| Temperature range |
+5 °C above ambient temperature to 300 °C
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Temperature uniformity at 150°C |
2.3 ± K
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Temperature fluctuation at 150°C |
0.4 ± K
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Heating up time to 150°C |
50 min
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Recovery time after door was opened for 30 s at 150°C |
19 min
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Air change rate at 100°C |
8 x/h
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Rated voltage |
230 V
p.2 ↗ |
ED056-230V |
| Rated voltage |
120 V
p.2 ↗ |
ED056UL-120V |
| Power frequency |
50/60 Hz
p.2 ↗ |
ED056-230V |
| Power frequency |
60 Hz
p.2 ↗ |
ED056UL-120V |
| Nominal power |
1,05 kW
p.2 ↗ |
ED056-230V |
| Nominal power |
1,15 kW
p.2 ↗ |
ED056UL-120V |
| Unit fuse |
6,3 A
p.2 ↗ |
ED056-230V |
| Unit fuse |
12,5 A
p.2 ↗ |
ED056UL-120V |
| Phase (nominal voltage) |
1~
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Interior volume |
57 L
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Net weight of the unit (empty) |
39 kg
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Load per rack |
30 kg
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Permitted load |
70 kg
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Wall clearance back |
160 mm
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Wall clearance sidewise |
100 mm
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Width net |
560 mm
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Height net |
625 mm
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Depth net |
565 mm
p.2 ↗ |
ED056-230V / ED056UL-120V |
| Interior width |
360 mm
p.3 ↗ |
ED056-230V / ED056UL-120V |