The diffusion-pump model cannot be upgraded to CI, so CI requirements force selection of a turbo-pump model.
Determines whether the chosen MSD model can later support chemical ionization.
Source
“Turbo pump models are field upgradeable to CI.”
| Reagent gas | Ion polarity | Abundance target | Peakwidth target | Maximum repeller | Maximum emission current, µA | Max electron energy, eV |
|---|---|---|---|---|---|---|
| Methane | Positive | 1x106 | 0.6 | 4 | 240 | 240 |
| Methane | Negative | 1x106 | 0.6 | 4 | 50 | 240 |
| Isobutane | Positive | N/A | N/A | N/A | N/A | N/A |
| Isobutane | Negative | 1x106 | 0.6 | 4 | 50 | 240 |
| Ammonia | Positive | N/A | N/A | N/A | N/A | N/A |
| Ammonia | Negative | 1x106 | 0.6 | 4 | 50 | 240 |
| Feature | Gas flows | Gas flows |
|---|---|---|
| High vacuum pump | Diffusion | Turbo |
| Optimal He column flow mL/min | 1 | 1 to 2 |
| Maximum recommended gas flow mL/min* | 1.5 | 4 |
| Maximum gas flow, mL/min† | 2 | 6.5 |
| Max column id | 0.53 mm (30 m) | 0.53 mm (30 m) |
| CI capability | No | Yes |
| CI reagent flows, mL/min | NA | 1 to 2 |
| JetClean option H2 flow | NA | 0.4 mL/min |
*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity.
| Feature | Gas flows | Gas flows |
|---|---|---|
| Diffusion | Turbo | |
| High vacuum pump | ||
| Optimal He column flow mL/min | 1 | 1 to 2 |
| Maximum recommended gas flow mL/min* | 1.5 | 4 |
| Maximum gas flow, mL/min† | 2 | 6.5 |
| Max column id | 0.53 mm (30 m) | 0.53 mm (30 m) |
| JetClean option H2 flow | NA | 0.4 mL/min |
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity.
| Action | Description |
|---|---|
| High Vacuum Pressure | Only with Micro-Ion vacuum gauge installed. |
| Turbo Pump Speed | Displays the turbo pump speed. |
| Foreline Pressure | Displays the foreline pressure. |
| MSD Fault Status | Reports a summary fault status code (number) in ‘dec’ (decimal) and ‘hex’ (hexadecimal) format covering all possible fault combinations. |
| Ion Source Temp, oC | Displays and sets the ion source temperature. |
| Mass Filter Temp, oC | Displays and sets the mass filter temperature. |
| CI Reagent | Displays CI reagent gas and flow rate (if installed). |
MS parameters cannot be set from the LCP while an online OpenLab CDS session is connected to the MSD.
| Item number | Item description | Part number (SS) | Part number (Inert) |
|---|---|---|---|
| 1 | Gold plated set screw | G1999-20022 | G1999-20022 |
| 2 | Gold plated screw | G3870-20021 | G3870-20021 |
| 3 | Interface socket | G1099-20136 | G1099-20136 |
| 4 | Source body | G1099-20130 | G2589-20043 |
| 5 | Drawout cylinder | G1072-20008 | G1072-20008 |
| 6 | Drawout plate | 05971-20134 | G2589-20100 |
| 7 | 4-turn filament | G7005-60061 | G7005-60061 |
| 8 | Spring washer | 3050-1374 | 3050-1374 |
| Item number | Item description | Part number (SS) | Part number (Inert) |
|---|---|---|---|
| 8 | Flat washer | 3050-0982 | 3050-0982 |
| 9 | Lens insulator | G3170-20530 | G3170-20530 |
| 10 | Entrance lens | G3170-20126 | G3170-20126 |
| 11 | Ion focus lens | 05971-20143 | 05971-20143 |
| 12 | Repeller insulator | G1099-20133 | G1099-20133 |
| 13 | Repeller | G3870-60172 | G3870-60173 |
| 14 | Flat washer | 3050-0627 | 3050-0627 |
| 15 | Belleville spring washer | 3050-1301 | 3050-1301 |
| 16 | Repeller nut | 0535-0071 | 0535-0071 |
| 17 | Source heater block assembly | G3870-60180 | G3870-60179 |
| 18 | Repeller block insert | G3870-20135 | G3870-20135 |
| Item number | Item description | Part number (SSL) | Part number (Inert) |
|---|---|---|---|
| 1 | Gold plated set screw | G1999-20022 | G1999-20022 |
| 2 | Gold plated screw | G3870-20021 | G3870-20021 |
| 3 | Interface socket | G1099-20136 | G1099-20136 |
| 4 | Source body | G1099-20130 | G2589-20043 |
| 5 | Drawout cylinder | G1072-20008 | G1072-20008 |
| Item number | Item description | Part number (SSL) | Part number (Inert) |
|---|---|---|---|
| 6 | Drawout plate | 05971-20134 | G2589-20100 |
| 7 | 4-turn filament | G7005-60061 | G7005-60061 |
| 8 | Spring washer | 3050-1374 | 3050-1374 |
| 8 | Flat washer | 3050-0982 | 3050-0982 |
| 9 | Lens insulator | G3170-20530 | G3170-20530 |
| 10 | Entrance lens | G3170-20126 | G3170-20126 |
| 11 | Ion focus lens | 05971-20143 | 05971-20143 |
| 12 | Repeller insulator | G1099-20133 | G1099-20133 |
| 13 | Repeller | G3870-60172 | G3870-60173 |
| 14 | Flat washer | 3050-0627 | 3050-0627 |
| 15 | Belleville spring washer | 3050-1301 | 3050-1301 |
| 16 | Repeller nut | 0535-0071 | 0535-0071 |
| 17 | Source heater block assembly | G3870-60180 | G3870-60179 |
| 18 | Repeller block insert | G3870-20135 | G3870-20135 |
| Feature | Diffusion | Turbo |
|---|---|---|
| High vacuum pump | Diffusion | Turbo |
| Optimal He column flow mL/min 1 | 1 to 2 | |
| Maximum recommended gas flow mL/min* | 1.5 | 4 |
| Maximum gas flow, mL/min† | 2 | 6.5 |
| Max column id | 0.25 mm (30 m) | 0.53 mm (30 m) |
| CI capability‡ | No | Yes |
| Inert ion sources available | Yes | Yes |
| GC compatibility | 9000 and 8890 | 9000 and 8890 |
| Foreline pumps available | Pfeiffer Duo | Pfeiffer Duo, Pfeiffer MVP-070-3x, Agilent IDP-3 24V |
| DIP** capability (3rd party) | Yes | Yes |
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity. ‡ Turbo pump models are field upgradeable to CI. ** Direct insertion probe.
| Item number | Item description | Part number (XTR) |
|---|---|---|
| 1 | Source finger grip | G7002-20008 |
| 2 | Filament block | G7002-20019 |
| 3 | Extractor lens (5)*, with 3 mm opening | G7004-20061 |
The diffusion-pump model cannot be upgraded to CI, so CI requirements force selection of a turbo-pump model.
Determines whether the chosen MSD model can later support chemical ionization.
“Turbo pump models are field upgradeable to CI.”
Diffusion-pump models accept only 0.25 mm columns and lower gas flows than turbo models, constraining column and flow choices.
Column diameter and carrier flow must match the pump type selected.
“Max column id 0.25 mm (30 m)”
The Intuvo 9000 GC cannot be used with CI on this MSD, so CI users must pair with an 8890 GC.
GC platform choice is constrained when CI operation is required.
“If you are using the Agilent Intuvo 9000 Gas Chromatograph with your MSD, Chemical Ionization (CI) is not currently supported.”
Columns must be conditioned before connection to the MSD interface to avoid source contamination and sensitivity loss.
Prevents column bleed deposits in the ion source that degrade sensitivity.
“Conditioning a column before it is connected to the GC/MSD interface is essential.”
Carrier gas purity must be at least 99.9995% to protect the column and MSD during installation and operation.
Gas purity affects column bleed, source contamination, and data quality.
“Always use carrier gas that is at least 99.9995% pure.”
New ferrules can loosen after thermal cycling, so tightness must be rechecked or self-tightening nuts used.
Prevents leaks at the interface after initial heat cycles.
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times.”
Intuvo guard and jumper chips are single-use consumables that deform on installation and cannot be reused or cleaned.
Spares must be stocked and misinstalled chips replaced, not reinstalled.
“The column Guard Chip and the Jumper Chip are both single-use, consumable parts.”
After pumpdown reaches 'Okay to run', a two-hour thermal equilibration wait is needed for reproducible data.
Acquiring before thermal equilibrium yields non-reproducible results.
“After the message Okay to run appears, wait 2 hours for the MS to reach thermal equilibrium.”
CI reagent gas flow is set as a percentage of maximum, recommended 20% for PCI and 40% for NCI.
Sets correct reagent gas flow for PCI versus NCI methods.
“The recommended flow is 20% for a PCI source and 40% for an NCI source.”
After switching from EI to CI or venting, the MSD must be baked out at least two hours before tuning.
Prevents contamination and failed tunes after mode changes.
“the MSD must be baked out for at least 2 hours before tuning.”
Hydrogen must be shut off at its source before MSD power is removed, or it can accumulate and explode.
Critical safety interlock when using hydrogen carrier or JetClean gas.
“the hydrogen gas flow must be off before turning off the MSD power.”
If a power failure occurs while unattended, hydrogen must be shut off and the manifold opened to atmosphere with a 10-minute wait.
Defines the safe restart sequence after an unattended power failure with hydrogen.
“Wait at least 10 minutes to allow any hydrogen to dissipate.”
The quadrupole mass filter needs no periodic maintenance and should generally never be disturbed; cleaning is service-only.
Avoids unnecessary invasive maintenance that could damage the analyzer.
“The mass filter (quadrupole) requires no periodic maintenance.”
The HED ceramic insulator must never be touched during analyzer maintenance.
Prevents damage to sensitive detector components during service.
“The HED ceramic insulator must never be touched.”