AX-65 Diffusion Pump

Contents

  1. Quality p. 3
    Warranty terms and conditions for the product
  2. Overview p. 5
    Table of contents listing document sections
  3. Overview p. 7
    List of figures referenced in the manual
  4. Overview p. 9
    List of tables referenced in the manual
  5. Overview p. 11
    Introduction and general information about the manual
  6. Procedure pp. 17-24
    Installation instructions and setup procedures
  7. Procedure pp. 25-26
    Operating instructions and procedures
  8. Procedure pp. 27-28
    Troubleshooting guide and fault diagnosis
  9. Procedure pp. 29-32
    Maintenance and service procedures
  10. Ordering pp. 33-37
    Spare parts list and replacement information
  11. p. 42
    Contact information for service and support
  12. p. 28 table, 16 rows
  13. p. 18 table, 28 rows
  14. p. 19 table, 12 rows
  15. p. 36 table, 14 rows
  16. p. 33 table, 18 rows
  17. p. 37 table, 4 rows
  18. p. 12 table, 8 rows
  19. p. 14 table, 10 rows

Table 3-1 Troubleshooting Guide

FaultProbable CauseCorrective Action
Poor system pressureLeaks in system, virtual or realLocate and repair
Poor ultimate pressureContaminated pump fluidExamine and clean pump; replace fluid
Low heat inputCheck voltage. Check for continuity, burned-out element, poor thermal contact
Excessive or too cold coolingCheck temperature and correct.
High forepressureCheck for leak in foreline, poor mechanical pump performance, breakdown of mechanical pump fluid
Low speed (Prolonged cycle after checking gas load conditions)Low heat inputCheck heaters
Low oil levelAdd oil
Improperly located jetsCheck and repair or replace
Inlet pressure surgesIncorrect heater voltageCheck and correct voltage
Fluid outgassingCondition fluid by operating the pump for a 24 hours
Leak in system ahead of pump inletCheck and correct
High chamber contamination of the pump fluidForepressure too highCheck for leak in foreline, poor mechanical pump performance, breakdown of pump fluid, and incorrect valve operation
Prolonged operation in overload range: 10-3 TorrAdhere to operating procedures
Improper system operation and air release proceduresAdhere to operating procedures
Pump will not startSafety circuits openInspect for electrical short or overtemperature condition
Main power offCheck main power source

Table 1-1 Operating Specifications

SpecificationUnitsAX-65 with cold capAX-65 with baffle
Operating RangeTorr3x10-3 to <5x10-88.0x10-3 to <5x10-8
mbar3.9x10-3 to <6.5x10-810.4x10-3 to <6.5x10-3
Pumping Speed, Maximuml/s, Air6530
l/s, Helium9065
Throughput (Air), Max. Optimum RangeTorr-l/s0.190.30
mbar-l/s0.250.40
Overload Range (@ 1x10-2 Torr)Torr-l/s
mbar-l/s
Forepressure, Max. No LoadTorr0.750.60
mbar1.000.78
Full LoadTorr
mbar
Compression Ratio Air4x107
He2x106
Backstreaming Rate at Pump Inletmg/cm2/min<2x10-4<2x10-5
Electrical PowerWatts200 W nominal, 260 W maximum250 W nominal, 320 W maximum
Warmup Timeminutes7
Cooldown Timeminutes10
Fluid Chargecc30
Recommended backing pump capacitycfm (m3/hr)1.5 (2.5)
Cooling Fancfm30
Inlet FlangeN/ANW50 or ISO-63-K
BodyN/AStainless steel
Jet AssemblyN/AThree-stage, self-aligning, aluminum, w/aluminum cold cap
Net Weightlbs (kg)8 (3.63)
EnvironmentalAmbient temperature: 5 °C (41 °F) to 45 °C (113 °F); Maximum relative humidity: 80% for temperatures up to 31 °C (88 °F) decreasing linearly to 50% relative humidity at 40 °C (104 °F)
InstallationIndoor use, Installation Category II, Pollution Degree 2
Altitudem (ft)2000 m (6562’)

Table 1-2 AX-65 Electrical Specifications

ModelVoltage VAC, 50/60 Hz 1 phase, ±10%Part Number Inlet Flange Type KF50Part Number Inlet Flange Type ISO63
AX-65 with standard cold cap115V, 250WL9670301L9707301
AX-65 with internal baffle115V, 250WL9670311L9707311
AX-65 with standard cold cap220V, 250WL9670302L9707302
AX-65 with internal baffle220V, 250WL9670312L9707312
AX-65 with standard cold cap115V, 200WL9670303L9707303
AX-65 with internal baffle115V, 200WL9670313L9707313
AX-65 with standard cold cap220V, 200WL9670304L9707304
AX-65 with internal baffle220V, 200WL9670314L9707314
AX-65 with standard cold cap90V, 250WL9670305L9707305
AX-65 with internal baffle90V, 250WL9670315L9707315
AX-65 with standard cold cap165V, 250WL9670306L9707306
AX-65 with internal baffle165V, 250WL9670316L9707316

Table B-3 Ordering Information

DescriptionContentsAgilent Part No.
NEOVAC SY1 liter/1,000 ccK6948301
NEOVAC SY1 gallon/3.8 litersK6948305
NEOVAC SY5 gallons/18.9 litersK6948315
DC-702 Silicone500 cc695472005
DC-702 Silicone1 gallon/3.8 liters695472008
DC-702 Silicone5 gallons/18.9 liters695472015
DC-704 Silicone500 cc695474005
DC-704 Silicone1 gallon/3.8 liters695474008
DC-704 Silicone5 gallons/18.9 liters695474015
DC-705 Silicone500 cc695475005
DC-705 Silicone1 gallon/3.8 liters695475008
Santovac 540 cc695405001
Santovac 565 cc695405002
Santovac 5500 cc695405005

Table B-1 AX-65 Diffusion Pump Replacement Parts List

ItemPart NumberDescription
1L9674301Jet Assembly
2L9675301Cold Cap
3L9706001Inlet Baffle
4L9995301Jet Alignment Kit
5L9994301Heater Harness, 90V, 250W
L9994303Heater Harness, 115V, 250W
L9994302Heater Harness, 165V, 250W
L9994304Heater Harness, 220V, 250W
L9994307Heater Harness, 115V, 200W
L9994308Heater Harness, 220V, 200W
6Not Available
7Not Available
8661300138Fan, Universal Voltage
9L9676301Foreline Baffle
10L9680301Centering Ring Assembly, NW50, with Viton O-ring
11KC16AVCentering Ring Assembly, NW16, with Viton O-ring
12L9701002Bottom Insulation (2 required with heater harness)
695405001Diffusion Pump Fluid, Santovac 5, 40 cc

RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:

NORTH AMERICA:PACIFIC RIM:EUROPE:
Fax: 1 781 860 9252please visit our website for individual office information http://www.DJLOHQW.comFax: 00 39 011 9979 330
Toll Free: 800 882 7426, Option 3Fax Free: 00 800 345 345 00
vpl-ra@agilent.comToll Free: 00 800 234 234 00
vpt-customercare@agilent.com

Таблица контактов по регионам; в тексте также присутствует заголовок 'Request for Return Health and Safety Certification'

Table 1 General Hazards

HazardSuggested Corrective Action
Loss of utility: water and/or electricityProvide sufficient backup water and power supply as necessary to effect a safe shutdown under worst case conditions
Overpressure in forelineProvide an interlock to ensure that the power supply to the pump heater cannot be activated if the foreline pump is not running and/or the pressure in foreline is above 0.5 Torr (0.67 mbar)
OvertemperatureFit temperature sensors and pump fluid level sensors with a feedback to an interlock on the heater power supply
Loss of electrical ground integrityIncorporate ground fault interrupt circuit into heater power supply
Positive pressure in pumping systemIntegrate pressure relief valve in vacuum system
High voltagePrevent personnel contact with high voltages; design and attach warnings
Toxicity and CorrosivityVent toxic and/or corrosive gases to a safe location; ensure adequate dilution or scrubbing to safe levels; take all actions required to meet air quality standards
ExplosionIntegrate pressure relief valves Do not use hydrocarbon-based pumping fluids

Table 2 Explosive Conditions

Prohibited ActionExplosion-Causing Condition
Do not run pump without cooling waterOvertemperature
Do not run pump with low level of pump fluidOvertemperature
Do not run pump without proper backing or holding pumpOverpressure
Do not run pump when not evacuated below 0.5 Torr (0.66 mbar)Overpressure
Do not admit air to, or rough through, a pump with hot boilerOverpressure plus strong oxidizer
Do not open drain or fill plug while pump is under vacuum, especially when it is hotOverpressure plus strong oxidizer
Do not contaminate pump with explosive vaporsLower explosive threshold of gas mixtures
Do not remove, defeat, or override safety counter-measures such as pressure and thermal switches and valve sequencer interlocksOvertemperature, overpressure, more combustible mixtures
Do not machine or weld any part of the pump without removing all fluid or solvent residue in pumpSource of ignition
Do not use unsuitable pumping fluidLower explosive threshold of gas mixture

Engineering Facts

14 facts · every claim linked to a page
Selection

Choose polyphenyl ether fluid (e.g., Santovac 5) for systems with internal electronics and pair it with a 250 W heater; use lower-boiling fluids like DC-704 only with 200 W heaters.

Fluid choice is tied to heater wattage and system electronics; wrong pairing risks overheating or poor performance.

Source
“When a diffusion pump is integrated into a system that has internal electronics, use a polyphenyl ether diffusion pump fluid such as Santovac 5.”
Installation

Mount the pump body vertical and plumb with the inlet mating flange horizontal within ±1° so the boiler plate stays level and avoids overheating.

Uneven fluid level from tilt can overheat the boiler plate and damage the pump.

Source
“Install the diffusion pump must be installed with the body vertical and plumb. Ensure that the pump inlet mating flange on the system is horizontal within ± 1°.”
Installation

Provide at least 1 inch clearance between the cooling fan and any closet surface, and mount the fan 90° from the foreline.

Insufficient airflow causes overheating; fan orientation affects cooling of the pump body.

Source
“Achieve adequate air flow by ensuring there is a minimum clearance of 1" between the fan and the closet surface.”
Installation

Install supplementary fusing no larger than 240 V, 15 A fast-acting for overcurrent protection when the pump is integrated into a system.

Correct fusing protects the heater circuit and prevents fire or equipment damage.

Source
“For overcurrent protection, the pump must have supplementary fusing when installed in a system. The fuse cannot be larger than 240 V, 15 A, fast acting.”
Operation

Never turn on the heater without fluid in the pump, and never run the heater unless the cooling fan is running.

Dry running or loss of cooling destroys heaters and overheats the pump fluid.

Source
“Do not turn on the heater without fluid in the pump. This may ruin the heaters and damage the pump.”
Operation

Keep forepressure below approximately 0.5 Torr during operation; higher forepressure causes poor pumping and excessive backstreaming.

Exceeding tolerable forepressure prevents proper operation and contaminates the chamber.

Source
“Do not operate the pump with forepressure (outlet pressure) above approximately 0.5 Torr”
Operation

After shutdown, let the fan run about 10 minutes (or 30 minutes if fan and heater are wired together) and confirm the body is at most 130 °F before venting.

Venting a hot boiler exposes hot fluid to oxidizer and can cause explosion or fluid breakdown.

Source
“The temperature of the pump body just above the boilerplate should be a maximum of 130 °F before the pump is vented.”
Maintenance

While running, keep fluid level within the FULL HOT zone; if low, cool down and add fluid to the FULL COLD zone (30 cm³ charge for an empty pump).

Low fluid level causes overheating and loss of pumping performance.

Source
“When the pump is running, the fluid level should remain within the zone indicated as FULL HOT.”
Maintenance

Do not use solvents on O-rings; wipe them with a clean lint-free cloth or a small amount of diffusion pump oil instead.

Solvent attack on O-rings causes leaks and loss of vacuum integrity.

Source
“Solvents degrade O-ring materials reducing their ability to hold vacuum. Do not use solvents on O-rings.”
Safety

Provide an interlock so heater power cannot be activated unless the foreline pump is running and foreline pressure is below 0.5 Torr.

Prevents overpressure operation that can cause explosion and equipment damage.

Source
“Provide an interlock to ensure that the power supply to the pump heater cannot be activated if the foreline pump is not running and/or the pressure in foreline is above 0.5 Torr (0.67 mbar)”
Safety

Never pressurize any part of the vacuum system for testing or any other purpose; vacuum components are not designed for positive pressure.

Pressurization can burst the pump body and expel shrapnel at lethal velocity.

Source
“Never pressurize any part of a vacuum system for test or any other purpose.”
Safety

Boiler surfaces reach 530 °F (275 °C); ensure surfaces have cooled to near room temperature before touching.

Prevents serious burns during maintenance and operation.

Source
“Boiler temperatures reach 530 °F (275 °C) which can cause serious burns.”
Performance

The gas load at the inlet must not exceed the maximum throughput of 0.30 Torr-liters/sec, or the diffusion pump will not function.

Exceeding throughput causes loss of pumping action and relies only on the backing pump.

Source
“The gas load must be reduced to the maximum throughput specification (0.30 Torr-liters/sec) or less.”
Selection

For leak detector applications, Agilent recommends using Santovac 5 fluid only.

Ensures compatibility and performance in sensitive leak detection systems.

Source
“For leak detectors, Agilent recommends the use of Santovac 5 only.”