Diffusion-pump models cannot be upgraded to CI, so CI capability must be chosen at purchase.
Determines whether the CI option can be added later or must be ordered with the instrument.
Source
“Turbo pump models are field upgradeable to CI.”
| Reagent gas | Methane | Methane | Isobutane | Isobutane | Ammonia | Ammonia |
|---|---|---|---|---|---|---|
| Ion polarity | Positive | Negative | Positive | Negative | Positive | Negative |
| Abundance target | 1x106 | 1x106 | N/A | 1x106 | N/A | 1x106 |
| Peakwidth target | 0.6 | 0.6 | N/A | 0.6 | N/A | 0.6 |
| Maximum repeller | 4 | 4 | N/A | 4 | N/A | 4 |
| Maximum emission current, µA | 240 | 50 | N/A | 50 | N/A | 50 |
| Max electron energy, eV | 240 | 240 | N/A | 240 | N/A | 240 |
N/A: Not available. There are no PFDTD ions formed in PCI with any reagent gas but methane, hence, CI autotune is not available with these configurations.
| Feature | Gas flows | |
|---|---|---|
| High vacuum pump | Diffusion | Turbo |
| Optimal He column flow mL/min | 1 | 1 to 2 |
| Maximum recommended gas flow mL/min* | 1.5 | 4 |
| Maximum gas flow, mL/min† | 2 | 6.5 |
| Max column id | 0.53 mm (30 m) | 0.53 mm (30 m) |
| CI capability | No | Yes |
| CI reagent flows, mL/min | NA | 1 to 2 |
| JetClean option H2 flow | NA | 0.4 mL/min |
*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity.
| Item number | Item description | Part number (SS) | Part number (Inert) |
|---|---|---|---|
| 1 | Gold plated set screw | G1999-20022 | G1999-20022 |
| 2 | Gold plated screw | G3870-20021 | G3870-20021 |
| 3 | Interface socket | G1099-20136 | G1099-20136 |
| 4 | Source body | G1099-20130 | G2589-20043 |
| 5 | Drawout cylinder | G1072-20008 | G1072-20008 |
| 6 | Drawout plate | 05971-20134 | G2589-20100 |
| 7 | 4-turn filament | G7005-60061 | G7005-60061 |
| 8 | Spring washer | 3050-1374 | 3050-1374 |
| Item number | Item description | Part number (SS) | Part number (Inert) |
|---|---|---|---|
| 8 | Flat washer | 3050-0982 | 3050-0982 |
| 9 | Lens insulator | G3170-20530 | G3170-20530 |
| 10 | Entrance lens | G3170-20126 | G3170-20126 |
| 11 | Ion focus lens | 05971-20143 | 05971-20143 |
| 12 | Repeller insulator | G1099-20133 | G1099-20133 |
| 13 | Repeller | G3870-60172 | G3870-60173 |
| 14 | Flat washer | 3050-0627 | 3050-0627 |
| 15 | Belleville spring washer | 3050-1301 | 3050-1301 |
| 16 | Repeller nut | 0535-0071 | 0535-0071 |
| 17 | Source heater block assembly | G3870-60180 | G3870-60179 |
| 18 | Repeller block insert | G3870-20135 | G3870-20135 |
| Item number | Item description | Part number (SSL) | Part number (Inert) |
|---|---|---|---|
| 1 | Gold plated set screw | G1999-20022 | G1999-20022 |
| 2 | Gold plated screw | G3870-20021 | G3870-20021 |
| 3 | Interface socket | G1099-20136 | G1099-20136 |
| 4 | Source body | G1099-20130 | G2589-20043 |
| 5 | Drawout cylinder | G1072-20008 | G1072-20008 |
| Item number | Item description | Part number (SSL) | Part number (Inert) |
|---|---|---|---|
| 6 | Drawout plate | 05971-20134 | G2589-20100 |
| 7 | 4-turn filament | G7005-60061 | G7005-60061 |
| 8 | Spring washer | 3050-1374 | 3050-1374 |
| 8 | Flat washer | 3050-0982 | 3050-0982 |
| 9 | Lens insulator | G3170-20530 | G3170-20530 |
| 10 | Entrance lens | G3170-20126 | G3170-20126 |
| 11 | Ion focus lens | 05971-20143 | 05971-20143 |
| 12 | Repeller insulator | G1099-20133 | G1099-20133 |
| 13 | Repeller | G3870-60172 | G3870-60173 |
| 14 | Flat washer | 3050-0627 | 3050-0627 |
| 15 | Belleville spring washer | 3050-1301 | 3050-1301 |
| 16 | Repeller nut | 0535-0071 | 0535-0071 |
| 17 | Source heater block assembly | G3870-60180 | G3870-60179 |
| 18 | Repeller block insert | G3870-20135 | G3870-20135 |
| Feature | Diffusion | Turbo |
|---|---|---|
| High vacuum pump | Diffusion | Turbo |
| Optimal He column flow mL/min 1 | 1 to 2 | |
| Maximum recommended gas flow mL/min* | 1.5 | 4 |
| Maximum gas flow, mL/min† | 2 | 6.5 |
| Max column id | 0.25 mm (30 m) | 0.53 mm (30 m) |
| CI capability‡ | No | Yes |
| Inert ion sources available | Yes | Yes |
| GC compatibility | 9000 and 8890 | 9000 and 8890 |
| Foreline pumps available | Pfeiffer Duo | Pfeiffer Duo, Pfeiffer MVP-070-3x, Agilent IDP-3 24V |
| DIP** capability (3rd party) | Yes | Yes |
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity. ‡ Turbo pump models are field upgradeable to CI. ** Direct insertion probe.
| Reagent gas/mode | Reagent ion masses | PFDTD Calibrant ions | Flow adj ions: Ratio |
|---|---|---|---|
| EI/PCI/NCI MSD Turbo pump Recommended flow: 20% PCI 40% NCI | |||
| Methane/PCI | 17, 29, 41* | 41, 267, 599 | 28/27: 1.5 – 5.0 |
| Methane/NCI | 17, 35, 235† | 185, 351, 449 | N/A |
| Isobutane/PCI | 39, 43, 57 | N/A | 57/43: 5.0 – 30.0 |
| Isobutane/NCI | 17, 35, 235 | 185, 351, 449 | N/A |
| Ammonia/PCI | 18, 35, 52 | N/A | 35/18: 0.1 – 1.0 |
| Ammonia/NCI | 17, 35, 235 | 185, 351, 517 | N/A |
* There are no PFDTD ions formed with any reagent gas but methane. Tune with methane and use the same parameters for the other gas. † There are no negative reagent gas ions formed. To pretune in negative mode, use background ions: 17 (OH-), 35 (Cl-), and 235 (ReO3-). These ions cannot be used for reagent gas flow adjustment. Set the flow to 40% for NCI and adjust as necessary to get acceptable results for your application.
| Item number | Item description | Part number (XTR) |
|---|---|---|
| 1 | Source finger grip | G7002-20008 |
| 2 | Filament block | G7002-20019 |
| 3 | Extractor lens (5)*, with 3 mm opening | G7004-20061 |
| Item number | Item description | Part number (XTR) |
|---|---|---|
| 4 | Ceramic insulator for extractor | G7002-20064 |
| 5 | Entrance lens assembly, Extended, HES (1)* | G7004-20065 |
| 6 | Ion focus lens (2)* | G7004-20068 |
| 7 | Lens insulator/holder | G7002-20074 |
| 8 | M2 x 0.4 screw x 12 mm long gold plated screw | G7002-20083 |
| 9 | Source body | G7002-20084 |
| 10 | Post extractor lens 2 (3)* | G7004-20090 |
| 11 | Post extractor lens 1 (4)* | G7004-20004 |
| 12 | M2 x 6 mm gold plated screw | G7002-20109 |
| 13 | Locking ring lens insulator | G7002-20126 |
| 14 | High efficiency dual filament | G7002-60001 |
| 15 | Ring heater/sensor assembly | G7002-60043 |
| 16 | Source mount 1.5 mm | G7002-60053 |
| 17 | Repeller assembly | G7002-67057 |
| Not shown | HES assembly | G7004-67056 |
* The number in parenthesis is the number engraved on the lens
Diffusion-pump models cannot be upgraded to CI, so CI capability must be chosen at purchase.
Determines whether the CI option can be added later or must be ordered with the instrument.
“Turbo pump models are field upgradeable to CI.”
The diffusion-pump model accepts only up to 1.5 mL/min total gas flow versus 4 mL/min for turbo models.
Limits column flow and reagent gas choices when selecting the pump configuration.
“Maximum recommended gas flow mL/min*”
The diffusion-pump model is limited to 0.25 mm ID columns, while turbo models accept 0.53 mm ID columns.
Constrains column selection and method transfer between MSD models.
“Max column id”
Column installation procedures differ from earlier MSDs, so using another instrument's procedure can damage the column or MSD.
Prevents installers from reusing older 5975/5977 procedures that may not fit the 5977C interface.
“Using the procedure from another instrument may not work, and may damage the column or the MSD.”
Carrier gas must be at least 99.9995% pure to protect the column and MSD.
Sets the minimum gas purity specification for the installation.
“Always use carrier gas that is at least 99.9995% pure.”
New ferrules may loosen after thermal cycling, so tightness must be rechecked or self-tightening nuts used.
Requires a leak check after the first heating cycles to avoid air leaks into the MSD.
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times.”
Hydrogen flow must be shut off before MSD power is removed, otherwise hydrogen can accumulate and explode.
Defines the mandatory shutdown sequence when hydrogen is used as carrier or JetClean gas.
“the hydrogen gas flow must be off before turning off the MSD power.”
The GC oven and GC/MSD interface must be cool before carrier gas is shut off to avoid column damage.
Sets the correct order of operations during venting and shutdown.
“Ensure the GC oven and the GC/MSD interface are cool before turning off carrier gas flow to prevent damage to the column.”
The tune file must match the installed ion source type.
Prevents wrong tuning and poor data when EI, XTR, HES, or CI sources are swapped.
“The tune file must match the type of ion source in the analyzer.”
In CI mode an air leak is detected by a visible m/z 32 peak, and operating with a leak rapidly contaminates the source.
Gives the acceptance criterion for leak checking before CI operation.
“A peak there indicates an air leak. Repair the leak before proceeding.”
In CI mode the m/z 19 peak must be less than 50% of the m/z 17 peak before proceeding.
Defines the water background criterion for a valid methane reagent gas setup.
“The peak at m/z 19 (protonated water) is less than 50% of the peak at m/z 17.”
EI performance must be verified before switching to CI, and CI must be set up in PCI first even for NCI work.
Establishes the required sequence for CI commissioning and troubleshooting.
“Always verify MSD performance in EI before switching to CI operation.”
An antistatic wrist strap must fit snugly to be effective.
Ensures ESD protection is actually effective when handling boards and the analyzer.
“To be effective, an antistatic wrist strap must fit snugly (not tight).”
Excess stress when removing the filament can crack the filament ceramic, and a defective filament must not be operated.
Warns service engineers about the handling limit during HES disassembly.
“Excessive stress can crack the filament ceramic.”