BINDER BF 400
BINDER ·Made in Germany · In Active Production

BINDER BF 400

BINDER BF 400 — Incubators with Temperature 5…100 °C.

The BINDER BF 400 is a 400 L forced-convection incubator with a stainless steel V2A (1.4301/AISI 304) interior and PID microprocessor control, operating from 5 °C above ambient to +100 °C. Within the BF series it is the mid-large 400 L modification, distinguished by a 0.85 kW single-phase 230 V supply and a 90 kg chamber load limit. It is intended for reproducible incubation and thermal conditioning of bulk laboratory samples in a temperature-controlled, fan-assisted air stream.

Temperature
5…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BF Series — Model Comparison & Selection

Model Temperature, °CHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kgPermitted load, kg
BF 400 Current 5–10012102.54003590
BF 115 8–1006572.511430150
BF 260 7–1003084.525740270
BF 56 7–1003562.5593070
BF 720 10–1004515973445315

Full specifications

Core Performance & Functional Specifications
Temperature 5...100 °C
Heating-up time to 100 °C 12 min
Interior volume 400 L
Load per rack 35 kg
Nominal power 0.85 kW
Permitted load 90 kg (BF400-230V / BF400UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.4 K
Back wall clearance 160 mm (BF400-230V / BF400UL-120V)
Side wall clearance 100 mm (BF400-230V / BF400UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 5 min
Dimensions, Weight & Mechanical Design
Depth 765 mm (BF400-230V / BF400UL-120V)
Height 1025 mm (BF400-230V / BF400UL-120V)
Interior width 1000 mm (BF400-230V / BF400UL-120V)
Width 1235 mm (BF400-230V / BF400UL-120V)
Net weight of the unit (empty) mfr 145 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 12.5 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior

Installation, control and operation

Integration & Operating Notes

Mounting
Respect 160 mm rear and 100 mm side clearance for heat dispersal

The chamber must not be installed in unventilated recesses. Prescribed minimum distances to walls and adjacent equipment are required so that heat from the 0.85 kW heating system can dissipate; otherwise the unit may overheat and the ambient-referenced lower temperature limit shifts.

Instruction Manual BF (E1) 01-2025, General safety instructions p.9 and chap. 3.4
Process Limits
Keep total chamber load below 90 kg and 35 kg per rack

Exceeding the 90 kg permitted load or the 35 kg per-rack limit can deform the racks and loading surfaces and disturb the forced-air circulation, invalidating the 0.4 K uniformity figure. Distribute samples evenly and account for the usable volume rather than the full 400 L interior.

Factory Specification Sheet BF 400 and Instruction Manual BF (E1) 01-2025, Technical Description
Process Limits
Do not load acids, halides or other corrosive substances

The V2A (1.4301/AISI 304) interior is not resistant to corrosive ingredients such as acids and halides. Damage caused by such loading material is excluded from BINDER liability, so process media and spillage must be checked for compatibility before use.

Instruction Manual BF (E1) 01-2025, Intended use p.11
Maintenance
Schedule annual maintenance and repeat electrical safety testing

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER, and annual repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is mandatory before initial startup and after any repair. Non-authorized servicing voids the warranty.

Instruction Manual BF (E1) 01-2025, Maintenance intervals p.38 and Maintenance p.37
Maintenance
Replace the door gasket only when the chamber is cold

The door gasket must be replaced only in the cold state; fitting a gasket on a hot chamber can damage the seal and compromise temperature uniformity. After cleaning or disinfection, allow the chamber to dry thoroughly and aerate before restarting.

Instruction Manual BF (E1) 01-2025, Maintenance intervals p.38 and Cleaning
Environment
After cold storage, wait one hour before start-up

Transport and storage are permitted from -10 °C to +60 °C at max. 70 % r.H. non-condensing. After storage in a cold location, the chamber must stand at ambient for at least one hour until it is completely dry, preventing condensation-related electrical hazards on commissioning.

Instruction Manual BF (E1) 01-2025, Transportation/storage p.17

Selection & Integration Q&A

When should I choose the BF 400 instead of the BF 260 or BF 720?

Pick the BF 400 for 400 L of usable volume with a +5 °C-above-ambient lower limit, the widest near-ambient range in the series. Choose the BF 260 for smaller loads or the BF 720 for larger batches, noting the BF 720's higher +10 °C lower limit.

What supply and protection does the BF400-230V need, and what changes for the UL variant?

The 230 V variant needs a grounded 230 V 1N~ 50/60 Hz supply with CEE 7/7 plug, sized for 3.7 A and 0.85 kW with 12.5 A unit fuse. The UL variant instead requires 115 V 1N~ 60 Hz via NEMA 5-15 receptacle.

What installation clearances and ambient conditions must I provide for the BF 400?

Allow 160 mm at the rear and 100 mm at the sides, avoid unventilated recesses, and set up splash-proof away from damp areas or puddles. The lower temperature limit is ambient +5 °C, so room temperature directly sets the achievable minimum.

What maintenance, accessories and consumables are not included with the base unit?

Annual maintenance by BINDER-authorized personnel and repeat electrical testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required but not included. APT-COM 4 software, analog temperature output, additional Pt100 sensor, internal socket and audible alarm are options.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model BF 400 en
Datasheet ·1,1 МБ ·7 pp.
View PDF
PDF
BF (E1) 01 2025 en
Manual ·5,7 МБ ·59 pp.
View PDF