Avantik LM8A
Avantik ·Made in USA · In Active Production

Avantik LM8A

Avantik LM8A — Microtomes.

The Avantik LM8A is a fully automatic rotary microtome with a motorized cutting drive, a 68 mm vertical specimen stroke, and a 30 mm horizontal feed driven by a stepping motor. It is distinguished within its series by its 0.5–400 mm/s continuously adjustable cutting speed and multi-stroke batch sectioning up to 100 sections. It is intended for high-throughput production of paraffin and similar soft-to-medium hardness sections in histology and research laboratories.

Full specifications

Core Performance & Functional Specifications
Blade type 6-12 degrees
Feed rate 0.5-400 mm/s
Section thickness 1.0-50 µm
Specimen size 30 mm
Dimensions, Weight & Mechanical Design
Weight 12 lbs
Width 21 x 17 x 13 inch
Electrical, Power & Drive Specifications
Power consumption 110-240 VAC

Description

Construction and positioning

The LM8A is a benchtop, fully automatic rotary microtome. Its architecture combines a motorized cutting drive with a stepping-motor horizontal feed and a manually adjustable specimen clamp. The specimen clamp position is settable from 0 to 30000 µm, and the cassette clamp can be rotated 360° with an 8° angular adjustment in all directions. The knife carrier angle is adjustable from 6° to 12°. ### Behavior under load and in different modes

The instrument supports two principal operating modes: manual sectioning via the handwheel and motorized sectioning via the cutting drive. In motorized mode, cutting speed is continuously adjustable from 0.5 to 400 mm/s, and the cutting window is defined by setting the lower edge of the specimen approximately 3 mm above the knife edge as the Start point. Multi-stroke mode allows programming 1 to 100 sections with automatic stop. Section thickness is selectable from 1.0 to 50 µm with resolution steps of 0.25 µm (up to 5 µm), 0.5 µm (5–10 µm), and 1.0 µm (10–50 µm). Trimming thickness ranges from 1.0 to 500.0 µm with coarser resolution at higher values. Coarse feed speed is controlled by press duration on the feed keys, with brief presses moving 5 µm. ### Installation, connection, and control

Integration & Operating Notes

Electrical
Use only 2 × T 2A fuses; wrong rating voids protection

The mains fuse drawer accepts two T 2A (time-lag) fuses. Substituting a different rating or a fast-blow fuse may not protect the internal electronics from overcurrent. Replacement requires switching off the unit, setting the on/off switch to ‘0’, and removing the power cord before pulling the fuse drawer.

User Manual, Section 5.1.1 Fuse Replacement, p.29
Mounting
Allow acclimatization and use a stable, level bench

The LM8A must be installed on a stable, level surface and allowed to acclimatize to the operating environment before power-on. Operating ambient temperature must remain between +5 °C and +40 °C. Failure to acclimatize or level the unit can affect sectioning precision and mechanical alignment.

User Manual, Section 3 Startup and Mounting
Control & I/O
Motorized cutting is blocked by brake or emergency stop

The motorized cutting drive will not function if the handwheel brake is applied or the emergency stop is engaged. Release the brake and reset the emergency stop before attempting motorized sectioning. This interlock prevents unexpected handwheel movement during automated cutting.

User Manual, Section 4.8.7 Motorized Cutting Drive, p.26
Process Limits
Set cutting window with specimen ~3 mm above knife edge

When defining the cutting window Start point, position the lower edge of the specimen approximately 3 mm above the knife edge. This clearance prevents the specimen from striking the knife during setup and ensures the motorized drive begins cutting only after the specimen has entered the safe cutting zone.

User Manual, Section 4.8.1 Cutting Window, p.24
Maintenance
Daily cleaning and annual factory service are mandatory

Clean the microtome daily according to the specified cleaning agents and procedures. Check guiding, locking, and clamping levers weekly. Every 12 months, arrange service by Avantik. These intervals are required to maintain sectioning accuracy and mechanical reliability; skipping them may lead to premature wear or alignment drift.

User Manual, Section 6.2 Cleaning and Maintenance, p.30
Control & I/O
Position memory and section counter reset on power cycle

The three position memory values (M1–M3) and the section counter are not retained after a restart. Users relying on stored positions or cumulative counts must re-enter or re-zero them after every power cycle. This behavior should be accounted for in batch workflows.

User Manual, Sections 4.9.1 and 4.9.3, pp.27–28

Selection & Integration Q&A

When should I choose the LM8A instead of a manual or semi-automatic microtome from the same series?

Choose the LM8A when you need motorized cutting with continuously adjustable speed (0.5–400 mm/s) and multi-stroke batch sectioning up to 100 sections.

What are the power supply and fusing requirements for integrating the LM8A into a laboratory bench?

The LM8A accepts 110–240 VAC, 50–60 Hz, and requires two T 2A time-lag fuses in the mains inlet. Ensure the supply is within this voltage and frequency range and that the fuse drawer is fitted with the correct rating. The unit does not include a power cord or external controller in the base delivery.

What environmental and duty-cycle limits apply when installing the LM8A on a lab bench?

Install on a stable, level surface in an ambient temperature of +5 °C to +40 °C. Allow acclimatization before first use. The motorized drive is interlocked against the handwheel brake and emergency stop, so duty cycle is limited by the need to release the brake before each motorized run. Storage temperature range is −20 °C to +50 °C.

What maintenance schedule and accessories are required, and what is not included in the base delivery?

Daily cleaning, weekly checks of guiding/locking/clamping levers, and annual service by Avantik are required. The base delivery does not include a foot switch, power cord, or spare fuses; these must be ordered separately. The foot switch is used for motorized cutting control, and fuses are 2 × T 2A.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Avantik-LM8A-Microtome-Sheet
Datasheet ·1,2 МБ ·2 pp.
View PDF
PDF
LM8A-Automatic-Microtome-User-Manual
Manual ·7,2 МБ ·34 pp.
View PDF