66 |
| External Control mfr | MassLynx Software with FractionLynx |
| Detectors mfr | Ultraviolet/visible (UV/Vis) absorbance; Evapora |
| Injector volume (fixed sample loop) mfr | Analytical valve: 20 μL loop standard; Preparati |
| Syringe sizes mfr | 1.0 mL (standard), 2.5 mL, 5 mL, and 12.5 mL |
| Height | 43.2 cm |
| Weight | 39.5 kg |
| Width | 28 cm |
Description
Construction and Architecture
- The system integrates a binary solvent delivery module (up to four solvents: A1 or A2 and B1 or B2), an autosampler with a standard 20 µL analytical loop and a 1.0 mL preparative loop, and a fraction collector, all under MassLynx with FractionLynx control. - Physical dimensions: 28 cm W × 43.2 cm H × 66 cm D; weight 39.5 kg. The compact footprint suits standard laboratory benches. - Delay volume of 0.35 mL minimizes gradient lag, enabling precise retention-time-based fraction cutting. - Detector compatibility includes UV/Vis absorbance, evaporative light scattering (ELS), photodiode array (PDA), and mass spectrometry (MS), allowing versatile detection for purification. ### Behavior Under Load and in Different Modes
- The pump delivers up to 150 mL/min at a maximum pressure of 6000 psi (≈41.4 MPa), supporting both analytical and preparative flow rates. - The standard 20 µL analytical loop is suited for method development and small-scale injections; the 1.0 mL preparative loop accommodates larger sample loads for purification. - Syringe sizes of 1.0 mL (standard), 2.5 mL, 5 mL, and 12.5 mL provide flexibility for sample handling and loop loading. - With up to four solvents, the system can perform binary high-pressure gradients or alternate solvent selection (A1/A2, B1/B2) for method scouting and cleaning. ### Installation, Connection, and Control
- External control is via MassLynx Software with FractionLynx, which manages gradient programming, injection sequences, and fraction collection based on time or detector signal. - The system requires connection to solvent reservoirs, waste, and a suitable detector (UV/Vis, ELS, PDA, or MS). Electrical requirements are not specified in the provided facts; refer to the installation manual. - The preparative valve and analytical valve are integrated; switching between them is software-controlled.
Integration & Operating Notes
Operating above the maximum pressure rating of 6000 psi (≈41.4 MPa) can damage pump seals, tubing, and fittings, and may cause leaks or unsafe operation. Always verify that the column and plumbing are rated for the intended pressure and set the pump's upper pressure limit accordingly in the software.
The preparative and analytical valves are switched automatically by FractionLynx based on the run method. Ensure that the method specifies the correct valve position and that the injector is configured for the installed loop size (20 µL analytical or 1.0 mL preparative) to avoid sample loss or overloading.
After using buffer-containing mobile phases, flush the entire flow path (pump, injector, valves, and tubing) with HPLC-grade water followed by an appropriate organic solvent (e.g., methanol or acetonitrile) to prevent salt precipitation and microbial growth. This is especially critical for the preparative valve and loop, which have small internal passages.
Selection & Integration Q&A
Choose this model when you need automated preparative purification with flow rates up to 150 mL/min and pressures to 6000 psi, plus flexibility for up to four solvents and multiple detectors (UV/Vis, ELS, PDA, MS). Sibling models may offer different flow or pressure ranges; select based on your required scale and detection needs.
The system supports UV/Vis absorbance, evaporative light scattering (ELS), photodiode array (PDA), and mass spectrometry (MS) detectors. Ensure the detector's flow cell is rated for the system's maximum pressure (6000 psi) and that its communication interface (e.g., Ethernet or analog) is compatible with MassLynx/FractionLynx control.
Operate in a temperature-controlled laboratory (typically 15–30 °C) with <80% relative humidity and away from direct sunlight and vibration. The system is designed for continuous duty but requires periodic maintenance (e.g., seal replacement) as specified in the manual; duty cycle limits are not explicitly stated in the provided facts.
The base system includes the pump, autosampler, fraction collector, and standard 20 µL analytical loop and 1.0 mL preparative loop. Detectors, columns, solvent reservoirs, waste containers, and additional syringes (2.5 mL, 5 mL, 12.5 mL) are not included and must be ordered separately. Routine maintenance parts such as pump seals and check valves are also consumables.