BINDER CBF 170
BINDER ·Made in Germany · In Active Production

BINDER CBF 170

BINDER CBF 170 — Incubators with Temperature 6…50 °C.

The CBF 170 is a 170 L CO2 incubator with hot-air sterilization and active humidity regulation (50–95 % RH), operating from +6 to 50 °C with 0.1 K fluctuation and 0.4 K uniformity. It is the mid-size CBF variant (sibling CBF 260 offers larger volume) and is distinguished by integrated steam humidification and a 6-minute recovery time after a 30 s door opening at 37 °C. It serves cell culture processes requiring controlled CO2, humidity, and periodic 10-hour sterilization cycles.

Temperature
6…50 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER CBF Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LPermitted load, kgWidth, cm
CBF 170 Current 6–5071.5871703068
CBF 260 6–5078.51022674074

Full specifications

Core Performance & Functional Specifications
Temperature 6...50 °C
CO2 recovery time 6 min (all variants)
Interior volume 170 L (CBF170, 50/60 Hz)
Load per rack 10 kg (all variants)
Nominal power 1.4 kW (all variants)
Permitted load 30 kg (all variants)
Temperature fluctuation 0.1 K
Temperature uniformity 0.4 K
Back wall clearance 100 mm (all variants)
Side wall clearance 50 mm (all variants)
Article Number mfr 9040
Recovery time after door was opened for 30 s at 37°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 715 mm (all variants)
Height 870 mm (all variants)
Width 680 mm (all variants)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (all variants)
Rated voltage 200...230 V (CBF170-230V)
Unit fuse 10.0 A (CBF170-230V)
Operating Conditions, Environment & Protection
Humidity range 50...95 % RH

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection, and control specifics

Integration & Operating Notes

Electrical
Verify supply voltage and fuse rating before connection

The standard CBF170-230V requires 200–230 V, 50/60 Hz, single-phase supply with a 10.0 A fuse. Connecting to a different voltage (e.g., 100–120 V) will damage the unit. UL variants are available for 100–120 V markets. Ensure the outlet is properly grounded and protected.

Data Sheet Model CBF 170 en.pdf; Operating Manual, Technical Description
Process Limits
Do not exceed 30 kg total load or 10 kg per rack

The permitted load is 30 kg for the entire chamber and 10 kg per rack. Overloading can compromise temperature uniformity, damage shelves, and affect door sealing. Distribute weight evenly and use appropriate racks for heavy items.

Technical Data, load specifications
Control & I/O
CO2 and humidity control are disabled during sterilization

During the hot-air sterilization cycle, CO2 and humidity control are automatically deactivated. They remain off until the chamber temperature falls below 80 °C and the user confirms the 'Successful sterilization' message. Do not expect gas or humidity control during this period; plan cultures accordingly.

Operating Manual, Section 21.3.5 Hot-air sterilization
Maintenance
Annual maintenance and electrical safety test are mandatory

Regular maintenance must be performed at least once a year, including the steam humidifier. An electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after any maintenance or repair. A maintenance message appears after 8760 operating hours or two years and cannot be permanently cleared by the user; BINDER Service must reset it.

Operating Manual, Sections 22.1 and 22.5
Cooling & Media
Water supply accessories are required for humidification

The CBF 170 does not include a water supply in base delivery. Humidification requires an external water source such as a BINDER-Water-Set, freshwater bag set, or water container set. Water quality affects humidifier maintenance intervals; use appropriate water to minimize scale.

Operating Manual, Sections 4.5.2–4.5.4

Selection & Integration Q&A

When should I choose the CBF 170 over the CBF 260 or a CB series model?

Choose CBF 170 for 170 L capacity with active humidity control (50–95 % RH) and hot-air sterilization. The CBF 260 offers larger volume (260 L) for higher throughput. CB series models lack active humidification and are suitable for dry CO2 incubation without humidity regulation.

What are the requirements for connecting the CO2 gas supply and water supply?

CO2 supply requires a gas cylinder connection kit (accessory) and a pressure regulator. For humidification, an external water source is needed: BINDER-Water-Set, freshwater bag set, or water container set. Ensure gas inlet pressure is within specified limits and use clean water to prevent scale.

What maintenance is required and what accessories are not included in base delivery?

Annual maintenance including humidifier service and electrical safety test is required. A maintenance reminder appears after 8760 h or 2 years. Base delivery excludes gas cylinder connection kits, water supply sets, and the BINDER-Water-Set; these must be ordered separately. CO2 sensor replacement requires recalibration.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model CBF 170 en
Datasheet ·414,7 КБ ·9 pp.
View PDF
PDF
CB+CBF (E7.1) 01 2026 en
Manual ·19,4 МБ ·201 pp.
View PDF