Waters PATROL UPLC Process Analysis System
Waters PATROL UPLC Process Analysis System — Chromatographs with Flow rate 2 milliliter/minute, Max pressure 0.713 megapascal.
The PATROL UPLC Process Analysis System is a full-loop injection chromatograph with binary or quaternary solvent delivery, up to 32 sample positions, and a 1:100 dilution range. It operates at flow rates up to 2.0 mL/min and 103.4 psi, with column temperature up to 90 °C and injection volumes of 1, 2, or 5 μL. It is intended for automated process and control sample analysis in a system, providing inject-to-inject cycle times under 15 seconds in binary configuration.
Full specifications
| Flow rate | 2.0 mL/min |
| Max pressure | 103.4 psi |
| Column temperature | 90.0 °C |
| Injection volume | 5 μL |
| Total system delay volume mfr | <120 μL with binary configuration, with standard |
| Cycle time mfr | <15 s inject-to-inject with binary configuration |
| Number of solvents mfr | Up to four, in any combination of two: A1 or A2, |
| Flow accuracy mfr | ±1.0% at 0.500 mL/min, as per Waters AQT/Systems |
| Flow precision mfr | 0.075 %RSD or ±0.01 min SD, (0.500 to 2.000 mL/m |
| Priming mfr | Wet priming runs at a flow rate of 4 mL/min |
| Pump compositional accuracy mfr | ±0.5% absolute (full scale), from 5% to 95% of f |
| Pump compositional precision mfr | 0.15 %RSD or ±0.04 min SD, whichever is greater, |
| Maximum sample capacity mfr | 32 available sites for process and control sampl |
| Internal sample capacity mfr | 32 positions for standard 8 mL vials |
| External sampling mfr | 1 port standard |
| Number of sample injections mfr | Up to 60 injections per sample vial |
| Injection mode mfr | Full-loop injection only |
| Injection loop volumes mfr | 1 μL, 2 μL, or 5 μL sample loops only |
| Dilution range mfr | Up to 1:100 |
| Injection precision (without dilution) mfr | < or =0.5 %RSD for caffeine peak area, full-loop |
| Injection precision with dilution mfr | < or =1.0 %RSD for caffeine peak area, full-loop |
| Dilution linearity mfr | 1 to 50 dilution factor > or =0.999 coefficient |
| Sample carryover mfr | 1 to 50 dilution factor < or =0.01% of the previ |
| Width | 762 mm |
| Power consumption | 450 VAC |
| pH range mfr | pH 2 to 10 |
Description
Construction and positioning
- The system integrates a binary or quaternary pump, an autosampler with 32 positions for 8 mL vials, and a column oven in a single unit 762 mm wide. - It uses full-loop injection only, with fixed loop volumes of 1, 2, or 5 μL, and supports up to 60 injections per vial. - Solvent selection allows up to four solvents in any combination of two: A1 or A2, and B1 or B2. - Total system delay volume is <120 μL with binary configuration and standard 50 μL mixer, or <400 μL with quaternary configuration and standard 100 μL mixer. ### Behavior under load and in different modes
- Flow accuracy is ±1.0% at 0.500 mL/min, and flow precision is 0.075 %RSD or ±0.01 min SD (0.500 to 2.000 mL/min), whichever is greater, using premixed solvent. - Pump compositional accuracy is ±0.5% absolute (full scale) from 5% to 95% of flow rates from 0.5 to 2.0 mL/min; compositional precision is 0.15 %RSD or ±0.04 min SD based on retention time. - Injection precision without dilution is ≤0.5 %RSD for caffeine peak area; with dilution, ≤1.0 %RSD. Dilution linearity from 1 to 50 dilution factor has a coefficient of deviation ≥0.999, and sample carryover is ≤0.01% of the previous injection for caffeine. - The system operates over a pH range of 2 to 10 and at column temperatures up to 90 °C. ### Installation, connection, and operation
- The system requires a 450 VAC power supply. - Priming is performed with wet priming at a flow rate of 4 mL/min. - External sampling is provided via one standard port. - Cycle time is <15 seconds inject-to-inject with binary configuration and <30 seconds with quaternary configuration. - No specific installation, control, or operation details are available from attached documents.
Integration & Operating Notes
The system is rated for 450 VAC; verify that the facility supply can deliver this voltage and the required current without significant voltage drop, as insufficient supply may cause erratic pump or autosampler operation. Ensure proper grounding and circuit protection according to local electrical codes.
Operating outside the pH range of 2 to 10 can degrade the analytical column and wetted parts, leading to loss of performance and potential leakage. Verify sample and mobile phase pH before introduction, and use appropriate buffers if necessary.
Exceeding the maximum pressure rating of 103.4 psi can cause seal failure, tubing rupture, or unsafe operation. Monitor system pressure during method development and ensure that column backpressure and flow rate combinations stay within this limit.
One external sampling port is standard, allowing connection to a process stream or external sample source. Ensure that the sample is compatible with the system's wetted materials and that the pressure and flow are within the specified limits.
Selection & Integration Q&A
Choose binary for faster cycle times (<15 s inject-to-inject) and lower delay volume (<120 μL). Choose quaternary for up to four solvents and higher delay volume (<400 μL) with cycle time <30 s. Select based on required solvent flexibility and throughput.