LM8A-Automatic-Microtome-User-Manual
Technical Reference Summary
Covered Models & Document Scope
This instruction manual covers the Avantik LM8A fully automatic rotary microtome, model LM8A. No other variants or ordering codes are described.
Section-by-Section Technical Breakdown
Lists electrical, environmental, dimensional, and microtome performance specifications for the LM8A. Includes power supply, fuse rating, section and trimming thickness ranges with resolutions, feed range, stroke, operating modes, and cutting speed.
- Power supply 110…240 VAC, 50…60 Hz, mains fuse 2 x T 2A
- Storage temperature -20°C to +50°C; operating conditions +5°C to +40°C
- Dimensions 430 mm W x 530 mm D x 310 mm H; weight 38 kg
- Section thickness 1.0–50 µm with 0.25/0.5/1.0 µm steps depending on range
- Trimming thickness 1.0–500.0 µm with 1/5/10 µm steps depending on range
- Horizontal feed range 30 mm by stepping motor; vertical specimen stroke 68 mm
- Modes: interval, single, multi, continuous stroke; cutting speed 0.5–400 mm/s
Identifies front and rear panel components, film keyboard keys, main display indicators, knife carrier parts, and object clamp controls. Provides the physical layout for operation and adjustment.
- Front: specimen, film keyboard, emergency stop, main switch, handwheel brake, handwheel, knife carrier, specimen clamping system, section waste tray
- Rear: connector for foot switch, USB communication, fan, SD card slot, RS232 communication
- Film keyboard includes menu, reset counter, retraction on/off, memory keys M1–M3, brake, start/stop, cutting speed +/-, crop window, object feed vertical +/-
- Main display shows trim, cut size, motorized cutting speed, AVR, mode, section counter, crop window on/off, retraction on/off, trim thickness, position of object feed
- Knife carrier has levers for fixing/loosening knife, blade, blade carrier, angle, and finger guard
- Object clamp has cassette clamp, lever for specimen, angle adjustment screws, and lever for cassette
Describes prerequisites for startup, including acclimatization time and stable installation surface. Covers mounting and dismounting precautions such as removing the blade and activating the handwheel brake.
- Leave unit at workspace for at least 2 hours to adjust to room temperature before operation
- Install on a stable, level, horizontally oriented base
- Use carry handles for transport
- Before dismounting, remove the microtome blade and activate the handwheel brake
Explains power-on sequence, menu navigation via turn-push button, and menu structure including mode, AVR, language, and display settings. Lists default display contrast, color, and brightness values.
- Set on/off switch to '0', plug in power cord, then switch to 'I'; ready when film keyboard display lights
- Menu accessed by menu button; navigate by turning and pressing turn-push button on left side
- Menu items: Mode, AVR, Language, Display (Contrast, Color, Brightness)
- Languages: English, German, French, Italian, Spanish; default English
- Default display settings: contrast 10, color 50, brightness 20
Details object orientation adjustment, specimen insertion and removal, and cassette clamp replacement. Specifies angular adjustment range and lever/screw functions.
- Cassette clamp angle adjustable to 8° in all directions and rotatable 360°
- Screw (1) vertical adjustment, screw (2) horizontal adjustment; lever (3) releases/fixes clamp
- Always release lever (3) before adjusting angle to avoid damage
- Keep cassette contact surfaces free from paraffin for optimal clamping
- To insert specimen, pull lever (1) toward user and insert into cassette clamp (2)
- Before removing cassette clamp, turn it to center position; ensure adapter hole aligns with object clamp hole on reinstallation
Covers insertion, adjustment, and changing of the microtome blade, knife carrier positioning, blade holder lateral adjustment, and knife carrier angle setting. Includes safety notes on finger guard and handwheel brake.
- Turn finger guard to front and press lever (1) forward to open blade insertion gap
- Insert blade carefully in gap (2), then press lever (1) back to fix
- Knife carrier can be moved forward/backward on rails or removed; pull up lever (1) to release, pull down to fix
- Blade holder moves left/right; pull down lever (1) to release, pull up to fix
- Knife carrier angle adjustable from 6 to 12 degrees using scale on left side; pull down lever (1) to release, pull up to fix
- Use cleaning brush with integrated magnet to remove blade
- After every operation or non-use, put finger guard over blade and activate handwheel brake
Describes operation of the handwheel brake lever and precautions to avoid damage. The brake fixes the handwheel for safety during adjustments.
- Tip lever from 'open' to 'close' position; lever moves automatically to 'close'
- Avoid overwinding the lever to the mechanical end position
- Turn down lever to release handwheel
- Do not rotate handwheel while brake is activated; device can be damaged
Explains setting section and trimming thickness, coarse feed speeds, trimming functions, and fine feed. Provides resolution tables for section and trimming thickness ranges.
- Turn pushbutton to adjust fine/trimming thickness; press to switch between 'Fine' and 'Trim' modes
- Specimen clamp position settable between 0 µm and 30000 µm
- Cutting speed adjustable in one-steps up to 10, then five-steps after 10
- Section thickness resolution: up to 5 µm in 0.25 µm steps; 5–10 µm in 0.5 µm steps; 10–50 µm in 1 µm steps
- Trimming thickness resolution: up to 20 µm in 1 µm steps; 20–100 µm in 5 µm steps; 100–500 µm in 10 µm steps
- Coarse feed forward speeds: speed 1 0–2.4 s, speed 2 2.4–3.2 s, speed 3 from 3.2 s
- Coarse feed reverse speeds: speed 1 0–0.8 s, speed 2 0.8–1.6 s, speed 3 from 1.6 s
- Brief press of coarse feed buttons moves object head 5 µm forward or backward
- Trim button has normal trimming and duration trim functions
Describes motorized sectioning including cutting window setup, operating modes, foot switch control, cutting speed setting, handwheel handle folding, start/stop, and brake interaction.
- Cutting window set by positioning lower edge of specimen about 3 mm above knife edge, then setting Start and Stop points
- Activated cutting window marked with 'C' in upper right of display
- Operating modes: tip stroke, single stroke, continuous stroke, multi stroke
- Multi stroke: number of sections 1–100 settable under 'Multi cut' menu
- Foot switch connected via socket on back; can start/stop motorized sectioning
- Press foot switch strongly to activate emergency stop; 'STOP' shown on display
- Cutting speed set via buttons, displayed 0–100; controlled speed 0–400 mm/s
- AVR (automatic velocity reduction) can be set to 'On' to reduce speed during fine cutting
- Fold handwheel handle down before motorized cutting for safety
- Motorized cutting drive will not function if brake or emergency stop is applied
Covers section counter, automatic retraction, position memory, and emergency stop functions. Includes display indicators and reset behavior.
- Section counter indicates cuts since power-on; reset by pressing section counter button; counter cleared after each restart
- Automatic retraction pulls specimen back during upward movement; activated by retraction button; 'R' appears in display when active
- Position memory stores 3 trimming positions via M1, M2, M3 keys; press >1 second until signal; values deleted after restart
- Emergency stop button on left side above operating panel; display shows 'STOP' when activated
- Handwheel brake cannot be activated when instrument is turned off
Provides a troubleshooting table for common problems and their remedies, and instructions for replacing the mains fuse. Includes safety steps for power disconnection.
- Troubleshooting table lists problems: knife carrier cannot be fixed, cutting movement too hard, piston cannot be driven, device does not turn on, screen does not turn on, keypad buttons do not respond
- Fuse replacement: switch off device, set on/off switch to '0', remove power cord, squeeze pins (1), pull out drawer (2), replace fuse (3), slide tray back until click
- Fuse data in chapter 2.2 Technical Data: 2 x T 2A
Specifies cleaning agents and procedures, maintenance intervals, and recycling guidance. Emphasizes daily cleaning and regular service checks.
- Clean specimen clamp, knife carrier, section waste tray, and visible guides after each use
- Use standardized non-scratching cleaning agents; use brush for sticking deposits; do not use sharp-edged blades or tools
- Switch off device and pull power cord before cleaning
- Daily: clean microtome from section and wax remains
- Weekly: check all guiding, locking, and clamping levers for perfect operation
- Every 12 months: contact Avantik service for maintenance of hidden parts
- At end of life, transfer unit to professional disposal or return to Avantik
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Power supply 110…240 VAC, 50…60 Hz, mains fuse 2 x T 2A p.8 ↗ | Ensure supply voltage and frequency are within range and use correct fuse rating for protection. | LM8A |
| Operating ambient temperature +5°C to +40°C; storage temperature -20°C to +50°C p.8 ↗ | Install in a controlled environment within these limits; allow acclimatization before operation. | LM8A |
| Section thickness range 1.0–50 µm with resolution 0.25 µm (up to 5 µm), 0.5 µm (5–10 µm), 1.0 µm (10–50 µm) p.8 ↗ | Select section thickness within range; resolution depends on selected range. | LM8A |
| Trimming thickness range 1.0–500.0 µm with resolution 1 µm (up to 20 µm), 5 µm (20–100 µm), 10 µm (100–500 µm) p.8 ↗ | Trimming thickness can be set up to 500 µm; resolution coarsens at higher values. | LM8A |
| Horizontal feed range 30 mm by stepping motor; vertical specimen stroke 68 mm p.8 ↗ | Specimen travel limits for positioning and cutting window setup. | LM8A |
| Cutting speed 0.5–400 mm/s, continuously adjustable p.8 ↗ | Motorized cutting speed can be set within this range for different sectioning needs. | LM8A |
| Cassette clamp angle adjustable to 8° in all directions and rotatable 360° p.15 ↗ | Allows precise specimen orientation; always release lever (3) before adjusting angle to avoid damage. | LM8A |
| Knife carrier angle adjustable from 6 to 12 degrees p.19 ↗ | Set clearance angle within this range for optimal cutting; use scale on left side. | LM8A |
| Specimen clamp position settable between 0 µm and 30000 µm p.21 ↗ | Object feed position can be adjusted over this range for coarse positioning. | LM8A |
| Coarse feed forward speeds: speed 1 0–2.4 s, speed 2 2.4–3.2 s, speed 3 from 3.2 s; reverse speeds: speed 1 0–0.8 s, speed 2 0.8–1.6 s, speed 3 from 1.6 s p.22 ↗ | Press duration determines coarse feed speed; brief press moves 5 µm. | LM8A |
| Multi stroke mode allows setting number of sections 1–100 under 'Multi cut' menu p.25 ↗ | Automated batch sectioning up to 100 sections with automatic stop. | LM8A |
| Cutting window: position lower edge of specimen about 3 mm above knife edge to set Start point p.24 ↗ | Ensures safe clearance before setting cutting window limits. | LM8A |
| Motorized cutting drive will not function if brake or emergency stop is applied p.26 ↗ | Release brake and reset emergency stop before starting motorized cutting. | LM8A |
| Position memory stores 3 values (M1, M2, M3); press key >1 second until signal; values deleted after restart p.28 ↗ | Use memory keys for quick repositioning; note values are not retained after power cycle. | LM8A |
| Section counter cleared after each restart; reset manually by pressing section counter button p.27 ↗ | Counter does not retain count across power cycles. | LM8A |
| Automatic retraction activated by retraction button; 'R' appears in display when active p.27 ↗ | Prevents knife and specimen damage during upward movement. | LM8A |
| Fuse replacement: switch off, set on/off to '0', remove power cord, squeeze pins (1), pull drawer (2), replace fuse (3), slide back until click p.29 ↗ | Follow procedure for safe fuse replacement; use 2 x T 2A fuses. | LM8A |
| Maintenance intervals: daily cleaning, weekly check of guiding/locking/clamping levers, every 12 months service by Avantik p.30 ↗ | Establish routine maintenance schedule to ensure reliable operation. | LM8A |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Knife Carrier cannot be fixed p.29 ↗ | Clamping lever has to be readjusted. | Readjust the clamping lever. | Chapter 5 Troubleshooting |
| Cutting movement (rotation) too hard p.29 ↗ | Guiding rolls dirty or defective. | Call Avantik service for help. | Chapter 5 Troubleshooting |
| Piston can not be driven forwards or backwards p.29 ↗ | Thrust / pullback-motor are defect. | Call Avantik service for help. | Chapter 5 Troubleshooting |
| Device does not turn on p.29 ↗ | Fuse defect. | Change the fuse (see chapter 5.1.1 – Change the Fuse). If the problem still exists, call Avantik service for help. | Chapter 5 Troubleshooting |
| Screen does not turn on p.29 ↗ | Display defect. | Call Avantik service for help. | Chapter 5 Troubleshooting |
| Keypad buttons do not respond / react incorrectly p.29 ↗ | Film tastatur defect. | Call Avantik service for help. | Chapter 5 Troubleshooting |