Waters SFC Prep 350 System
Waters SFC Prep 350 System — Chromatographs with Flow rate 350 milliliter/minute, Max pressure 32 megapascal.
The Waters SFC Prep 350 System is a preparative supercritical fluid chromatography (SFC) instrument configured with a P-350 CO2 pump (35–350 mL/min) and a P-200X co-solvent pump (20–200 mL/min), operating up to 320 bar. It is distinguished within the series by its high flow rate and 2 mL standard injection loop, enabling milligram-to-gram scale purifications. The system is used for fraction collection in pharmaceutical and natural product purification processes.
Full specifications
| Flow rate | 350 mL/min |
| Max pressure | 320 bar |
| Column temperature | 25 °C |
| Injection volume | 2 mL |
| Co-solvent percentage range mfr | 5% to 55% |
| Recommended column size mfr | 3 and 5 cm I.D. |
| Fraction capacity mfr | 5+1 |
| Flow rate range for CO2 pump (P-350) mfr | 35 to 350 mL/min |
| Flow rate range for co-solvent pump (P-200X) mfr | 20 to 200 mL/min |
| Wavelength range mfr | 190 to 700 nm |
| UV lamp mfr | Deuterium |
| Visible lamp mfr | Tungsten/halogen |
| Flow cell mfr | Preparative standard |
| Pathlength mfr | Preparative 0.5 mm |
| Separators mfr | Six 1 L separators for collecting five fractions |
| Height | 135 cm |
| Weight | 195 kg |
| Width | 107 cm |
| Maximum pressure (Mass Flow Meter) mfr | 100 bar (1450 psi) |
| Voltage mfr | 110 to 220 VAC |
| Frequency mfr | 50 to 60 Hz |
| Liquid CO2 inlet pressure mfr | 55 to 100 bar (797 to 1450 psi) |
| Liquid CO2 inlet temperature mfr | 5 to 25 °C |
| Cell material of construction mfr | Sapphire PEEK, 316 stainless-steel |
| Cooling of CO2 pump mfr | Circulating coolant |
Description
Construction and positioning
The SFC Prep 350 System is a preparative-scale SFC instrument built around a P-350 CO2 pump and a P-200X co-solvent pump, with a maximum pressure rating of 320 bar. It is positioned as the highest-flow model in the series, delivering up to 350 mL/min of CO2, compared to lower-flow siblings. The system includes a preparative UV detector with a 0.5 mm pathlength, 2.0 µL flow cell, and a fraction collection module with six 1 L separators (five fractions plus one waste). ### Behavior under load and media
- The CO2 pump requires liquid CO2 at 55–100 bar inlet pressure and 5–25 °C, with cooling provided by a circulating coolant to maintain CO2 in the liquid phase. - Co-solvent percentage can be programmed from 5% to 55%, allowing polarity adjustment for diverse compound classes. - The system operates at a maximum pressure of 320 bar; the mass flow meter is limited to 100 bar. - Column temperature is maintained at 25 °C (ambient) with no active heating or cooling specified. ### Installation, connection, and operation
- The system requires a 110–220 VAC, 50–60 Hz power supply. - The injection loop is 2 mL standard, with a 5 mL syringe for sample loading. - Recommended column sizes are 3 and 5 cm I.D., typical for preparative SFC. - The UV detector uses a deuterium lamp (190–700 nm) and a tungsten/halogen lamp for visible range, with a preparative flow cell (sapphire PEEK, 316 stainless steel). - Fraction collection is automated via the six-separator module, enabling unattended collection of five fractions plus waste.
Integration & Operating Notes
Operating the CO2 pump outside the specified inlet pressure range (55–100 bar) can cause cavitation or pump damage. Ensure the CO2 supply is regulated to maintain liquid phase and is cooled to 5–25 °C before entering the pump.
The CO2 pump is cooled by a circulating coolant to prevent vapor lock and maintain consistent flow. The coolant supply must be connected and maintained at a temperature that keeps the pump head below the CO2 vaporization point under all operating pressures.
The CO2 and co-solvent pumps contain wear parts (seals, check valves) that must be replaced at intervals specified in the service manual to prevent leaks and flow inconsistencies. Use only Waters-approved spare parts to maintain performance and safety.
Selection & Integration Q&A
Choose the 350 when your purification requires flow rates up to 350 mL/min and pressures up to 320 bar, typical for gram-scale separations. Lower-flow models are better for analytical or small-scale prep where sample mass is limited.
The CO2 must be supplied as a liquid at 55–100 bar and 5–25 °C, and the pump requires a circulating coolant to maintain temperature. Ensure proper pressure regulation and cooling to avoid pump cavitation and inconsistent flow.