BF (E1) 01 2025
Technical Reference Summary
Covered Models & Document Scope
Covers BINDER BF / BF-UL incubators with forced convection and microprocessor temperature controller: BF 400 (BF400-230V, Art. No. 9010-0241, 9110-0241) and BF 400-UL (BF400UL-120V, Art. No. 9010-0242, 9110-0242). Options include disconnectable audible over-temperature alarm, APT-COM 4 software, analog temperature output, additional Pt100 sensor, and internal socket.
Section-by-Section Technical Breakdown
Describes the BF/BF-UL incubator construction: electronic PID controller with digital display (0.1 °C resolution), electrical heating with fan-assisted forced-air circulation, APT.line preheating chamber system, and stainless steel V2A (1.4301/AISI 304) interior. The chamber operates from 5 °C above ambient up to +100 °C and includes a DIN 12880:2007 temperature safety device. A serial RS 422 interface is provided for computer communication.
- Temperature range: 5 °C / 9 °F above ambient up to +100 °C / 212 °F
- Electronic PID controller with digital display, accuracy of a tenth of a degree
- APT.line preheating chamber system with fan-assisted forced-air circulation
- Inner chamber, preheating chamber and door interiors made of stainless steel V2A (1.4301, AISI 304)
- Housing powder-coated RAL 7035 or RAL 9003
- Serial interface RS 422 for computer communication (e.g. APT-COM 4)
- Chamber overview items: display, set-point key, selector keys, time management key, ON/OFF switch, ventilation slide lever, safety device, door handle, internal socket switch (option), buzzer switch (option)
Specifies electrical connection requirements for BF 400 (grounded plug, 1N~) and BF-UL 400 (NEMA 5-15P, 115 V at 60 Hz, 1N~). Installation must avoid unventilated recesses and observe minimum distances; an optional suction plant connection is described. The chamber must not be installed in damp areas or puddles and must be set up splash-proof.
- BF 400: grounded plug, current type 1N~
- BF-UL 400: NEMA 5-15P, 115 V at 60 Hz, 1N~
- Do NOT install in unventilated recesses; ensure sufficient ventilation for heat dispersal
- Observe prescribed minimum distances when installing (chap. 3.4)
- Optional connection to a suction plant
- Do NOT install in damp areas or puddles; set up splash-proof
Covers turning on the chamber by pressing the switch until the display lights up, the heating operation display, and air change via the ventilation slide. If the temperature safety device has tripped (chap. 6.3), the chamber is inactive (no heating) and the display shows an alarm value.
- Press switch until display lights up
- Heating operation display indicates active heating
- Air change controlled via ventilation slide lever
- If safety device tripped (chap. 6.3), chamber is inactive (no heating) and display shows alarm value
Describes display and entry of temperature and ventilation set-points with and without ramp function, continuous vs. timer operation, timer value setting, and user-level settings including °C/°F unit change, temperature ramp entry, chamber addressing, timer function selection, and interface mode/printer interval. Includes a temperature programming example.
- Display/entry of temperature and ventilation set-points without ramp function (chap. 6.1)
- Display/entry with selected temperature ramp (chap. 6.2)
- Time functions: continuous operation and timer operation (chap. 6.3)
- Switching between continuous and timer operation (chap. 6.3.1)
- Setting timer values (chap. 6.3.3)
- User level settings: °C/°F change, temperature ramp, chamber addressing, timer function, interface mode and printer interval (chap. 6.4)
- Temperature programming example (chap. 6.5)
The chamber is equipped with a temperature safety device class 3.1 according to DIN 12880:2007 as an additional safety controller. Visual and audible (buzzer) signals indicate temperature exceeding. The safety device is accessible on the chamber overview (item 7).
- Temperature safety device class 3.1 acc. to DIN 12880:2007
- Visual and audible (buzzer) signals indicate temperature exceeding
- Safety device is easily accessible (chamber overview item 7)
- If tripped, chamber is inactive (no heating) and display shows alarm value
Lists available options: disconnectable audible over-temperature alarm, APT-COM 4 Multi Management Software, analog output for temperature, additional Pt100 temperature sensor, and internal socket. The audible alarm option adds a buzzer switch (chamber overview item 10).
- Disconnectable audible over-temperature alarm (option) with buzzer switch
- APT-COM 4 Multi Management Software (option) for RS 422 communication
- Analog output for temperature (option)
- Additional Pt100 temperature sensor (option)
- Internal socket (option) with ON/OFF switch (chamber overview item 9)
Provides procedures for cleaning and chemical disinfection/decontamination of the chamber. After using disinfectant spray, allow the chamber to dry thoroughly and aerate sufficiently. Do not empty disinfectant into drains; wear protective goggles.
- Cleaning procedures (chap. 9.1)
- Decontamination / chemical disinfection (chap. 9.2)
- After disinfectant spray, allow chamber to dry thoroughly and aerate sufficiently
- Do NOT empty disinfectant into drains
- Wear protective goggles when handling disinfectant spray
Covers maintenance intervals (at least once a year), personnel qualification, simple troubleshooting by operating personnel, and requirements for repair and electrical testing. Annual repeat inspection and testing before initial startup and after maintenance/repair are required per EN 50678/VDE 0701 and EN 50699/VDE 0702. The door gasket must be replaced only when cold.
- Maintenance work at least once a year by licensed electricians or experts authorized by BINDER
- Warranty void if maintenance by non-authorized personnel
- Replace door gasket only when cold
- Annual repeat electrical inspection and test before initial startup and after maintenance/repair per EN 50678/VDE 0701 and EN 50699/VDE 0702
- Disconnect power plug before maintenance work; do NOT remove rear panel
- Simple troubleshooting by operating personnel does not require technical intervention or disassembly
- Detailed troubleshooting and repair by BINDER Service or qualified partners per Service Manual
- After maintenance, chamber must be tested prior to resuming operation
Provides factory calibration and adjustment information, definition of usable volume, overcurrent protection, technical data for BF/BF-UL 400, and extracts of equipment/options and accessories/spare parts. Nominal temperature is 100 °C (212 °F); nominal power 0.85 kW; nominal current 3.7 A; nominal voltage 230 V / 50 Hz or 230 V / 60 Hz, 1N~; IP protection 20.
- Factory calibration and adjustment (chap. 12.1)
- Definition of usable volume (chap. 12.2)
- Overcurrent protection (chap. 12.3)
- Technical data BF/BF-UL 400 (chap. 12.4)
- Equipment and options extract (chap. 12.5)
- Accessories and spare parts extract (chap. 12.6)
- Nominal temperature 100 °C / 212 °F; nominal power 0.85 kW; nominal current 3.7 A; nominal voltage 230 V / 50 Hz or 230 V / 60 Hz, 1N~; IP protection 20
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Temperature range: 5 °C / 9 °F above ambient temperature up to +100 °C / 212 °F. p.15 ↗ | Selection must account for ambient temperature; minimum achievable temperature is ambient +5 °C. | BF / BF-UL series |
| Nominal temperature 100 °C / 212 °F; nominal power 0.85 kW; nominal current 3.7 A; nominal voltage 230 V / 50 Hz or 230 V / 60 Hz, 1N~; IP protection 20. p.8 ↗ | Electrical supply must match 230 V ±10% at indicated frequency; overcurrent protection and wiring sized for 3.7 A. | BF 400 (BF400-230V) |
| BF-UL 400 electrical connection: NEMA 5-15P, 115 V at 60 Hz, 1N~. p.18 ↗ | UL variant requires 115 V / 60 Hz supply with NEMA 5-15 receptacle. | BF 400-UL (BF400UL-120V) |
| Permissible ambient temperature range during transport and storage: -10 °C up to +60 °C / 14 °F up to 140 °F. p.17 ↗ | Transport and storage conditions must remain within this range; after cold storage, wait at least one hour before start-up. | BF / BF-UL series |
| Permissible ambient humidity during storage: max. 70 % r.H., non-condensing. p.17 ↗ | Storage location must be dry and non-condensing to avoid moisture damage. | BF / BF-UL series |
| After storage in a cold location, wait at least one hour before start-up until chamber attains ambient temperature and is completely dry. p.17 ↗ | Prevents condensation-related electrical hazards and damage on commissioning. | BF / BF-UL series |
| Do NOT lift chambers size 400 by hand; use fork lifter from the rear in the middle of the chamber, with all lateral supports on the forks. p.16 ↗ | Mechanical lifting required for BF 400; improper lifting risks injury and chamber damage. | BF 400 / BF 400-UL |
| Transport chambers size 400 ONLY with the original transport pallet; set fork lifter only to the pallet. p.17 ↗ | Without pallet the chamber is in imminent danger of overturning. | BF 400 / BF 400-UL |
| Maintenance work at least once a year by licensed electricians or experts authorized by BINDER. p.38 ↗ | Annual service required; warranty void if maintenance by non-authorized personnel. | BF / BF-UL series |
| Annual repeat electrical inspection and test before initial startup and after maintenance/repair per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.37 ↗ | Electrical safety testing is mandatory to prevent shock risk; documentation required. | BF / BF-UL series |
| Replace the door gasket only when cold. p.38 ↗ | Replacing a hot gasket may damage it. | BF / BF-UL series |
| Temperature safety device class 3.1 acc. to DIN 12880:2007 with visual and audible (buzzer) signals. p.14 ↗ | Provides over-temperature protection; if tripped, chamber is inactive (no heating) and display shows alarm value. | BF / BF-UL series |
| Serial interface RS 422 for computer communication, e.g. via APT-COM 4 Multi Management Software. p.15 ↗ | Enables external monitoring and data logging when APT-COM 4 option is used. | BF / BF-UL series |
| Inner chamber, preheating chamber and inside of doors made of stainless steel V2A (German material no. 1.4301, US equivalent AISI 304). p.15 ↗ | Material compatibility: avoid corrosive ingredients such as acids and halides. | BF / BF-UL series |
| Do NOT install the chamber in unventilated recesses; ensure sufficient ventilation for heat dispersal and observe prescribed minimum distances (chap. 3.4). p.9 ↗ | Prevents overheating and damage to the chamber. | BF / BF-UL series |
| Chamber must NOT become wet during operation, cleaning, or maintenance; do NOT install in damp areas or puddles; set up splash-proof. p.10 ↗ | Prevents deadly electric shock from water ingress. | BF / BF-UL series |
| If persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance of 30 cm from field sources, influence can be excluded with high probability. p.14 ↗ | Maintain 30 cm distance from electrical equipment for implant wearers. | BF / BF-UL series |
| Loading material must not contain corrosive ingredients, in particular acids and halides. p.11 ↗ | Corrosive damage caused by such ingredients is excluded from liability by BINDER GmbH. | BF / BF-UL series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber is inactive (no heating); display shows alarm value p.20 ↗ | Temperature safety device (chap. 6.3) has tripped | Check and reset temperature safety device per chap. 6.3; if fault persists, contact BINDER Service | Chap. 5.1 |