Induztr
Engineering Digest Document Viewer Related Models

CB+CBF (E7.1) 01 2026

Technical Reference Summary

Covered Models & Document Scope

This operating manual (E7.1, 01/2026) covers BINDER CO2 incubators of the CB / CB-UL series (sizes 56, 170, 260) and CBF / CBF-UL series (sizes 170, 260) with sterilizable NDIR CO2 sensor and MB2 touchscreen program controller. Variants include 200-230 V and 100-120 V (UL) models, left-hinged doors (-L), O2 control 0-20 vol.-% (-O), and CBF humidity control. Article numbers range from 9040/9140-0197 to 9040/9140-0218.

Section-by-Section Technical Breakdown

pp. 25-30
Chamber Description

Describes the chamber overview, inner chamber layout for CB/CB-UL and CBF/CBF-UL, the control panel on the rear, and the instrument panel. It provides the physical orientation of components needed before installation.

  • CB/CB-UL inner chamber differs from CBF/CBF-UL inner chamber
  • Control panel is located on the rear of the chamber
  • Instrument panel with MB2 touchscreen controller
pp. 36-58
Installation and Connections

Covers shelves, Permadry water pan for CB/CB-UL, O2 sensor connection, gas connections for CO2/O2/N2, water supply and dehumidification for CBF/CBF-UL, electrical connection, and optional divided inner door. Includes gas cylinder connection kits and water supply accessories.

  • Permadry water pan used on CB/CB-UL for humidity
  • O2 sensor connection required for chambers with O2 control
  • Gas connections for CO2, O2, and N2 with specific cylinder kits
  • CBF/CBF-UL water supply via prefilled bottle, freshwater bag, or freshwater can
  • Electrical connection details and divided inner door option
pp. 59-68
Functional Overview of the MB2 Chamber Controller

Explains operating functions in normal display, display views, controller icons, operating modes, menu structure, entry principles, and behavior during power failures and door opening. This section is the basis for configuring and operating the controller.

  • Normal display, program display, and chart-recorder display views
  • Operating modes include fixed value and program modes
  • Main menu, Settings submenu, and Service submenu structure
  • Controller behavior during power failures and door opening
pp. 68-72
Start Up

Covers turning on the chamber, controller settings at start up, factory settings, equilibration time, DuoDoor door lock operation, required gas supply for O2 control, activating/deactivating CO2/O2/humidity control, and CB/CB-UL humidity control of the Permadry system.

  • Factory settings and equilibration time after start up
  • DuoDoor door lock operating instructions
  • O2 control ranges: hypoxic 0.2-20 vol.-% (regular) or 10-90 vol.-% (option)
  • Activation/deactivation of CO2, O2, and humidity control
  • Permadry humidity control for CB/CB-UL
pp. 73-76
Set-Point Entry in Fixed Value Operating Mode

Describes entering setpoints through the Setpoints menu or directly via the normal display, special controller functions, and safety instructions for setting high gas concentrations.

  • Setpoint entry via Setpoints menu or direct normal display
  • Special controller functions available in fixed value mode
  • Safety instructions for high gas concentrations
pp. 77-92
Timer Program and Time Programs

Covers the stopwatch timer function and time programs, including starting, stopping, pausing, cancelling, creating programs, program and section editors, value entry for sections, setpoint ramps and steps, tolerance ranges, and repeating sections.

  • Stopwatch timer program operation
  • Time program creation with program and section editors
  • Setpoint ramp and setpoint step modes
  • Tolerance range and repeat functions within time programs
pp. 92-100
Week Programs

Describes starting, cancelling, and creating week programs, program and section management, and value entry including setpoint ramp/step modes, weekday, start time, setpoint, and special controller functions.

  • Week program creation and management
  • Weekday and start time settings
  • Setpoint ramp and step modes for week programs
pp. 101-107
Notification and Alarm Functions

Provides an overview of notifications and alarm messages, state of alarm, resetting alarms and active alarm list, audible alarm settings, tolerance range and alarm delay settings, and zero-voltage relay alarm output.

  • Alarm messages and notifications overview
  • Resetting alarms and list of active alarms
  • Tolerance range settings and alarm delay times
  • Zero-voltage relay alarm output
pp. 108-110
Temperature Safety Devices

Covers the over temperature protective device (class 1) and overtemperature safety controller class 3.1, including safety controller modes, setting, alarm state messages and measures, and function check.

  • Over temperature protective device class 1
  • Overtemperature safety controller class 3.1
  • Safety controller modes and setting
  • Function check of safety devices
pp. 111-120
User Management

Describes authorization levels and password protection, login/logout, user change, password assignment and change, deleting passwords, and activation codes.

  • Authorization levels and password protection
  • Login, logout, and user change procedures
  • Password assignment, change, and deletion
  • Activation code entry
pp. 121-129
General Controller Settings

Covers menu language selection, date and time setting, temperature unit, display configuration and touchscreen calibration, network and communication settings (RS485, Ethernet, e-mail), and USB data transfer and firmware update.

  • Menu language, date/time, and temperature unit settings
  • Touchscreen calibration
  • Ethernet configuration and MAC address display
  • USB data transfer and controller firmware update
pp. 132-135
Chart Recorder Display

Describes chart recorder views, showing/hiding legend, switching legend pages, showing/hiding specific indications, history display, and setting parameters.

  • Chart recorder display views and legend control
  • History display and parameter settings
pp. 136-137
CBF / CBF-UL Humidification System

Explains the function of the humidifying system and the dehumidifying system for CBF/CBF-UL chambers.

  • Humidifying system function
  • Dehumidifying system function
pp. 137-143
Options and Accessories

Lists options and accessories including APT-COM 4 software, silicone access ports, quick sample access, interior socket, analog outputs, access port for extra-low voltage, gas supply service, stands, stacking adapters, and water supply sets.

  • APT-COM 4 Multi Management Software
  • Silicone access ports and quick sample access
  • Analog outputs for temperature and CO2
  • Stands, stacking adapters, and water supply sets
pp. 144-145
Reference Measurements

Covers CO2 reference measurement methods (indirect via pH, direct via indicator tubes or electronic infrared device) and temperature reference measurement.

  • CO2 measurement via pH of cell medium
  • CO2 measurement via chemical indicator tubes
  • CO2 measurement with electronic infrared device
  • Temperature reference measurement
pp. 148-158
Cleaning, Decontamination / Disinfection, and Sterilization

Covers cleaning, chemical disinfection of the chamber and CO2 sensor, disinfection of CBF fresh water supply components, and hot-air sterilization including preparation, running the cycle, premature termination, and completion.

  • Cleaning and chemical disinfection procedures
  • Disinfection of CO2 sensor and CBF fresh water supply components
  • Hot-air sterilization cycle: 4-hour effective phase plus 6-hour cooling to 37 °C
  • Premature termination effects before and after 4 hours
pp. 158-169
Maintenance and Service, Troubleshooting, Repair, Testing

Covers general maintenance information and personnel qualification, replacement of CBF humidifying components, gas inlet fine filters, CO2 sensor, maintenance intervals, simple troubleshooting by subsystem, and returning the chamber to BINDER.

  • Annual maintenance required; CBF steam humidifier maintained at least once a year
  • Replacement procedures for silicone tube, pump head, gas fine filters, and CO2 sensor
  • Simple troubleshooting for heating, gas cylinder pressure, gas control, sterilization, humidity, controller, and open door
  • Electrical testing per EN 50678/VDE 0701 and EN 50699/VDE 0702
pp. 174-185
Technical Description

Provides factory calibration and adjustment, over current protection, definition of usable volume, technical data, equipment/options/accessories extracts, conversion data for non-SI units, gas inlet pressure conversion table, and dimensions for chamber sizes 56, 170, and 260.

  • Technical data for all models
  • Over current protection details
  • Conversion tables for non-SI units and gas inlet pressures (bar to psi)
  • Dimensional drawings for sizes 56, 170, and 260

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
CB 56 and CB 56-UL are available in 200-230 V and 100-120 V versions respectively; CB 170/260 and CBF 170/260 also have 200-230 V and 100-120 V (UL) versions. p.1 ↗ Select model voltage according to local supply; UL models for 100-120 V markets. CB / CB-UL, CBF / CBF-UL series
O2 control option provides a control range of 0-20 vol.-% O2; an alternative control range of 10-90 vol.-% O2 is available as an option. p.1 ↗ Choose O2 control variant based on required oxygen range; standard hypoxic range is 0.2-20 vol.-%. Chambers with O2 control (-O variants)
Left-hinged door option (-L) is available for CB 170 and CB 170-UL models. p.1 ↗ Order -L variant if door opening direction must be left-hinged. CB 170, CB 170-UL
Hot-air sterilization cycle: effective sterilization phase (heating up and holding) finishes automatically after 4 hours, followed by a defined 6-hour cooling phase to 37 °C / 98.6 °F; total cycle 10 hours. p.158 ↗ Plan sterilization cycles with 10-hour duration; CO2/O2 control is deactivated until temperature falls below 80 °C and message is confirmed. All models with hot-air sterilization
CO2/O2 control is deactivated during sterilization and only reactivated after confirming 'Successful sterilization' and when temperature has cooled below 80 °C. p.158 ↗ Do not expect gas control during sterilization or until cooling below 80 °C. All models
Tolerance range alarm example: temperature setpoint 37 °C with tolerance range +/- 2 °C defines a range from 35 °C to 39 °C; function activates only after setpoint has been reached once. p.105 ↗ Configure tolerance range and alarm delay to suit process; alarm will not trigger until setpoint is first reached. All models
Gas sensors are individually adjusted; when exchanging a sensor, control adjustment for CO2 and O2 (chamber with O2 control) must be repeated. p.162 ↗ After sensor replacement, perform CO2 and O2 control adjustment to maintain accuracy. All models
After 8760 operating hours or two years, a maintenance message appears; after confirmation it reappears every two weeks until reset by BINDER Service. p.162 ↗ Schedule sensor/maintenance service at 8760 hours or 2 years; message cannot be permanently cleared by user. All models
Annual repeat electrical inspection and test before initial startup and after maintenance/repair are required; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.158 ↗ Include electrical safety testing in commissioning and annual maintenance schedule. All models
Regular maintenance work must be performed at least once a year; CBF/CBF-UL steam humidifier must be maintained at least once a year. p.162 ↗ Plan annual maintenance; humidifier interval depends on water quality and steam production. All models; CBF/CBF-UL humidifier
Door gasket must be replaced only when cold; otherwise it may become damaged. p.162 ↗ Allow chamber to cool before door gasket replacement. All models
Premature termination of sterilization after less than 4 hours is ineffective; termination after more than 4 hours (during cooling-down phase) is successful sterilization. p.156 ↗ Do not abort sterilization before 4 hours if sterilization is required; after 4 hours cycle can be stopped with successful result. All models with hot-air sterilization
O2 control requires gas supply; regular hypoxic range is 0.2 vol.-% up to 20 vol.-% O2; alternative range 10 vol.-% up to 90 vol.-% is an option. p.71 ↗ Ensure appropriate O2 and N2 gas supplies for the selected control range. Chambers with O2 control
CBF/CBF-UL water supply can be via prefilled water bottle (BINDER-Water-Set accessory), freshwater bag (tubular bag set accessory), or freshwater can (water container set accessory). p.2 ↗ Select water supply accessory based on lab setup and water quality requirements. CBF / CBF-UL
Gas cylinder connection kits are available as accessory for CO2, O2, and N2 connections. p.2 ↗ Order appropriate gas cylinder connection kit for safe gas supply installation. All models requiring gas supply
Zero-voltage relay alarm output resets together with the alarm. p.105 ↗ External alarm systems connected to this relay will see reset when alarm is acknowledged or condition ends. All models
Audible alarm (buzzer) can be set to 'off' or 'on' via Main menu > Settings > Chamber. p.105 ↗ Configure audible alarm according to local noise requirements. All models
Alarm delay times can be set for temperature, CO2, O2 (chamber with O2 control), and humidity (CBF/CBF-UL) deviations, and for door opening. p.105 ↗ Adjust delay times to avoid nuisance alarms during door openings or transient deviations. All models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Gas cylinder pressure too low p.164 ↗ Gas cylinder pressure too low See chapter 22.6.3 for troubleshooting; replace gas cylinder. Chapter 22.6.3
Heating malfunction p.163 ↗ Heating system fault See chapter 22.6.2 for troubleshooting. Chapter 22.6.2
Gas control malfunction p.165 ↗ Gas control system fault See chapter 22.6.4 for troubleshooting. Chapter 22.6.4
Sterilization malfunction p.167 ↗ Sterilization system fault See chapter 22.6.5 for troubleshooting. Chapter 22.6.5
Humidity malfunction (CB / CB-UL) p.167 ↗ Humidity system fault See chapter 22.6.6 for troubleshooting. Chapter 22.6.6
Humidity malfunction (CBF / CBF-UL) p.168 ↗ Humidity system fault See chapter 22.6.7 for troubleshooting. Chapter 22.6.7
Controller malfunction p.168 ↗ Controller fault See chapter 22.6.8 for troubleshooting. Chapter 22.6.8
Open door p.169 ↗ Door open condition See chapter 22.6.9 for troubleshooting. Chapter 22.6.9