Agilent Technologies 5977C GC/MSD
Agilent Technologies 5977C GC/MSD — Gas chromatography mass spectrometry gc ms.
The 5977C GC/MSD is a single-quadrupole mass-selective detector configured here as an ethylene and trace-gas analysis system with a HydroInert ion source and hydrogen carrier gas. It quantifies permanent gases and greenhouse gases at ppb detection limits (SF6 0.63 ppb, CH4 1.2 ppb) and aromatics per ASTM D5769, distinguishing it from triple-quadrupole and Q-TOF siblings by its scan-mode single-quadrupole architecture.
Agilent Technologies GC MS Instruments Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Carrier gas | Transfer line temperature, °C | Sample volume, milliliter | Ion source temperature, °C | Analysis time, min |
|---|---|---|---|---|---|---|
| 5977C GC/MSD Current | 1 | Hydrogen | 230 | 3×10⁻⁴ | 230 | 7.5 |
| 7000E GC/TQ | 6×10⁻⁵ | Helium | 300 | 0.002 | 280 | 10 |
| 7010D GCMS | — | Helium | 300 | 0.002 | 280 | — |
| Quadrupole Time-of-Flight GC/MS, Mass Screening GC/Q-TOF | — | 1.5 | 280 | — | — | — |
Full specifications
| Pumping speed | 1 flow |
| Analysis time | 7.5 min |
| Carrier gas | Hydrogen |
| Column Film Thickness | 0.4 µm |
| Column Flow | 0.9 mL/min |
| Column temperature | 250 °C |
| Foreline Reading | 34.7 mTorr |
| Gas Flow | 2 mL/min |
| Ion source | Agilent HydroInert source |
| Ion Vacuum Gauge Reading | 3.18 Torr |
| Maximum Recommended Gas Flow | 1.5 mL/min |
| Optimal He Column Flow | 1 mL/min |
| Optional Gauge Reading | 2.18 Torr |
| Resolution | 4.77 milligram |
| Sample volume | 0.3 µL |
| Scan speed | 2 |
| Septum Purge | 3 mL/min |
| Septum Purge Flow | 2 mL/min |
| Total Run Time | nine minutes |
| Acquisition mode mfr | Scan Scan |
| Gain factor mfr | 0.5 |
| Chromatographic separation time mfr | 7.5 minutes |
| Detection limit (MDL), CO2 mfr | 2.5 ppb |
| Detection limit (MDL), N2O mfr | 2.0 ppb |
| Detection limit (MDL), C2H4 mfr | 2.1 ppb |
| Detection limit (MDL), SF6 mfr | 0.63 ppb |
| Detection limit (MDL), CH4 mfr | 1.2 ppb |
| Detection limit (MDL), CO mfr | 4.1 ppb |
| Linearity R2, CO2 mfr | 0.99934554 |
| Linearity R2, N2O mfr | 0.99980122 |
| Linearity R2, C2H4 mfr | 0.99998106 |
| Linearity R2, SF6 mfr | 0.99976908 |
| Linearity R2, CH4 mfr | 0.99981775 |
| Linearity R2, CO mfr | 0.99988739 |
| Inlet mode mfr | Multimode inlet , constant flow flow |
| Inlet liner mfr | 4 mm |
| Equilibrium time mfr | 1 min |
| Ionization mode mfr | 70 eV |
| Number of aromatic species quantified mfr | 24 species |
| Linearity (R2) for toluene mfr | 0.99996 |
| Extraction lens diameter mfr | 9 mm |
| Inlet septum mfr | Advanced Green, non-stick stick |
| Benzene average content mfr | 1.99 % mass |
| Benzene %RSD mfr | 0.37 |
| Toluene average content mfr | 15.17 % mass |
| Toluene %RSD mfr | 1.04 |
| Ethylbenzene average content mfr | 3.54 % mass |
| Ethylbenzene %RSD mfr | 0.49 |
| p-Xylene/m-Xylene average content mfr | 8.51 % mass |
| p-Xylene/m-Xylene %RSD mfr | 0.42 |
| o-Xylene average content mfr | 4.37 % mass |
| o-Xylene %RSD mfr | 0.48 |
| Total aromatics (vol %) mfr | 14.46 |
| Total aromatics (vol %) reference mfr | 14.82 |
| Benzene (vol %) mfr | 0.47 |
| Benzene (vol %) reference mfr | 0.50 |
| Toluene (vol %) mfr | 2.74 |
| Max column id mfr | 0.25 mm |
| CI capability mfr | No No |
| Inert ion sources available mfr | Yes Yes |
| GC compatibility mfr | 9000 |
| Foreline pumps available mfr | Pfeiffer Duo Duo |
| DIP capability (3rd party) mfr | Yes Yes |
| Peak area RSD, CO2 mfr | 2.2 |
| Peak area RSD, N2O mfr | 2.5 |
| Peak area RSD, C2H4 mfr | 3.3 |
| Peak area RSD, SF6 mfr | 0.7 |
| Peak area RSD, CH4 mfr | 1.5 |
| Peak area RSD, CO mfr | 4.9 |
| Linearity (R2), toluene mfr | 0.99996 |
| Linearity (R2), benzene mfr | 0.99998 |
| Linearity (R2), o-xylene mfr | 1.00000 |
| Total aromatics, measured mfr | 14.82 |
| Total aromatics, reference mfr | 14.46 |
| Benzene, measured mfr | 0.50 |
| Benzene, reference mfr | 0.47 |
| Toluene, measured mfr | 2.74 |
| Benzene, average mfr | 1.00 % m/m |
| Toluene, average mfr | 9.00 % m/m |
| Ethylbenzene, average mfr | 3.00 % m/m |
| p-Xylene/m-Xylene, average mfr | 2.98 % m/m |
| o-Xylene, average mfr | 2.98 % m/m |
| 1,2,4-Trimethylbenzene, average mfr | 2.95 % m/m |
| 1,2,4,5-Tetramethylbenzene, average mfr | 3.00 % m/m |
| Benzene, %RSD mfr | 0.37 |
| Toluene, %RSD mfr | 1.04 |
| High vacuum pump mfr | Diffusion Diffusion |
| Instrument mfr | 8890 GC |
| Column Oven Ramp Program mfr | 40 min |
| Carrier gas purity mfr | 99.9995 |
| Ionization energy mfr | 70 eV |
| Group 1 mfr | 4.25 |
| Group 2 mfr | 5.60 |
| Group 3 mfr | 8.00 |
| Group 4 mfr | 8.80 |
| Group 5 mfr | 10.45 |
| Group 6 mfr | 12.00 |
| Group 7 mfr | 13.70 |
| PCB 118 detection limit mfr | 10 ppb |
| Sampler mfr | 7650 Sampler |
| Hydrogen Leak Sensor mfr | 2 GC |
| Extractor Lens mfr | 9 mm |
| Acquisition Mode mfr | Scan Scan |
| Inlet Septum mfr | Advanced Green, non-stick stick |
| Inlet Liner mfr | Deactivated, low pressure drop split liner w/ gl |
| ALS Syringe mfr | 5 cone |
| Carrier Gas Filter mfr | 1 in |
| Runtime mfr | 9 min |
| Mode mfr | 700 |
| Gas Saver mfr | 20 minutes |
| Initial Hold mfr | 1 min |
| Ramp Rate mfr | 25 °C/min |
| Final Hold mfr | 1 min |
| Solvent Delay mfr | 2 min |
| Scan Range mfr | 40...200 amu |
| Tuning Algorithm mfr | atune atune |
| Gain Factor mfr | 0.3 |
| Trace Ion Detection (TID) mfr | Enabled Enabled |
| Extraction Lens Diameter mfr | 9 mm |
| Toluene Calibration Concentration mfr | 16.3 |
| 1,4-Diethylbenzene Concentration mfr | 100 ppm |
| Method Resolution Requirement mfr | 2 |
| Total Runtime mfr | nine minutes minutes |
| CI capability‡ mfr | No No |
| DIP** capability (3rd party) mfr | Yes Yes |
| Methane/PCI mfr | 17 |
| Methane/NCI mfr | 17 |
| Isobutane/PCI mfr | 39 |
| Isobutane/NCI mfr | 17 |
| Ammonia/PCI mfr | 18 |
| Ammonia/NCI mfr | 17 |
| Not shown mfr | HES assembly assembly |
| Intuvo column conditioning ramp rate mfr | 10...15 °C/min |
| Intuvo column conditioning hold time mfr | 30 minutes |
| NCI methane flow mfr | 40 |
| NCI OFN injection volume mfr | 2 μL |
| NCI OFN concentration mfr | 100 fg/μL |
| Column Inner Diameter | 0.18 mm |
| Column Length | 40 m |
| Foreline Pump Weight | 11 kg |
| Weight | 39 kg |
| Housing height mfr | 41 cm |
| Housing width mfr | 30 cm |
| Housing depth mfr | 54 cm |
| Total system footprint mfr | 2 systems |
| GC footprint mfr | 8890 System |
| MSD housing height mfr | 41 cm |
| MSD housing width mfr | 30 cm |
| MSD housing depth mfr | 54 cm |
| Diffusion pump mainframe weight mfr | 39 kg |
| EI/CI turbo pump mainframe weight mfr | 46 kg |
| Primary fuse rating mfr | 12.5 V |
| Ionization Voltage mfr | 70 eV |
| Inlet Temperature | 300 °C |
| Ion Source Temperature | 230 °C |
| Quadrupole Temperature | 150 °C |
| Quad Temperature | 150 °C |
| Sound pressure | 70 dB |
| Sound Pressure Lp | 70 dB |
| Source Temperature | 250 °C |
| Transfer Line Temperature | 230 °C |
| Valve Oven Temperature | 50 °C |
| PCB 118 signal-to-noise ratio mfr | 3 |
| Standard EI turbo pump mainframe weight mfr | 44 kg |
Description
Construction and positioning in the series
The 5977C GC/MSD is the single-quadrupole member of Agilent's GC/MSD line. The sibling models listed — 7000E and 7010D triple quadrupoles and the 7250 GC/Q-TOF — are tandem and high-resolution instruments; the 5977C is the scan-mode single-quadrupole platform used where unit-mass resolution with full-scan acquisition is sufficient. This configuration pairs the MSD with a multimode inlet operated in constant-flow mode, a 4 mm inlet liner, a 20 µm column, and the HydroInert ion source, which is specified for hydrogen carrier gas service.
Behavior in the configured method
- Carrier gas: hydrogen; column flow 0.9 mL/min; septum purge 2 mL/min.
- Thermal setpoints: inlet 300 °C, ion source 230 °C, quadrupole 150 °C, GC transfer line 300 °C, valve oven 50 °C; oven program starts at 50 °C with 1 min equilibration.
- Ionization: 70 eV, scan acquisition, gain factor 0.5, 9 mm extraction lens.
- Chromatography: 7.5 min separation within a nine-minute total run.
- Detection limits (MDL): SF6 0.63 ppb, CH4 1.2 ppb, N2O 2.0 ppb, C2H4 2.1 ppb, CO2 2.5 ppb, CO 4.1 ppb.
- Linearity R² across CO2, N2O, C2H4, SF6, CH4 and CO ranges from 0.99934 to 0.99998.
Application coverage from the attached documents
The document set covers three distinct uses of the same MSD platform: trace permanent and greenhouse gases with the high-efficiency source (10 species including H2, O2, N2, Ar), PCB determination by US EPA Method 1628 with the HydroInert source and a DB-XLB column on an 8890 GC, and aromatic quantification in gasoline per ASTM D5769 on an 8850 GC with the optional 9 mm extraction lens. In the aromatics method, 24 species are quantified; benzene averages 1.99 % mass at 0.37 %RSD and toluene 15.17 % mass at 1.04 %RSD, with toluene linearity R² 0.99996. These are method results, not instrument ratings, and depend on the column, inlet hardware and carrier gas named in each document.
Control and operation
The digests describe method parameters and results only; they contain no installation, connection or control-section content, so wiring, communication interfaces and startup sequencing cannot be stated from the verified facts.
Integration & Operating Notes
The configured ion source is the Agilent HydroInert source, and the method facts specify hydrogen as carrier gas. Method performance figures (MDLs, linearity) were obtained under that carrier/source pairing; substituting helium or another carrier without re-validating the method invalidates the quoted detection limits.
Inlet and transfer line are held at 300 °C, ion source at 230 °C and quadrupole at 150 °C in the configured method. These are operating setpoints for the described analysis; the verified facts give no independent maximum ratings, so do not treat them as absolute limits when developing other methods.
The method specifies a 4 mm inlet liner, a 20 µm column and an Advanced Green non-stick septum, and the aromatics method uses an optional 9 mm extraction lens. These are method-specific consumables and hardware; confirm each against the ordered configuration, since the verified facts do not establish them as included accessories.
Detection limits from 0.63 ppb (SF6) to 4.1 ppb (CO) and linearity R² values were generated with the specified column flow, oven program, 70 eV ionization and scan/gain settings. Any change to carrier, column dimensions, inlet mode or acquisition parameters requires re-establishing these figures.
Selection & Integration Q&A
Pick the 5977C when unit-mass full-scan data at ppb levels suffice, as in the trace-gas method (SF6 0.63 ppb, CH4 1.2 ppb) and ASTM D5769 aromatics. The triple quadrupoles and Q-TOF siblings target MRM trace quantitation and high-resolution exact-mass work, which this scan-mode configuration does not provide.
The method pairs the MSD with a multimode inlet in constant-flow mode at 0.9 mL/min hydrogen, a 4 mm inlet liner, a 20 µm column, an Advanced Green non-stick septum and, for ASTM D5769 aromatics, the optional 9 mm extraction lens. The HydroInert source is specified for hydrogen carrier; confirm column and liner part numbers against the ordered configuration.
The 4 mm inlet liner, 20 µm analytical column, Advanced Green non-stick septum and the optional 9 mm extraction lens are method-specific items named in the application documents rather than confirmed base-delivery contents. The verified facts contain no maintenance-interval or spare-parts schedule, so service intervals must be taken from the instrument manual.
Dimensional Drawings & Technical Documentation
Dimensional drawings
Click drawing to inspect vector dimensions