Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Contents

  1. Overview p. 2
    Background on PFAS compounds and analytical challenge
  2. Procedure p. 3
    Sample collection and preparation workflow
  3. p. 4
    Analytical results and interpretation of findings
  4. Spec p. 5
    Detection limits per compound and method
  5. p. 6
    Analysis results from wastewater effluent samples
  6. p. 7
    Summary of findings and method suitability
  7. pp. 8-10
    Supplementary compound group data tables
  8. p. 11
    Cited literature and sources
  9. p. 4 table, 11 rows
  10. p. 5 table, 40 rows
  11. p. 7 table, 40 rows
  12. p. 8 table, 30 rows
  13. p. 9 table, 40 rows
  14. p. 10 table, 40 rows

Table 1. LC instrument conditions.

ParameterValue
LCAgilent 1290 Infinity II LC with G7120A High Speed Pump (G7120A), Multisampler (G7167B), Multicolumn Thermostat column compartment (G7116B)
UHPLC Guard ColumnAgilent ZORBAX Eclipse Plus C18, 2.1 × 5 mm, 1.8 μm (part number 821725-901)
Analytical ColumnAgilent ZORBAX Eclipse Plus C18, 2.1 × 100 mm, 1.8 μm (part number 959758-902)
Delay ColumnAgilent InfinityLab PFC Delay Column, 4.6 × 30 mm (part number 5062-8100)
Column Temperature40 °C
Injection Volume5 µL
Mobile PhaseA) 2 mM Ammonium acetate in 95% water (LC grade), 5% acetonitrile B) Acetonitrile (LC grade)
Gradient Flow Rate0.4 mL/min
GradientTime (min) % B 0.0 2 0.2 2 10 95
Stop Time12.2 min
Post Time2.0 min

Table 3. Initial method detection limits achieved with the Agilent Bond Elut PFAS WAX SPE extraction of low-level spikes and lab reagent blanks (LRB).

AnalyteBond Elut PFAS WAX Initial MDL (ng/L)EPA 1633 Aq. MDL (ng/L)*
PFBA0.380.79
PFPeA0.710.54
PFHxA0.090.46
PFHpA0.190.37
PFOA0.230.54
PFNA0.150.45
PFDA0.150.52
PFUnA0.130.45
PFDoA0.290.40
PFTrDA0.330.46
PFTeDA0.380.49
PFBS0.120.37
PFPeS0.150.50
PFHxS0.260.54
PFHpS0.210.50
PFOS0.390.63
PFNS0.150.47
PFDS0.230.60
PFDoS0.440.60
4:2FTS0.431.69
6:2FTS0.582.45
8:2FTS0.842.50
PFOSA0.200.32
NMeFOSA0.090.43
NEtFOSA0.170.45
NMeFOSAA0.270.68
NEtFOSAA0.340.59
NMeFOSE1.413.81
NEtFOSE1.974.84
HFPO-DA0.600.51
ADONA0.260.50
PFMPA0.361.46
PFMBA0.131.41
NFDHA0.380.75
9Cl-PF3ONS0.521.38
11Cl-PF3OUdS1.381.67
PFEESA0.261.17
3:3 FTCA0.822.47
5:3 FTCA3.779.59
7:3 FTCA3.278.71

* EPA Method 1633 — Table 9, multi-lab validation results.

Table 4. Spiked concentration, average recovery (accuracy), and range of recoveries of PFAS in wastewater samples.

AnalyteSpiked Concentration (ng/L)Average Accuracy (Recovery)Low Average Recovery RangeHigh Average Recovery Range
PFBA5091%86%99%
PFPeA2588%84%95%
PFHxA12.593%85%103%
PFHpA12.597%91%104%
PFOA12.596%87%107%
PFNA12.594%85%101%
PFDA12.596%90%103%
PFUnA12.595%90%101%
PFDoA12.592%87%95%
PFTrDA12.590%85%96%
PFTeDA12.596%88%103%
PFBS12.5104%92%114%
PFPeS12.592%86%98%
PFHxS12.591%88%94%
PFHpS12.599%90%105%
PFOS12.597%84%105%
PFNS12.582%76%87%
PFDS12.570%65%76%
PFDoS12.548%45%50%
4:2FTS5095%90%104%
6:2FTS5095%92%99%
8:2FTS5089%77%105%
PFOSA12.596%88%103%
NMeFOSA12.589%83%93%
NEtFOSA12.592%81%99%
NMeFOSAA12.5105%100%115%
NEtFOSAA12.596%92%103%
NMeFOSE12592%86%101%
NEtFOSE12593%89%98%
HFPO-DA50109%90%120%
ADONA50111%91%122%
PFMPA25103%95%111%
PFMBA25102%93%106%
NFDHA2548%39%54%
9Cl-PF3ONS50100%82%109%
11Cl-PF3OUdS5074%61%84%
PFEESA2596%93%101%
3:3 FTCA50114%97%127%
5:3 FTCA250153%136%169%
7:3 FTCA250141%117%156%

dMRM transitions monitored

Compound GroupCompound NamePrecursor IonProduct IonRT (min)FragmentorCollision EnergyCell Accelerator Voltage
AcidPFBA213168.93.56082
AcidPFPeA2632194.57242
AcidPFHxA313268.95.27082
AcidPFHxA3131195.270182
AcidPFHpA362.93195.77242
AcidPFHpA362.91695.772142
AcidPFOA4133696.26942
AcidPFOA4131696.269122
AcidPFNA4634196.66642
AcidPFNA4632196.666172
AcidPFDA5134697.072122
AcidPFDA5132197.072202
AcidPFUnA5635197.5100122
AcidPFUnA5632697.5100202
AcidPFDoA6135697.910082
AcidPFDoA6133197.9100202
AcidPFTrDA6636198.2100122
AcidPFTrDA6631698.2100322
AcidPFTeDA712.96698.6100122
AcidPFTeDA712.91698.6100322
Ether Sulfonic AcidsPFEESA314.9134.95.6110242
Ether Sulfonic AcidsPFEESA314.9835.6124205
Ether Sulfonic AcidsPFEESA314.9695.6110602
Ether Sulfonic Acids9Cl-PF3ONS530.9350.97.6145282
Ether Sulfonic Acids9Cl-PF3ONS530.9837.6145322
Ether Sulfonic Acids11Cl-PF3OUdS630.9838.4160322
Ether Sulfonic Acids11Cl-PF3OUdS630.9450.98.4165322
Flurotelomer Carboxylic Acid3-3 FTCA2411774.27443
Flurotelomer Carboxylic Acid3-3 FTCA2411174.274443
Flurotelomer Carboxylic Acid5-3 FTCA3412375.684123

Compound Group, Compound Name, Precursor Ion, Product Ion, RT (min), Fragmentor, Collision Energy, Cell Accelerator Voltage

Compound GroupCompound NamePrecursor IonProduct IonRT (min)FragmentorCollision EnergyCell Accelerator Voltage
Flurotelomer Carboxylic Acid5-3 FTCA3412175.684243
Flurotelomer Carboxylic Acid7-3 FTCA4413376.676123
Flurotelomer Carboxylic Acid7-3 FTCA4413176.676243
FTS4:2FTS327306.95.0125202
FTS4:2FTS32780.95.0125362
FTS6:2FTS427406.86.0125242
FTS6:2FTS42780.96.0125402
FTS8:2FTS5275076.8200304
FTS8:2FTS52780.96.8170402
ISTD13C3-PFBA216171.93.56582
ISTD13C2-PFHxA3152705.27082
ISTD13C4-PFOA4171726.269122
ISTD18O2-PFHxS40383.96.4100492
ISTD13C5-PFNA4684236.66642
ISTD13C2-PFDA5154707.08142
ISTD13C4-PFOS503807.3148542
ISTD13C4-PFOS502.998.97.3180482
Perfluorooctane Sulfonamide EthanolsNMeFOSE616599.882154
Perfluorooctane Sulfonamide EthanolsNEtFOSE6305910.1124454
Perfluorooctane SulfonamidesPFOSA497.94788.6150363
Perfluorooctane SulfonamidesPFOSA497.9788.6150363
Perfluorooctane SulfonamidesPFOSA497.9488.61501103
Perfluorooctane SulfonamidesNMeFOSA5122199.9156285
Perfluorooctane SulfonamidesNMeFOSA5121699.9156325
Perfluorooctane SulfonamidesNEtFOSA52626910.3160285
Perfluorooctane SulfonamidesNEtFOSA52621910.3160285
Perfluorooctane SulfonamidesNEtFOSA52616910.3160285
Perfluorooctane Sulfonamidoacetic AcidsNMeFOSAA570482.97.1150162
Perfluorooctane Sulfonamidoacetic AcidsNMeFOSAA5704197.1150202
Perfluorooctane Sulfonamidoacetic AcidsNEtFOSAA5845267.3100202
Perfluorooctane Sulfonamidoacetic AcidsNEtFOSAA5844197.3100202
Polyfluoroether Carboxylic AcidsPFMPA22984.94.060122
Polyfluoroether Carboxylic AcidsPFMBA2792354.88015
Polyfluoroether Carboxylic AcidsPFMBA27984.94.870122
Polyfluoroether Carboxylic AcidsNFDHA2952015.17552
Polyfluoroether Carboxylic AcidsNFDHA295855.1120325
Polyfluoroether Carboxylic AcidsNFDHA201855.170155
Polyfluoroether Carboxylic AcidsHFPO-DA2851855.450205
Polyfluoroether Carboxylic AcidsHFPO-DA2851695.45045
Polyfluoroether Carboxylic AcidsADONA377250.95.980122

Compound Group / Compound Name / Precursor Ion / Product Ion / RT (min) / Fragmentor / Collision Energy / Cell Accelerator Voltage

Compound GroupCompound NamePrecursor IonProduct IonRT (min)FragmentorCollision EnergyCell Accelerator Voltage
SulfonatePFHxS398.9806.4100492
SulfonatePFHpS448.998.76.9100442
SulfonatePFHpS448.9806.9148502
SulfonatePFOS498.9997.3100502
SulfonatePFOS498.9807.3100502
SulfonatePFNS548.9997.7148522
SulfonatePFNS548.9807.7148562
SulfonatePFDS598.9998.1148562
SulfonatePFDS598.9808.1148602
SulfonatePFDoS698.9998.8156622
SulfonatePFDoS698.9808.8156672
Surrogate13C4-PFBA2171723.56082
Surrogate13C5-PFPeA2682234.56082
Surrogate13C2-4:2FTS3293095.0125202
Surrogate13C2-4:2FTS329815.0150322
Surrogate13C5-PFHxA3182735.27082
Surrogate13C5-PFHxA3181205.272242
Surrogate13C3-PFBS302995.3130322
Surrogate13C3-PFBS302805.3130442
Surrogate13C3-HFPO-DA2871855.464205
Surrogate13C3-HFPO-DA2871695.46445
Surrogate13C4-PFHpA3673225.77242
Surrogate13C4-PFHpA3671695.772162
Surrogate13C2-6:2FTS4294096.0125242
Surrogate13C2-6:2FTS429816.0150402
Surrogate13C8-PFOA4213766.26942
Surrogate13C8-PFOA4211726.272202
Surrogate13C3-PFHxS402996.4156442
Surrogate13C3-PFHxS402806.4100452
Surrogate13C9-PFNA4724276.66642
Surrogate13C9-PFNA4722236.672162
Surrogate13C2-8:2FTS5295096.8170282
Surrogate13C2-8:2FTS529816.8200524
Surrogate13C6-PFDA5194747.07282
SurrogateD3-NMeFOSAA573.24197.1150202
SurrogateD5-NEtFOSAA589.24197.2100202
Surrogate13C8-PFOS507997.3148522
Surrogate13C8-PFOS507807.3100502
Surrogate13C7-PFUnA5705257.510082
Surrogate13C2-PFDoA6155707.990122

Engineering Facts

13 facts · every claim linked to a page
Selection

The method requires weak anion exchange (WAX) SPE and carbon cleanup for all matrices, so the cartridge chemistry must be WAX and a carbon cleanup step must be included.

Determines the SPE sorbent chemistry and the need for a carbon cleanup consumable when selecting the extraction workflow.

Source
“all the methods within EPA Method 1633 specify SPE using weak anion exchange (WAX) and carbon cleanup.”
Selection

The 150 mg, 6 mL Bond Elut PFAS WAX cartridge is the format used for aqueous matrices in this method, so this bed mass and volume should be selected for 500 mL aqueous samples.

Specifies the exact cartridge size to choose for method-compliant aqueous PFAS extraction.

Source
“using a 150 mg, 6 mL Agilent Bond Elut PFAS WAX cartridge for SPE and Carbon S for cleanup.”
Selection

Bond Elut PFAS WAX cartridges are PFAS-specific and undergo PFAS-specific QC testing, so they should be chosen over general-purpose WAX cartridges to control background contamination.

Justifies selecting PFAS-certified cartridges rather than generic WAX to avoid false positives and poor recovery.

Source
“The cartridges undergo PFAS-specific QC testing to ensure optimal cleanliness and recovery.”
Installation

A PFAS-specific delay column must be installed between the pump and the multisampler to separate background contaminants without significantly increasing backpressure.

Defines the required LC plumbing position of the delay column to keep PFAS background low without overpressuring the system.

Source
“The PFAS-specific delay column was placed between the pump and multisampler to separate background contaminants from compounds originating in the sample vial without significantly increasing backpressure.”
Operation

Samples must be loaded onto the SPE cartridge at a slow flow rate of 5 mL/min, so the loading step must be flow-controlled rather than run at full vacuum.

Sets the maximum sample loading flow to preserve analyte recovery during extraction.

Source
“Passthrough cartridge at 5 mL/min.”
Operation

The SPE cartridge must be conditioned with 15 mL of 1% methanolic ammonium hydroxide followed by 5 mL of 0.3 M formic acid before loading the sample.

Defines the conditioning sequence and volumes needed to activate the WAX sorbent correctly.

Source
“15 mL of 1% methanolic ammonium hydroxide”
Operation

After elution, 25 µL of concentrated acetic acid must be added to each eluate and vortexed before Carbon S cleanup.

Specifies the acidification step required before carbon cleanup to achieve the correct recovery.

Source
“Add 25 µL of concentrated acetic acid to each sample eluate and vortex.”
Operation

Carbon S cleanup uses 10 mg per sample with hand-shaking for less than 5 minutes, vortexing for 30 seconds, and centrifugation for 10 minutes at 2,800 rpm.

Gives the exact carbon dose and mixing/centrifugation conditions needed for reproducible matrix removal.

Source
“Add 10 mg of Carbon S to each sample.”
Operation

The final extract must be filtered through a nylon syringe filter into a collection tube containing the nonextracted internal standard (NIS).

Ensures the NIS is added at the correct point and that particulates are removed before LC/TQ analysis.

Source
“Filter the entire extract into a NIS collection tube and vortex.”
Performance

The chromatographic separation between TDCA and PFOS must be at least 1 minute; the conditions used achieved 5.7 and 7.1 minutes respectively.

Sets a mandatory chromatographic resolution criterion that the LC method and column must satisfy.

Source
“EPA Method 1633 specifies that the chromatographic separation between the bile salts (TDCA) and PFOS must be at least 1 minute.”
Performance

Initial MDLs with Bond Elut PFAS WAX SPE were equivalent to or better than the method-documented MDLs for most analytes, supporting method-compliant sensitivity.

Confirms the cartridge can meet required detection limits, which is a key selection criterion for trace PFAS work.

Source
“Initial MDLs achieved following extraction of low-level spikes and lab reagent blanks (LRB) with the Bond Elut PFAS WAX SPE are equivalent or better than those documented in the method.”
Performance

In wastewater effluent, the FTS class of extracted surrogates had consistently high recoveries, showing that matched labeled internal standards are needed to correct matrix effects.

Warns that matrix effects can bias FTS results unless appropriate isotope-labeled internal standards are used.

Source
“However, the FTS class of PFAS extracted surrogates had consistently high recoveries in matrix, demonstrating the importance of matched and labeled internal standards to accurately correct for matrix effects.”
Compliance

The method was followed closely with only permitted changes in supplies and instrumentation parameters, so any deviation in consumables or instrument settings must be within the method's allowed changes.

Reminds the engineer that consumable and instrument substitutions must remain within EPA Method 1633 allowances for compliance.

Source
“The method was followed closely; only permitted changes in supplies and instrumentation parameters were made.”